Back to search
FDAFDA · Food and drugs·Recall number Z-2080-2015

Sorin Stockert Heater-Cooler 3T, 120 V / 60 Hz Temperature control for extracorporeal perfusion of durations up to 6 hours.

Published July 22, 2015 · Recall initiated: June 15, 2015

Summary

FDA records medical device Class II recall Z-2080-2015, published July 22, 2015 and initiated June 15, 2015. Recalling firm: Sorin Group USA, Inc.; Status: Terminated; Reported quantity: 1755; Lot: Product code 16-02-85 Serial number 16S10958-16S15634. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Potential colonization of organisms, including Mycobacteria, in Sorin Heater Cooler Devices, if proper disinfection and maintenance is not performed per Instructions for Use.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Sorin Group USA, Inc.
Distributor
Worldwide Distribution - US (nationwide) and Internationally to AL, AT, BE, BG, CH, CY, CZ, DE, DK, EE, ES, FI, FR, GB, GR, HR, HU, IE, IT, LT, LU, LV, MC, MK, MT, NL, NO, PL, PT, RO, RS, SE, SK, AE, AR, AU, AZ, BD, BH, BR, BY, CA, CL, CN, CO, CR, DZ, EC, EG, ET, GE, GY, HK, ID, IL, IN, IQ, IR, JO, JP, KR, KW, KZ, LB, LK, LY, MA, MN, MU, MX, MY, NG, NP, NZ, OM, PA, PE, PH, PK, PR, PS, QA, RE, RU, SA, SG, SV, SY, TH, TN, TR, TT, TW, UA, VN, ZA.
Lot
Product code 16-02-85 Serial number 16S10958-16S15634
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was Sorin Stockert Heater-Cooler 3T, 120 V / 60 Hz Temperature control for extracorporeal perfusion of durations up to 6 hours. recalled?

Yes. FDA published recall Z-2080-2015 on July 22, 2015. Reported hazard: Potential colonization of organisms, including Mycobacteria, in Sorin Heater Cooler Devices, if proper disinfection and maintenance is not performed per Instructions for Use.

What should I do if I have this product?

Voluntary: Firm initiated · Letter