FDAFDA · Food and drugs·Recall number Z-2093-2026

See RES for complete list. Medline Convenience Kits: 1) OPEN HEART CDS, Model Number:CDS840015AQ; 2) OPEN HEART CDS, Model Number:CDS840015AQ; 3) OPEN HEART CDS, Model Number:CDS840023T; 4) OFF PUMP CABG CDS, Model Number:CDS840087AI; 5) OFF PUMP CABG CDS, Model Number:CDS840087AK; 6) OFF PUMP CABG CDS, Model Number:CDS840087AL; 7) OFF PUMP CABG CDS, Model Number:CDS840087AM; 8) OPEN HEART, Model Number:CDS840150S; 9) OPEN HEART, Model Number:CDS840150T;

Published May 13, 2026 · Recall initiated: January 7, 2026

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.

What to do

Voluntary: Firm initiated · Letter

Recall details

Manufacturer
Medline Industries, LP
Distributor
Worldwide - US Nationwide distribution.
Lot
1) CDS840015AQ, UDI-DI: 10195327210144(each), 40195327210145(case), Lot Number: 23FLA191; 2) CDS840015AQ, UDI-DI: 10195327210144(each), 40195327210145(case), Lot Number: 24ILB176; 3) CDS840023T, UDI-DI: 10193489682069(each), 40193489682060(case), Lot Number: 21CKA394; 4) CDS840023T, UDI-DI: 10193489682069(each), 40193489682060(case), Lot Number: 21EBN817; 5) CDS840023T, UDI-DI: 10193489682069(each), 40193489682060(case), Lot Number: 21GBH886; 6) CDS840023T, UDI-DI: 10193489682069(each), 40193489682060(case), Lot Number: 21IBG586; 7) CDS840023T, UDI-DI: 10193489682069(each), 40193489682060(case), Lot Number: 21KBW667; 8) CDS840023T, UDI-DI: 10193489682069(each), 40193489682060(case), Lot Number: 21LBV113; 9) CDS840023T, UDI-DI: 10193489682069(each), 40193489682060(case), Lot Number: 22CBC309; 10) CDS840023T, UDI-DI: 10193489682069(each), 40193489682060(case), Lot Number: 22EBK678; 11) CDS840023T, UDI-DI: 10193489682069(each), 40193489682060(case), Lot Number: 22GBQ917; 12) CDS840023T, UDI-DI: 10193489682069(each), 40193489682060(case), Lot Number: 22HBU935; 13) CDS840023T, UDI-DI: 10193489682069(each), 40193489682060(case), Lot Number: 22JBT770; 14) CDS840023T, UDI-DI: 10193489682069(each), 40193489682060(case), Lot Number: 22JBU353; 15) CDS840023T, UDI-DI: 10193489682069(each), 40193489682060(case), Lot Number: 22KBJ798; 16) CDS840023T, UDI-DI: 10193489682069(each), 40193489682060(case), Lot Number: 22KBL256; 17) CDS840023T, UDI-DI: 10193489682069(each), 40193489682060(case), Lot Number: 22KBW565; 18) CDS840023T, UDI-DI: 10193489682069(each), 40193489682060(case), Lot Number: 22OBM690; 19) CDS840023T, UDI-DI: 10193489682069(each), 40193489682060(case), Lot Number: 23ABE738; 20) CDS840023T, UDI-DI: 10193489682069(each), 40193489682060(case), Lot Number: 23ABM709; 21) CDS840023T, UDI-DI: 10193489682069(each), 40193489682060(case), Lot Number: 23CBO830; 22) CDS840023T, UDI-DI: 10193489682069(each), 40193489682060(case), Lot Number: 23GBA105; 23) CDS840023T, UDI-DI: 10193489682069(each), 40193489682060(case), Lot Number: 23IBJ571; 24) CDS840023T, UDI-DI: 10193489682069(each), 40193489682060(case), Lot Number: 23IBJ759; 25) CDS840023T, UDI-DI: 10193489682069(each), 40193489682060(case), Lot Number: 23JBU616; 26) CDS840023T, UDI-DI: 10193489682069(each), 40193489682060(case), Lot Number: 23KBM369; 27) CDS840023T, UDI-DI: 10193489682069(each), 40193489682060(case), Lot Number: 23LBM199; 28) CDS840087AI, UDI-DI: 10193489847413(each), 40193489847414(case), Lot Number: 21DLA296; 29) CDS840087AI, UDI-DI: 10193489847413(each), 40193489847414(case), Lot Number: 21ELA419; 30) CDS840087AK, UDI-DI: 10195327212353(each), 40195327212354(case), Lot Number: 23BLA242; 31) CDS840087AK, UDI-DI: 10195327212353(each), 40195327212354(case), Lot Number: 23BLA602; 32) CDS840087AK, UDI-DI: 10195327212353(each), 40195327212354(case), Lot Number: 23CLB259; 33) CDS840087AK, UDI-DI: 10195327212353(each), 40195327212354(case), Lot Number: 23ELA870; 34) CDS840087AK, UDI-DI: 10195327212353(each), 40195327212354(case), Lot Number: 23FLA144; 35) CDS840087AK, UDI-DI: 10195327212353(each), 40195327212354(case), Lot Number: 23GLA152; 36) CDS840087AK, UDI-DI: 10195327212353(each), 40195327212354(case), Lot Number: 23GLB213; 37) CDS840087AK, UDI-DI: 10195327212353(each), 40195327212354(case), Lot Number: 23HLB120; 38) CDS840087AK, UDI-DI: 10195327212353(each), 40195327212354(case), Lot Number: 23ILA911; 39) CDS840087AK, UDI-DI: 10195327212353(each), 40195327212354(case), Lot Number: 23LLA114; 40) CDS840087AK, UDI-DI: 10195327212353(each), 40195327212354(case), Lot Number: 24ALA030; 41) CDS840087AL, UDI-DI: 10195327590406(each), 40195327590407(case), Lot Number: 24ALA755; 42) CDS840087AL, UDI-DI: 10195327590406(each), 40195327590407(case), Lot Number: 24CLA098; 43) CDS840087AL, UDI-DI: 10195327590406(each), 40195327590407(case), Lot Number: 24CLA726; 44) CDS840087AL, UDI-DI: 10195327590406(each), 40195327590407(case), Lot Number: 24DLA309; 45) CDS840087AL, UDI-DI: 10195
Risk classification
Class II
Area (FDA)
device
Country
United States