DeRoyal Insufflation Tubing sets: 1) REF 28-0206 Insufflation Tubing, Laparoscopic Rx Only 2) REF 28-0214 Hi-Flow Insufflation Tubing Rx Only 3) REF 28-0216 Tubing, Insuf w/.3 Micr Filter Rx Only 4) REF 28-0217 Insufflation Tubing, Laparoscopic Rx Only 5) REF 28-0208 Insufflation Tubing w/Vent Rx Only 6) REF 28-0212H Heated Hi-Flow Insufflation Tubing Rx Only. Used during laparoscopic surgical procedures to aid in the insufflation of the abdomen.
Published September 4, 2013 · Recall initiated: July 15, 2013
Summary
FDA records medical device Class II recall Z-2113-2013, published September 4, 2013 and initiated July 15, 2013. Recalling firm: DeRoyal Industries Inc; Status: Terminated; Reported quantity: 506,632 sets in US; 17,854 sets International; Lot: All Lots of the Insufflation Tubing Sets are as follows: REF 28-0206 Insufflation Tubing, Laparoscopic, Lot #29074214; REF 28-0214 Hi-Flow Insufflation Tubing Lot #3194170; REF 28-0216 Tubing, Insuf w/.3 Micr Filter Lot #29074249; REF 28-0217 Insufflation Tubing, Laparoscopic Lot #32305010; DeRoyal REF 28-0208 Insufflation Tubing w/Vent Lot #29314302; DeRoyal REF 28-0212H Heated Hi-Flow Insufflation Tubing Lot #29000354. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
DeRoyal Industries is recalling their product Insufflation Tubing Sets due to materials incompatibility which could possibly result in occlusion at or near the filter housing which could prevent the flow of gas through the device.
What to do
Voluntary: Firm initiated · Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- DeRoyal Industries Inc
- Distributor
- Worldwide Distribution-USA (nationwide) including DC and the states of AL, AR, AZ, CA, CO, CT, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, WA, WI, WV, and the countries of Dubai, Arab Emirates, Quebec, Canada, Japan, and Ireland.
- Lot
- All Lots of the Insufflation Tubing Sets are as follows: REF 28-0206 Insufflation Tubing, Laparoscopic, Lot #29074214; REF 28-0214 Hi-Flow Insufflation Tubing Lot #3194170; REF 28-0216 Tubing, Insuf w/.3 Micr Filter Lot #29074249; REF 28-0217 Insufflation Tubing, Laparoscopic Lot #32305010; DeRoyal REF 28-0208 Insufflation Tubing w/Vent Lot #29314302; DeRoyal REF 28-0212H Heated Hi-Flow Insufflation Tubing Lot #29000354.
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAREF: 81-030415 Myocardial Probe Non-Pyrogenic , 400 SeriesSeptember 30, 2026
- FDADeRoyal Industries Inc. Ref 46-406, Covaderm Plus, Vascular Access Dressing 6"x6" (15cmx15cm)June 10, 2026
- FDADeRoyal CHILDREN'S LAPTRACEPACK, 89-9362.06November 13, 2024
- FDADeRoyal C-SECTION TRACEPACK, REF 89-9338-06November 13, 2024
- FDAPrepStain Tecan US-II. For use in the screening and detection of cervical cancer.September 4, 2013
- FDAADVIA 120 Hematology Analyzer, Catalog Numbers SMN 10360958 and 10360959 . The ADVIA 120 is a hematologySeptember 4, 2013
Questions about this recall
Was DeRoyal Insufflation Tubing sets: 1) REF 28-0206 Insufflation Tubing, Laparoscopic Rx Only 2) REF 28-0214 Hi-Flow Insufflation Tubing Rx Only 3) REF 28-0216 Tubing, Insuf w/.3 Micr Filter Rx Only 4) REF 28-0217 Insufflation Tubing, Laparoscopic Rx Only 5) REF 28-0208 Insufflation Tubing w/Vent Rx Only 6) REF 28-0212H Heated Hi-Flow Insufflation Tubing Rx Only. Used during laparoscopic surgical procedures to aid in the insufflation of the abdomen. recalled?
Yes. FDA published recall Z-2113-2013 on September 4, 2013. Reported hazard: DeRoyal Industries is recalling their product Insufflation Tubing Sets due to materials incompatibility which could possibly result in occlusion at or near the filter housing which could prevent the flow of gas through the device.
What should I do if I have this product?
Voluntary: Firm initiated · Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit