HeartMate® II LVAS with Pocket Controller; 106762 - HMII POCKET CONTROLLER, NA; UDI: 00813024011286 Intended to provide hemodynamic support in patients with end-stage, refractory left ventricular heart failure; intended for use inside or outside the hospital.

Published May 31, 2017 · Recall initiated: March 30, 2017

Summary

FDA records medical device Class I recall Z-2053-2017, published May 31, 2017 and initiated March 30, 2017. Recalling firm: Thoratec Corporation; Status: Terminated; Reported quantity: 24,077 active units in US; Lot: Serial Numbers PC-10255-B, PC-10329-B, PC-10357-B, PC-10476-B, PC-10478-B, PC-10486-B, PC-10518-B, PC-10519-B, PC-10524-B, PC-10525-B, PC-10538-B, PC-10540-B, PC-10544-B, PC-10626-B, PC-10629-B, PC-10973-C, PC-10976-C, PC-10977-C, PC-10986-C, PC-10988-C, PC-11017-C, PC-11018-C, PC-11019-C, PC-11027-C, PC-11041-C, PC-11044-C, PC-11050-C, PC-11055-C, PC-11056-C, PC-11063-C, PC-11068-C, PC-11073-C, PC-11077-C, PC-11082-C, PC-11083-C, PC-11085-C, PC-11086-C, PC-11090-C, PC-11104-C, PC-11105-C, PC-11106-C, PC-11136-C, PC-11142-C, PC-11150-C, PC-11151-C, PC-11152-C, PC-11156-C, PC-11163-C, PC-11164-C, PC-11168-C, PC-11200-C, PC-11209-C, PC-11228-C, PC-11231-C, PC-11233-C, PC-11234-C, PC-11236-C, PC-11237-C, PC-11239-C, PC-11240-C, PC-11246-C, PC-11248-C, PC-11249-C, PC-11251-C, PC-11253-C, PC-11255, PC-11256-C, PC-11258, PC-11261-C, PC-11262-C, PC-11263-C, PC-11264-C, PC-11265-C, PC-11268-C, PC-11270-C, PC-11272-C, PC-11273-C, PC-11275-C, PC-11276-C, PC-11278-C, PC-11279-C, PC-11282-C, PC-11283-C, PC-11340-C, PC-11343-C, PC-11349-C, PC-11351-C, PC-11352-C, PC-11353-C, PC-11357-C, PC-11361-C, PC-11373-C, PC-11378-C, PC-11379-C, PC-11380-C, PC-11382-C, PC-11389-C, PC-11390-C, PC-11391-C, PC-11392-C, PC-11393-C, PC-11394-C, PC-11396-C, PC-11397-C, PC-11398-C, PC-11404-C, PC-11414-C, PC-11417-C, PC-11419-C, PC-11422-C, PC-11428-C, PC-11432-C, PC-11436-C, PC-11437-C, PC-11439-C, PC-11443-C, PC-11455-C, PC-11456-C, PC-11457-C, PC-11459-C, PC-11468-C, PC-11469-C, PC-11474-C, PC-11476-C, PC-11477-C, PC-11481-C, PC-11489-C, PC-11495-C, PC-11497-C, PC-11499-C, PC-11500-C, PC-11501-C, PC-11502-C, PC-11506-C, PC-11507-C, PC-11508-C, PC-11509-C, PC-11510-C, PC-11511-C, PC-11518-C, PC-11519-C, PC-11520-C, PC-11532-C, PC-11536-C, PC-11541-C, PC-11542-C, PC-11547-C, PC-11551-C, PC-11552-C, PC-11553-C, PC-11554-C, PC-11559-C, PC-11560-C, PC-11572-C, PC-11576-C, PC-11579-C, PC-11583-C, PC-11596-C, PC-11597-C, PC-11598-C, PC-11603-C, PC-11607-C, PC-11614-C, PC-11632-C, PC-11637-C, PC-11638-C, PC-11642-C, PC-11656-C, PC-11673-C, PC-11690-C, PC-11693-C, PC-11694-C, PC-11695-C, PC-11701-C, PC-11704-C, PC-11706-C, PC-11707-C, PC-11714-C, PC-11716-C, PC-11717-C, PC-11723-C, PC-11724-C, PC-11727-C, PC-11729-C, PC-11734-C, PC-11736-C, PC-11739-C, PC-11740-C, PC-11741-C, PC-11744-C, PC-11745-C, PC-11746-C, PC-11747-C, PC-11749-C, PC-11771-C, PC-11773-C, PC-11775-C, PC-11776-C, PC-11777-C, PC-11780-C, PC-11790-C, PC-11793-C, PC-11794-C, PC-11796-C, PC-11797-C, PC-11798-C, PC-11801-C, PC-11803-C, PC-11804-C, PC-11810-C, PC-11812-C, PC-11816-C, PC-11819-C, PC-11842-C, PC-11869-C, PC-11886-C, PC-11888-C, PC-11889-C, PC-11905-C, PC-11907-C, PC-11909-C, PC-11910-C, PC-11913-C, PC-11915-C, PC-11918-C, PC-11922-C, PC-11923-C, PC-11927-C, PC-11928-C, PC-11929-C, PC-11930-C, PC-11931-C, PC-11939-C, PC-11940-C, PC-11988, PC-11993, PC-11997, PC-11998, PC-12002, PC-12003, PC-12004, PC-12005, PC-12006, PC-12009, PC-12017, PC-12018, PC-12019, PC-12020, PC-12021, PC-12029, PC-12034, PC-12036, PC-12038, PC-12040, PC-12041, PC-12047, PC-12049, PC-12053, PC-12054, PC-12063, PC-12067, PC-12068, PC-12076, PC-12077-D, PC-12078, PC-12081, PC-12083, PC-12087, PC-12091, PC-12092, PC-12093, PC-12096, PC-12098, PC-12101, PC-12108, PC-12111, PC-12113, PC-12115, PC-12118, PC-12121, PC-12123, PC-12131, PC-12136, PC-12145, PC-12148, PC-12154, PC-12155, PC-12158, PC-12159-D, PC-12160, PC-12165, PC-12166, PC-12168, PC-12169, PC-12171, PC-12175, PC-12176, PC-12177, PC-12181, PC-12183, PC-12184, PC-12185, PC-12187, PC-12190, PC-12203, PC-12207, PC-12209, PC-12211, PC-12216, PC-12218, PC-12219, PC-12221, PC-12222, PC-12227, PC-12229, PC-12230, PC-12231, PC-12232, PC-12235, PC-12237, PC-12239, PC-12241, PC-12242, PC-12243, PC-12253, PC-12256, PC-12258, PC-12265, PC-12266, PC-12269, PC-12270, PC-12272, PC-12273, PC-12274, PC-12276, PC-12283-E, PC-12296-E, PC-12304, PC-12305, PC-12306, PC-12307, PC-12308, PC-12310, PC-12312-E, PC-12313-E, PC-12335, PC-12339, PC-12340, PC-12341, PC-12343, PC-12344. Class I means a reasonable probability of serious adverse health consequences or death.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

