Maquet Getinge Group Operating Table System Magnus Carbon-Fibre Table Top Version 1180.16 The carbon-fibre MAGNUS table top (1180-16XX) serves to support and position patients for surgical procedures (e.g. cardiovascular surgery) and interventional examination immediately before, during and after the surgical phase. The table top is radiotranslucent and enables intraoperative use of X-ray equipment.
Published August 8, 2012 · Recall initiated: April 13, 2012
Summary
FDA records medical device Class II recall Z-2108-2012, published August 8, 2012 and initiated April 13, 2012. Recalling firm: Maquet Medical Systems, Usa; Status: Terminated; Reported quantity: 164 Units located in 27 Countries; Lot: Model # 1180.16F2 Serial #s 1,3,7, & 8; Model # 1180.16F3 Serial # 1; Model #1180.16F5 Serial #4; Model # T285.7000 Serial #s 1,2,3,4, & 7. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
As a result of a small number of customer complaints, MAQUET conducted an investigation and identified a potential problem with specific MAGNUS Carbon-Fibre Table Tops manufactured between February 23, 2010 and February 28, 2012 and distributed between November 11, 2010 and April 6, 2012. There is a potential issue with the motor brakes for the longitudinal shift drive, and the tilt drive.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Maquet Medical Systems, Usa
- Distributor
- Worldwide Distribution - USA (nationwide) and the countries of United Arab Emirates, Austria, Australia, Belgium, Brazil, Canada, Switzerland, China, Germany, Denmark, Spain, France, Hong Kong, Ireland, Israel, India, Italy, Japan, South Korea, Netherlands, Poland, Russia, Saudi Arabia, Sweden, Singapore, and Thailand.
- Lot
- Model # 1180.16F2 Serial #s 1,3,7, & 8; Model # 1180.16F3 Serial # 1; Model #1180.16F5 Serial #4; Model # T285.7000 Serial #s 1,2,3,4, & 7.
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDASmiths Medical, Medfusion® Syringe Infusion Pump, Model 4000, Software Version V1.0 & V1.1, Reorder NumbersAugust 8, 2012
- FDABeekley S-Spots Floral Skin Markers -Order Number: 750 Product Usage: Mammography Skin Markers for marking:August 8, 2012
- FDAACCESSORY BASE STARTER KIT,IS3000; Product Usage: The intended use of the Instrument arm drape is to protectAugust 8, 2012
- FDABeekley A-Spots Floral Skin Markers -Order Number: 751 Product Usage: Mammography Skin Markers for marking:August 8, 2012
- FDAKIT,DISPOSABLE ACCESSORY,4ARM,IS3000,5 PACK; Product Usage: The intended use of the Instrument arm drape is toAugust 8, 2012
- FDABeekley O-Spots Floral Skin Markers -Order Number: 752 Product Usage: Mammography Skin Markers for marking:August 8, 2012
Questions about this recall
Was Maquet Getinge Group Operating Table System Magnus Carbon-Fibre Table Top Version 1180.16 The carbon-fibre MAGNUS table top (1180-16XX) serves to support and position patients for surgical procedures (e.g. cardiovascular surgery) and interventional examination immediately before, during and after the surgical phase. The table top is radiotranslucent and enables intraoperative use of X-ray equipment. recalled?
Yes. FDA published recall Z-2108-2012 on August 8, 2012. Reported hazard: As a result of a small number of customer complaints, MAQUET conducted an investigation and identified a potential problem with specific MAGNUS Carbon-Fibre Table Tops manufactured between February 23, 2010 and February 28, 2012 and distributed between November 11, 2010 and April 6, 2012. There is a potential issue with the motor brakes for the longitudinal shift drive, and the tilt drive.
What should I do if I have this product?
Voluntary: Firm initiated · Letter