BD BBLTM Gram Crystal Violet 250mL-- IVD Primary stain for Staining Microorganisms, Catalog Number: 212525 - Product Usage: Gram Stain Kits and Reagents are used to stain microorganisms from cultures or specimens by the differential Gram method.
Published May 27, 2020 · Recall initiated: April 14, 2020
Summary
FDA records medical device Class II recall Z-2046-2020, published May 27, 2020 and initiated April 14, 2020. Recalling firm: Becton Dickinson & Co.; Status: Terminated; Reported quantity: 10,568 bottles; Lot: Lot Number/ Exp Date: 9269800 04/30/21; 9296589 05/31/21; 9317733 06/30/21; 9352487 07/31/21; Expanded 05/12/20- Lot Numbers: 9352484 07/31/21 9352492 07/31/21. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
May exhibit increased levels of artifacts which could be difficult to distinguish from gram positive cocci, increased levels of artifacts could result in a false-positive Gram stain
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Becton Dickinson & Co.
- Distributor
- Worldwide distribution - US Nationwide distribution and the countries of Canada, Taiwan, Hong Kong, Chile, EMEA, Thailand, China, Japan, Australia, New Zealand, Korea, Costa Rica, Mexico, Canada, and Singapore.
- Lot
- Lot Number/ Exp Date: 9269800 04/30/21; 9296589 05/31/21; 9317733 06/30/21; 9352487 07/31/21; Expanded 05/12/20- Lot Numbers: 9352484 07/31/21 9352492 07/31/21
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDABD Phoenix Panel NMIC-312. Catalog Number: 449082.October 21, 2026
- FDABD Phoenix Panel NMIC-306. Catalog Number: 449292.October 21, 2026
- FDABD GasPak EZ CO2 Pouch System. Catalog Number: 260684. for microbiologic useMay 6, 2026
- FDABD GasPak EZ Campy Pouch System. Catalog Number: 260685. for microbiologic useMay 6, 2026
- FDASELECTSILICONE" 100% SILICONE FOLEY CATHETER 2-WAY 8 Fr 3 ml - Product Usage: Medline Silicone Foley catheterMay 27, 2020
- FDAProfessional Urine Drug Control, Item code 88010 - Product Usage: The Alere Professional Cup Urine DrugMay 27, 2020
Questions about this recall
Was BD BBLTM Gram Crystal Violet 250mL-- IVD Primary stain for Staining Microorganisms, Catalog Number: 212525 - Product Usage: Gram Stain Kits and Reagents are used to stain microorganisms from cultures or specimens by the differential Gram method. recalled?
Yes. FDA published recall Z-2046-2020 on May 27, 2020. Reported hazard: May exhibit increased levels of artifacts which could be difficult to distinguish from gram positive cocci, increased levels of artifacts could result in a false-positive Gram stain
What should I do if I have this product?
Voluntary: Firm initiated · Letter