St. Jude Medical is providing all users of their HeartMate II LVAS with Pocket Controller with new software and updates to hardware to make the exchange to a backup controller easier. This is related to recall Z-1227/1230-2014.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Thoratec Corporation
Distributor
Worldwide Distribution -- US and Austria, Belgium, Brazil, Canada, Cayman Islands, Chile, Columbia, Croatia, Czech Republic, Denmark, Egypt, Estonia, France, Germany, Greece, Hong Kong, Hungary, India, Iran, Ireland, Israel, Italy, Kazakhstan, Kuwait, Latvia, Lebanon, Liechtenstein, Lithuania, Malaysia, Orman, Palastine, Poland, Qatar, Russian Federation, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, Spain, Sweden, Switzerland, Thailand, The Netherlands, Turkey, United Arab Emirates, and United Kingdom.
Lot
Serial Numbers PC-10255-B, PC-10329-B, PC-10357-B, PC-10476-B, PC-10478-B, PC-10486-B, PC-10518-B, PC-10519-B, PC-10524-B, PC-10525-B, PC-10538-B, PC-10540-B, PC-10544-B, PC-10626-B, PC-10629-B, PC-10973-C, PC-10976-C, PC-10977-C, PC-10986-C, PC-10988-C, PC-11017-C, PC-11018-C, PC-11019-C, PC-11027-C, PC-11041-C, PC-11044-C, PC-11050-C, PC-11055-C, PC-11056-C, PC-11063-C, PC-11068-C, PC-11073-C, PC-11077-C, PC-11082-C, PC-11083-C, PC-11085-C, PC-11086-C, PC-11090-C, PC-11104-C, PC-11105-C, PC-11106-C, PC-11136-C, PC-11142-C, PC-11150-C, PC-11151-C, PC-11152-C, PC-11156-C, PC-11163-C, PC-11164-C, PC-11168-C, PC-11200-C, PC-11209-C, PC-11228-C, PC-11231-C, PC-11233-C, PC-11234-C, PC-11236-C, PC-11237-C, PC-11239-C, PC-11240-C, PC-11246-C, PC-11248-C, PC-11249-C, PC-11251-C, PC-11253-C, PC-11255, PC-11256-C, PC-11258, PC-11261-C, PC-11262-C, PC-11263-C, PC-11264-C, PC-11265-C, PC-11268-C, PC-11270-C, PC-11272-C, PC-11273-C, PC-11275-C, PC-11276-C, PC-11278-C, PC-11279-C, PC-11282-C, PC-11283-C, PC-11340-C, PC-11343-C, PC-11349-C, PC-11351-C, PC-11352-C, PC-11353-C, PC-11357-C, PC-11361-C, PC-11373-C, PC-11378-C, PC-11379-C, PC-11380-C, PC-11382-C, PC-11389-C, PC-11390-C, PC-11391-C, PC-11392-C, PC-11393-C, PC-11394-C, PC-11396-C, PC-11397-C, PC-11398-C, PC-11404-C, PC-11414-C, PC-11417-C, PC-11419-C, PC-11422-C, PC-11428-C, PC-11432-C, PC-11436-C, PC-11437-C, PC-11439-C, PC-11443-C, PC-11455-C, PC-11456-C, PC-11457-C, PC-11459-C, PC-11468-C, PC-11469-C, PC-11474-C, PC-11476-C, PC-11477-C, PC-11481-C, PC-11489-C, PC-11495-C, PC-11497-C, PC-11499-C, PC-11500-C, PC-11501-C, PC-11502-C, PC-11506-C, PC-11507-C, PC-11508-C, PC-11509-C, PC-11510-C, PC-11511-C, PC-11518-C, PC-11519-C, PC-11520-C, PC-11532-C, PC-11536-C, PC-11541-C, PC-11542-C, PC-11547-C, PC-11551-C, PC-11552-C, PC-11553-C, PC-11554-C, PC-11559-C, PC-11560-C, PC-11572-C, PC-11576-C, PC-11579-C, PC-11583-C, PC-11596-C, PC-11597-C, PC-11598-C, PC-11603-C, PC-11607