Back to search
FDAFDA · Food and drugs·Recall number Z-2121-2013

Microscan Synergies Plus Negative Breakpoint Combo 7 Product Usage: MicroScan® Synergies plus® Panels & MicroScan® Rapids plus (Negative) - For use in determining quantitative and/or qualitative antimicrobial agent susceptibility and/or identification for gram-negative bacteria.

Published September 18, 2013 · Recall initiated: August 21, 2013

Summary

FDA records medical device Class I recall Z-2121-2013, published September 18, 2013 and initiated August 21, 2013. Recalling firm: Siemens Healthcare Diagnostics, Inc.; Status: Terminated; Reported quantity: 4319 boxes, 86, 380 panels; Lot: Internal Number B1025-109, Catalog Number - 10444748 Lot code/expiration date: 2012-08-17, 2012-08-17; 2012-10-06, 2012-10-06; 2012-12-08, 2012-12-08; 2013-01-09, 2013-01-09; 2013-03-02, 2013-03-02; 2013-04-12, 2013-04-12; 2013-06-15, 2013-06-15; 2013-08-06, 2013-08-06; 2013-09-24, 2013-09-24; 2013-12-07, 2013-12-07; 2014-01-08, 2014-01-08; 2014-01-31, 2014-01-31; 2014-03-25, 2014-03-25; 2014-05-09, 2014-05-09; 2014-07-01, 2014-07-01. Class I means a reasonable probability of serious adverse health consequences or death.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Confirmed increase in false susceptible and false intermediate misreads for Imipenem and/or meropenum on Synergies plus Negative and rap ID/S plus Negative panels read on Walk Away System.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Siemens Healthcare Diagnostics, Inc.
Distributor
Worldwide Distribution - USA Nationwide in the states of: AK, SC, IA, MO GA, WI, TX, KS, CA, IL, CO, AZ, SD, IN, MD, FL, MN and the countries of:Japan, France, Algeria, Germany, Austria, Romania, Russian Federation, South Africa, Kazakhstan, Belgium, Great Britain, Turkey, Spain, Italy, Saudi Arabia, Qutar, Poland, Iraq, Unite Arab Emrites, Algeria, Oman, Greece, Portugal, and Ireland.
Lot
Internal Number B1025-109, Catalog Number - 10444748 Lot code/expiration date: 2012-08-17, 2012-08-17; 2012-10-06, 2012-10-06; 2012-12-08, 2012-12-08; 2013-01-09, 2013-01-09; 2013-03-02, 2013-03-02; 2013-04-12, 2013-04-12; 2013-06-15, 2013-06-15; 2013-08-06, 2013-08-06; 2013-09-24, 2013-09-24; 2013-12-07, 2013-12-07; 2014-01-08, 2014-01-08; 2014-01-31, 2014-01-31; 2014-03-25, 2014-03-25; 2014-05-09, 2014-05-09; 2014-07-01, 2014-07-01.
Risk classification
Class I
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was Microscan Synergies Plus Negative Breakpoint Combo 7 Product Usage: MicroScan® Synergies plus® Panels & MicroScan® Rapids plus (Negative) - For use in determining quantitative and/or qualitative antimicrobial agent susceptibility and/or identification for gram-negative bacteria. recalled?

Yes. FDA published recall Z-2121-2013 on September 18, 2013. Reported hazard: Confirmed increase in false susceptible and false intermediate misreads for Imipenem and/or meropenum on Synergies plus Negative and rap ID/S plus Negative panels read on Walk Away System.

What should I do if I have this product?

Voluntary: Firm initiated · Letter