00 4309-049-00 Trabecular Metal Reverse Shoulder Glenosphere Distractor Rx, Non sterile (sterilize before use) The Glenosphere Distractor must be able to attach, connect, or mate with the Glenosphere and remove it without dislodging or loosening the base plate from the glenoid, or damaging the glenoid. Pulling the trigger releases a weight inside the Distractor. This force is transmitted to the bottom of the glenosphere dislodging it from the base plate.
Published September 4, 2013 · Recall initiated: August 1, 2013
Summary
FDA records medical device Class II recall Z-2055-2013, published September 4, 2013 and initiated August 1, 2013. Recalling firm: Zimmer, Inc.; Status: Terminated; Reported quantity: 771; Lot: Item # 00-4309-049-00 and Lot Numbers: 60452884, 60461858, 60531990, 60549559, 60605842, 60612781, 60684768, 60754281, 60815778, 60909547, 60959086, 60978338, 61029048, 61135953, 61213252, 61274877, 61334929, 61431672, 61566464, 61594740, 61690097, 61756058, 61857748, 61858851, 61889174, 61906350, 61910781, 61918666 62039450, 62096379, 62101249, 62141887, 62160965, 62215378, 62254111, 62269856, 62304206, and 62311096. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Zimmer is initiating a lot specific recall of the Trabecular Metal" Reverse Glenosphere Distractor due to the potential of the device not functioning properly should it be utilized without proper lubrication. As a result, there is a potential for the device to fail to provide impact to the Glenosphere when the trigger is pulled. In addition, there is a potential for the foot of the device to fract
What to do
Voluntary: Firm initiated
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Zimmer, Inc.
- Distributor
- Worldwide Distribution - USA (nationwide) including the countries of Singapore, Brazil, Taiwan, Australia, Canada, Germany(with further distribution to UAE, Austria, Belgium, Switzerland, Czech Republic, Denmark, Spain, Finland, France, Greece, Israel, Italy, Malta, Norway, Portugal, Sweden and South Africa,
- Lot
- Item # 00-4309-049-00 and Lot Numbers: 60452884, 60461858, 60531990, 60549559, 60605842, 60612781, 60684768, 60754281, 60815778, 60909547, 60959086, 60978338, 61029048, 61135953, 61213252, 61274877, 61334929, 61431672, 61566464, 61594740, 61690097, 61756058, 61857748, 61858851, 61889174, 61906350, 61910781, 61918666 62039450, 62096379, 62101249, 62141887, 62160965, 62215378, 62254111, 62269856, 62304206, and 62311096.
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAOff-Axis Comprehensive Shoulder System Small, Augment, Off-Axis, Reamer Guide, Model/Catalog Number: 110040240July 1, 2026
- FDABrand Name: Persona Revision Product Name: Persona Revision® Trabecular Metal® Femoral Distal AugmentJuly 1, 2026
- FDAOff-Axis Alliance Glenoid, Left, 4-Peg, Augment, Reamer Guide, Model/Catalog Number: 110040444July 1, 2026
- FDAOff-Axis Comprehensive Shoulder System, Medium, Augment, Off-Axis, Reamer Guide Model/Catalog Number:July 1, 2026
- FDAStryker Part Numbers 79-43903 (3cc), 79-43905 (5cc), 79-43910 (10 cc), 79-43915 (15cc), 397003 (3cc), 397005September 4, 2013
- FDAIntegra Leyla Ball Joint Clamp, also as included in Leyla Table Mounting Hardware. The Leyla RetractionSeptember 4, 2013
Questions about this recall
Was 00 4309-049-00 Trabecular Metal Reverse Shoulder Glenosphere Distractor Rx, Non sterile (sterilize before use) The Glenosphere Distractor must be able to attach, connect, or mate with the Glenosphere and remove it without dislodging or loosening the base plate from the glenoid, or damaging the glenoid. Pulling the trigger releases a weight inside the Distractor. This force is transmitted to the bottom of the glenosphere dislodging it from the base plate. recalled?
Yes. FDA published recall Z-2055-2013 on September 4, 2013. Reported hazard: Zimmer is initiating a lot specific recall of the Trabecular Metal" Reverse Glenosphere Distractor due to the potential of the device not functioning properly should it be utilized without proper lubrication. As a result, there is a potential for the device to fail to provide impact to the Glenosphere when the trigger is pulled. In addition, there is a potential for the foot of the device to fract
What should I do if I have this product?
Voluntary: Firm initiated