-C, PC-11614-C, PC-11632-C, PC-11637-C, PC-11638-C, PC-11642-C, PC-11656-C, PC-11673-C, PC-11690-C, PC-11693-C, PC-11694-C, PC-11695-C, PC-11701-C, PC-11704-C, PC-11706-C, PC-11707-C, PC-11714-C, PC-11716-C, PC-11717-C, PC-11723-C, PC-11724-C, PC-11727-C, PC-11729-C, PC-11734-C, PC-11736-C, PC-11739-C, PC-11740-C, PC-11741-C, PC-11744-C, PC-11745-C, PC-11746-C, PC-11747-C, PC-11749-C, PC-11771-C, PC-11773-C, PC-11775-C, PC-11776-C, PC-11777-C, PC-11780-C, PC-11790-C, PC-11793-C, PC-11794-C, PC-11796-C, PC-11797-C, PC-11798-C, PC-11801-C, PC-11803-C, PC-11804-C, PC-11810-C, PC-11812-C, PC-11816-C, PC-11819-C, PC-11842-C, PC-11869-C, PC-11886-C, PC-11888-C, PC-11889-C, PC-11905-C, PC-11907-C, PC-11909-C, PC-11910-C, PC-11913-C, PC-11915-C, PC-11918-C, PC-11922-C, PC-11923-C, PC-11927-C, PC-11928-C, PC-11929-C, PC-11930-C, PC-11931-C, PC-11939-C, PC-11940-C, PC-11988, PC-11993, PC-11997, PC-11998, PC-12002, PC-12003, PC-12004, PC-12005, PC-12006, PC-12009, PC-12017, PC-12018, PC-12019, PC-12020, PC-12021, PC-12029, PC-12034, PC-12036, PC-12038, PC-12040, PC-12041, PC-12047, PC-12049, PC-12053, PC-12054, PC-12063, PC-12067, PC-12068, PC-12076, PC-12077-D, PC-12078, PC-12081, PC-12083, PC-12087, PC-12091, PC-12092, PC-12093, PC-12096, PC-12098, PC-12101, PC-12108, PC-12111, PC-12113, PC-12115, PC-12118, PC-12121, PC-12123, PC-12131, PC-12136, PC-12145, PC-12148, PC-12154, PC-12155, PC-12158, PC-12159-D, PC-12160, PC-12165, PC-12166, PC-12168, PC-12169, PC-12171, PC-12175, PC-12176, PC-12177, PC-12181, PC-12183, PC-12184, PC-12185, PC-12187, PC-12190, PC-12203, PC-12207, PC-12209, PC-12211, PC-12216, PC-12218, PC-12219, PC-12221, PC-12222, PC-12227, PC-12229, PC-12230, PC-12231, PC-12232, PC-12235, PC-12237, PC-12239, PC-12241, PC-12242, PC-12243, PC-12253, PC-12256, PC-12258, PC-12265, PC-12266, PC-12269, PC-12270, PC-12272, PC-12273, PC-12274, PC-12276, PC-12283-E, PC-12296-E, PC-12304, PC-12305, PC-12306, PC-12307, PC-12308, PC-12310, PC-12312-E, PC-12313-E, PC-12335, PC-12339, PC-12340, PC-12341, PC-12343, PC-12344
Risk classification
Class I
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was HeartMate® II LVAS with Pocket Controller; 106762 - HMII POCKET CONTROLLER, NA; UDI: 00813024011286 Intended to provide hemodynamic support in patients with end-stage, refractory left ventricular heart failure; intended for use inside or outside the hospital. recalled?

Yes. FDA published recall Z-2053-2017 on May 31, 2017. Reported hazard: St. Jude Medical is providing all users of their HeartMate II LVAS with Pocket Controller with new software and updates to hardware to make the exchange to a backup controller easier. This is related to recall Z-1227/1230-2014.

What should I do if I have this product?

Voluntary: Firm initiated · Letter