Back to search
FDAFDA · Food and drugs·Recall number Z-2095-2026

Medline Convenience Kits: 1) KIT SURG ONC EXCISION BREAST, Model Number:DYKMBNDL84A; 2) KIT SURG ONC EXPLORATORY LAP, Model Number:DYKMBNDL87A; 3) CT BIOPSY TRAY, Model Number:DYNDH1143C; 4) BIOPSY SPECIALS PACK, Model Number:DYNDH1873; 5) BIOPSY SPECIALS PACK, Model Number:DYNDH1934; 6) BREAST BIOPSY PACK, Model Number:DYNJ17489I; 7) C T BIOPSY TRAY-LF, Model Number:DYNJ17558D; 8) RIGHT HEART BIOPSY TRAY, Model Number:DYNJ19063K; 9) VOR PACK-LF, Model Number:DYNJ48539; 10) VOR PACK-LF, Model Number:DYNJ48539A; 11) IR BIOPSY PACK, Model Number:DYNJ59897B; 12) BREAST PACK, Model Number:DYNJ82998; 13) LIVER BIOPSY KIT, Model Number:DYNJ910654

Published May 13, 2026 · Recall initiated: January 7, 2026

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.

What to do

Voluntary: Firm initiated · Letter

Recall details

Manufacturer
Medline Industries, LP
Distributor
Worldwide - US Nationwide distribution.
Lot
1) DYKMBNDL84A, UDI-DI: 10195327240486(each), 40195327240487(case), Lot Number: 22IBQ267; 2) DYKMBNDL84A, UDI-DI: 10195327240486(each), 40195327240487(case), Lot Number: 22IBU924; 3) DYKMBNDL84A, UDI-DI: 10195327240486(each), 40195327240487(case), Lot Number: 23ABK216; 4) DYKMBNDL87A, UDI-DI: 10195327240479(each), 40195327240470(case), Lot Number: 22LBK913; 5) DYKMBNDL87A, UDI-DI: 10195327240479(each), 40195327240470(case), Lot Number: 22LBM302; 6) DYKMBNDL87A, UDI-DI: 10195327240479(each), 40195327240470(case), Lot Number: 23ABU170; 7) DYKMBNDL87A, UDI-DI: 10195327240479(each), 40195327240470(case), Lot Number: 23CBC364; 8) DYKMBNDL87A, UDI-DI: 10195327240479(each), 40195327240470(case), Lot Number: 23CBL439; 9) DYNDH1143C, UDI-DI: 10193489193336(each), 40193489193337(case), Lot Number: 22GBZ268; 10) DYNDH1873, UDI-DI: 10195327486464(each), 40195327486465(case), Lot Number: 23KBK476; 11) DYNDH1934, UDI-DI: 10195327519728(each), 40195327519729(case), Lot Number: 24FMF173; 12) DYNJ17489I, UDI-DI: 10195327224783(each), 40195327224784(case), Lot Number: 23CBL337; 13) DYNJ17558D, UDI-DI: 10195327504205(each), 40195327504206(case), Lot Number: 23JLB107; 14) DYNJ19063K, UDI-DI: 10888277828025(each), 40888277828026(case), Lot Number: 24DBL079; 15) DYNJ48539, UDI-DI: 10888277906228(each), 40888277906229(case), Lot Number: 20LME482; 16) DYNJ48539, UDI-DI: 10888277906228(each), 40888277906229(case), Lot Number: 20LMF497; 17) DYNJ48539, UDI-DI: 10888277906228(each), 40888277906229(case), Lot Number: 21JMC430; 18) DYNJ48539, UDI-DI: 10888277906228(each), 40888277906229(case), Lot Number: 22BMH767; 19) DYNJ48539, UDI-DI: 10888277906228(each), 40888277906229(case), Lot Number: 22EMD675; 20) DYNJ48539, UDI-DI: 10888277906228(each), 40888277906229(case), Lot Number: 23AMC919; 21) DYNJ48539, UDI-DI: 10888277906228(each), 40888277906229(case), Lot Number: 23DMD933; 22) DYNJ48539, UDI-DI: 10888277906228(each), 40888277906229(case), Lot Number: 23GBI270; 23) DYNJ48539, UDI-DI: 10888277906228(each), 40888277906229(case), Lot Number: 23HBR688; 24) DYNJ48539, UDI-DI: 10888277906228(each), 40888277906229(case), Lot Number: 23KBD033; 25) DYNJ48539, UDI-DI: 10888277906228(each), 40888277906229(case), Lot Number: 24ABL751; 26) DYNJ48539, UDI-DI: 10888277906228(each), 40888277906229(case), Lot Number: 24DBK656; 27) DYNJ48539, UDI-DI: 10888277906228(each), 40888277906229(case), Lot Number: 24EBQ366; 28) DYNJ48539, UDI-DI: 10888277906228(each), 40888277906229(case), Lot Number: 24GBD309; 29) DYNJ48539A, UDI-DI: 10198459112133(each), 40198459112134(case), Lot Number: 24IBT900; 30) DYNJ48539A, UDI-DI: 10198459112133(each), 40198459112134(case), Lot Number: 25EBA588; 31) DYNJ48539A, UDI-DI: 10198459112133(each), 40198459112134(case), Lot Number: 25HBK043; 32) DYNJ48539A, UDI-DI: 10198459112133(each), 40198459112134(case), Lot Number: 25IBA545; 33) DYNJ48539A, UDI-DI: 10198459112133(each), 40198459112134(case), Lot Number: 25JBA908; 34) DYNJ59897B, UDI-DI: 10195327043179(each), 40195327043170(case), Lot Number: 22LBH614; 35) DYNJ59897B, UDI-DI: 10195327043179(each), 40195327043170(case), Lot Number: 23EBQ760; 36) DYNJ59897B, UDI-DI: 10195327043179(each), 40195327043170(case), Lot Number: 23FBT785; 37) DYNJ59897B, UDI-DI: 10195327043179(each), 40195327043170(case), Lot Number: 23LBA546; 38) DYNJ82998, UDI-DI: 10195327222987(each), 40195327222988(case), Lot Number: 22JMG410; 39) DYNJ82998, UDI-DI: 10195327222987(each), 40195327222988(case), Lot Number: 22LMF004; 40) DYNJ910654, UDI-DI: 10198459055843(each), 40198459055844(case), Lot Number: 25CBM821; 41) DYNJ910654, UDI-DI: 10198459055843(each), 40198459055844(case), Lot Number: 25DBP029; 42) DYNJ910654, UDI-DI: 10198459055843(each), 40198459055844(case), Lot Number: 25EBU142; 43) DYNJ910654, UDI-DI: 10198459055843(each), 40198459055844(case), Lot Number: 25GBO011; 44) DYNJ910654, UDI-DI: 10198459055843(each), 40198459055844(case), Lot Number: 25HBG682
Risk classification
Class II
Area (FDA)
device
Country
United States