FDAFDA · Food and drugs·Recall number Z-2108-2012
Maquet Getinge Group Operating Table System Magnus Carbon-Fibre Table Top Version 1180.16 The carbon-fibre MAGNUS table top (1180-16XX) serves to support and position patients for surgical procedures (e.g. cardiovascular surgery) and interventional examination immediately before, during and after the surgical phase. The table top is radiotranslucent and enables intraoperative use of X-ray equipment.
Published August 8, 2012 · Recall initiated: April 13, 2012
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
As a result of a small number of customer complaints, MAQUET conducted an investigation and identified a potential problem with specific MAGNUS Carbon-Fibre Table Tops manufactured between February 23, 2010 and February 28, 2012 and distributed between November 11, 2010 and April 6, 2012. There is a potential issue with the motor brakes for the longitudinal shift drive, and the tilt drive.
What to do
Voluntary: Firm initiated · Letter
Recall details
- Manufacturer
- Maquet Medical Systems, Usa
- Distributor
- Worldwide Distribution - USA (nationwide) and the countries of United Arab Emirates, Austria, Australia, Belgium, Brazil, Canada, Switzerland, China, Germany, Denmark, Spain, France, Hong Kong, Ireland, Israel, India, Italy, Japan, South Korea, Netherlands, Poland, Russia, Saudi Arabia, Sweden, Singapore, and Thailand.
- Lot
- Model # 1180.16F2 Serial #s 1,3,7, & 8; Model # 1180.16F3 Serial # 1; Model #1180.16F5 Serial #4; Model # T285.7000 Serial #s 1,2,3,4, & 7.
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States