IRISpec" CA/CB two-part urine chemistry control twin set, Iris Diagnostics part number 800-0074. Intended Use: IRISpec CA/CB is an assayed QC material for monitoring of the urine chemistry analytes and devices as listed on the package insert. For in vitro diagnostic (IVD) use only.

Published August 8, 2012 · Recall initiated: July 9, 2012

Summary

FDA records medical device Class III recall Z-2124-2012, published August 8, 2012 and initiated July 9, 2012. Recalling firm: Iris Diagnostics; Status: Terminated; Reported quantity: 862; Lot: Iris Diagnostics part number 800-0074, lot number 131-12. Class III means the product is not likely to cause adverse health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

The firm recalled after it received complaints of IRISpec CA/CB two part Urine Chemistry Control, 101131-12, where the nitrite results read negative for the part B control, which should be positive.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Iris Diagnostics
Distributor
Nationwide Distribution
Lot
Iris Diagnostics part number 800-0074, lot number 131-12.
Risk classification
Class III
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was IRISpec" CA/CB two-part urine chemistry control twin set, Iris Diagnostics part number 800-0074. Intended Use: IRISpec CA/CB is an assayed QC material for monitoring of the urine chemistry analytes and devices as listed on the package insert. For in vitro diagnostic (IVD) use only. recalled?

Yes. FDA published recall Z-2124-2012 on August 8, 2012. Reported hazard: The firm recalled after it received complaints of IRISpec CA/CB two part Urine Chemistry Control, 101131-12, where the nitrite results read negative for the part B control, which should be positive.

What should I do if I have this product?

Voluntary: Firm initiated · Letter