Normed General Instruments 503004541: Werber Countersink Cannulated for micro CBS Screws, AO 503004341: Werber Countersink Cannulated for micro CBS Screws, round-shaft The Countersink is an instrument which is used over the guide wire to prepare for adequate space in the cortical bone rim to sink the screw head in to bones/tissues for different Foot, Ankle and Hand implant systems. The instrument is used in several Normed implant systems. The instrument is used to avoid soft tissue irritation

Published May 24, 2017 · Recall initiated: May 11, 2017

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

the devices used a different raw material to produce two countersink products which has higher hardness and higher carbon content resulting in device damage.

What to do

Voluntary: Firm initiated · Letter

Recall details

Manufacturer
Zimmer Gmbh
Distributor
Single US consignee CO.
Lot
Lots: 13065 13569 13571
Risk classification
Class II
Area (FDA)
device
Country
Switzerland