PrepStain Tecan US-I. For use in the screening and detection of cervical cancer.
Published September 4, 2013 · Recall initiated: July 8, 2013
Summary
FDA records medical device Class II recall Z-2111-2013, published September 4, 2013 and initiated July 8, 2013. Recalling firm: Becton Dickinson & Co.; Status: Terminated; Reported quantity: 890 instruments; Lot: Catalog Numbers - 799-13000-00 and 799-13000-00R. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
PrepStain system rack setting was incorrectly entered on one PrepStain system during a routine preventive maintenance visit by a BD Field Service Representative.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Becton Dickinson & Co.
- Distributor
- Worldwide Distribution-USA (nationwide) and Puerto Rico and the country of Canada.
- Lot
- Catalog Numbers - 799-13000-00 and 799-13000-00R
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDABD Phoenix Panel NMIC-312. Catalog Number: 449082.October 21, 2026
- FDABD Phoenix Panel NMIC-306. Catalog Number: 449292.October 21, 2026
- FDABD GasPak EZ CO2 Pouch System. Catalog Number: 260684. for microbiologic useMay 6, 2026
- FDABD GasPak EZ Campy Pouch System. Catalog Number: 260685. for microbiologic useMay 6, 2026
- FDAPrepStain AG. For use in the screening and detection of cervical cancer.September 4, 2013
- FDAPrepStain Tecan US-II. For use in the screening and detection of cervical cancer.September 4, 2013
Questions about this recall
Was PrepStain Tecan US-I. For use in the screening and detection of cervical cancer. recalled?
Yes. FDA published recall Z-2111-2013 on September 4, 2013. Reported hazard: PrepStain system rack setting was incorrectly entered on one PrepStain system during a routine preventive maintenance visit by a BD Field Service Representative.
What should I do if I have this product?
Voluntary: Firm initiated · Letter