Vigilant Agilia, Vigilant Drug'Lib, REF Z073476, versions 1.0 and 1.1 Product Usage: Vigilant Drug Lib is Dose Error Reduction Software that, when used in combination with Volumat MC Agilia infusion pumps, is intended to reduce drug administration errors at the bedside. Vigilant Drug Lib Agilia is a software application running on a computer and intended to configure and generate multiple device configurations, drug libraries, care areas and data sets that can then be uploaded to the Volumat MC Agilia infusion pumps. Intended patient population: The Vigilant Drug Lib software must only be used to create drug libraries for patients whose weight is 250g to 250 kg (0.6 lbs. to 551 lbs.).
Published August 21, 2019 · Recall initiated: June 24, 2019
Summary
FDA records medical device Class I recall Z-2065-2019, published August 21, 2019 and initiated June 24, 2019. Recalling firm: Fenwal Inc; Status: Terminated; Reported quantity: 83 units; Lot: All serial numbers distributed in the US. Class I means a reasonable probability of serious adverse health consequences or death.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
The firm is correcting four software anomalies and KVO (keep vein open) end of infusion alarm priority in Software versions 1.7 and 1.9a of Volumat MC Agilia Software and versions 1.0 and 1.1 of Vigilant DrugLib
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Fenwal Inc
- Distributor
- US Nationwide Distribution
- Lot
- All serial numbers distributed in the US
- Risk classification
- Class I
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAFreshness Guaranteed Chocolate Chunk cookie Clear plastic clam shell Net wt. 14 oz. UPC: 0007874209234August 21, 2019
- FDAOrthopedic Salvage System (OSS) Hybrid Poly Tibia Item # 151811August 21, 2019
- FDASlim Chickens Breader, Item Number 30753878.49, Net Wt. 50 lb, Distributed by Kerry Ingredients & Flavours,August 21, 2019
- FDAOrthopedic Salvage System (OSS) Hybrid Poly Tibia Item # 151812August 21, 2019
- FDAPREDUST G4627, Item Number G4627, Net Wt. 50 lb, Distributed by Kerry Ingredients & Flavours, Beloit, WIAugust 21, 2019
- FDAAMICUS Exchange Kit; Product Usage: This kit is designed for use with the AMICUS separator for the TherapeuticJune 19, 2013
Questions about this recall
Was Vigilant Agilia, Vigilant Drug'Lib, REF Z073476, versions 1.0 and 1.1 Product Usage: Vigilant Drug Lib is Dose Error Reduction Software that, when used in combination with Volumat MC Agilia infusion pumps, is intended to reduce drug administration errors at the bedside. Vigilant Drug Lib Agilia is a software application running on a computer and intended to configure and generate multiple device configurations, drug libraries, care areas and data sets that can then be uploaded to the Volumat MC Agilia infusion pumps. Intended patient population: The Vigilant Drug Lib software must only be used to create drug libraries for patients whose weight is 250g to 250 kg (0.6 lbs. to 551 lbs.). recalled?
Yes. FDA published recall Z-2065-2019 on August 21, 2019. Reported hazard: The firm is correcting four software anomalies and KVO (keep vein open) end of infusion alarm priority in Software versions 1.7 and 1.9a of Volumat MC Agilia Software and versions 1.0 and 1.1 of Vigilant DrugLib
What should I do if I have this product?
Voluntary: Firm initiated · Letter