Microscan Synergies Plus Negative Urine Combo 5 Product Usage: MicroScan® Synergies plus® Panels & MicroScan® Rapids plus (Negative) - For use in determining quantitative and/or qualitative antimicrobial agent susceptibility and/or identification for gram-negative bacteria.
Published September 18, 2013 · Recall initiated: August 21, 2013
Summary
FDA records medical device Class I recall Z-2123-2013, published September 18, 2013 and initiated August 21, 2013. Recalling firm: Siemens Healthcare Diagnostics, Inc.; Status: Terminated; Reported quantity: 522 boxes, 10, 440 panels; Lot: Internal Number - B1025-115; Catalog Number - 10483101 Lot code/expiration date: 2012-09-22, 2012-09-22; 2013-01-20, 2013-01-20; 2013-05-07, 2013-05-07; 2013-11-01, 2013-11-01; 2014-03-05, 2014-03-05. Class I means a reasonable probability of serious adverse health consequences or death.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Confirmed increase in false susceptible and false intermediate misreads for Imipenem and/or meropenum on Synergies plus Negative and rap ID/S plus Negative panels read on Walk Away System.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Siemens Healthcare Diagnostics, Inc.
- Distributor
- Worldwide Distribution - USA Nationwide in the states of: AK, SC, IA, MO GA, WI, TX, KS, CA, IL, CO, AZ, SD, IN, MD, FL, MN and the countries of:Japan, France, Algeria, Germany, Austria, Romania, Russian Federation, South Africa, Kazakhstan, Belgium, Great Britain, Turkey, Spain, Italy, Saudi Arabia, Qutar, Poland, Iraq, Unite Arab Emrites, Algeria, Oman, Greece, Portugal, and Ireland.
- Lot
- Internal Number - B1025-115; Catalog Number - 10483101 Lot code/expiration date: 2012-09-22, 2012-09-22; 2013-01-20, 2013-01-20; 2013-05-07, 2013-05-07; 2013-11-01, 2013-11-01; 2014-03-05, 2014-03-05.
- Risk classification
- Class I
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAAtellica CH Revised C-Reactive Protein (RCRP); Siemens Material Number (SMN): 11537223;October 7, 2026
- FDABeckman Coulter, Emit 2000 Quinidine Assay REF: OSR4Q229 Contents: R1: 2 x 23 mL;R2 2x13 mLAugust 26, 2026
- FDASyva EMIT 2000 Quinidine Assay REF: 4Q019UL Contents: 1x28 mL, 1x14mLAugust 26, 2026
- FDASyva Emit tox Serum Tricyclic Antidepressant Calibrator/Controls SMN 10445390 Catalog No. 7C219ULAugust 12, 2026
- FDAMicroscan Synergies Plus Negative Urine Combo 2 Product Usage: MicroScan® Synergies plus® Panels &September 18, 2013
- FDAMicroscan plus Negative Combo 3 Sold outside of US only. Product Usage: MicroScan® Synergies plus® PanelsSeptember 18, 2013
Questions about this recall
Was Microscan Synergies Plus Negative Urine Combo 5 Product Usage: MicroScan® Synergies plus® Panels & MicroScan® Rapids plus (Negative) - For use in determining quantitative and/or qualitative antimicrobial agent susceptibility and/or identification for gram-negative bacteria. recalled?
Yes. FDA published recall Z-2123-2013 on September 18, 2013. Reported hazard: Confirmed increase in false susceptible and false intermediate misreads for Imipenem and/or meropenum on Synergies plus Negative and rap ID/S plus Negative panels read on Walk Away System.
What should I do if I have this product?
Voluntary: Firm initiated · Letter