Microfil Composite Instruments are packaged in an unsealed plastic sleeve, which is then placed in a cardboard box. Microfil Composite Instruments are intended to be sterilized by the dentist prior to each use. It is used for composite restoration development. The thin paddles shape light-cured composites. Paddles slide under the gum line to shape the margin. The flat ends contour the surfaces of the teeth and the thin edge is used to create esthetic highlights. Models: #96041 Gold Microfil #96042 Green Microfil #96043 Blue Microfil #96044 Set of 3 Microfil (Gold, Green, Blue)
Published May 17, 2017 · Recall initiated: April 20, 2017
Summary
FDA records medical device Class II recall Z-2069-2017, published May 17, 2017 and initiated April 20, 2017. Recalling firm: Almore International Inc; Status: Terminated; Reported quantity: 1773 devices (US =1570; International = 203); Lot: no lot codes. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Microfil Composite Instruments were distributed with a sterile symbol on their packaging label inserts, indicating that the products are sold as sterile devices. These products are not sterile and they are intended to be sterile prior to each use.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Almore International Inc
- Distributor
- Worldwide distributed - US Nationwide in AZ, CA, CO, CT, FL, GA, IA, IL, IN, MA, MN, MO, NC, NJ, NV, NY, OH, PA, TN, TX, WA, and WI, and internationally in Canada, Netherlands, and South Korea.
- Lot
- no lot codes
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDA5-MTHF *10ML* MDV 5MG/ML INJ, Injection, 5mg/mL, Rx only, 10mL Glass/Multiple Dose vial, Prepared by KeyMay 24, 2017
- FDARegenerex Patella RGX 3 PEG SER A PATELLA 28MM RGX 3 PEG SER A PATELLA 31MM RGX 3 PEG SER A PATELLA 34MM RGX 3May 17, 2017
- FDAMentor Smooth Round SPECTRUM Post- Operatively Adjustable Saline Breast Implant, Product Code: 350-1490,May 17, 2017
- FDAHealon Pro, Part No. 10270012May 17, 2017
- FDAStryker PROFESS Registration Kit. The Stryker PROFESS Registration Kit is a member of the Navigation portfolioMay 17, 2017
- FDAHealon 5 Pro, Part No. 10270015May 17, 2017
Questions about this recall
Was Microfil Composite Instruments are packaged in an unsealed plastic sleeve, which is then placed in a cardboard box. Microfil Composite Instruments are intended to be sterilized by the dentist prior to each use. It is used for composite restoration development. The thin paddles shape light-cured composites. Paddles slide under the gum line to shape the margin. The flat ends contour the surfaces of the teeth and the thin edge is used to create esthetic highlights. Models: #96041 Gold Microfil #96042 Green Microfil #96043 Blue Microfil #96044 Set of 3 Microfil (Gold, Green, Blue) recalled?
Yes. FDA published recall Z-2069-2017 on May 17, 2017. Reported hazard: Microfil Composite Instruments were distributed with a sterile symbol on their packaging label inserts, indicating that the products are sold as sterile devices. These products are not sterile and they are intended to be sterile prior to each use.
What should I do if I have this product?
Voluntary: Firm initiated · Letter