94" (239cm) 20 Drop Admin Set w/High Pressure 4-Way (300 psi) Stopcock, Rotating Luer, Non-DEHP Tubing
Published August 8, 2012 · Recall initiated: April 17, 2012
Summary
FDA records medical device Class II recall Z-2130-2012, published August 8, 2012 and initiated April 17, 2012. Recalling firm: ICU Medical, Inc.; Status: Terminated; Reported quantity: N/A; Lot: Part Number B30059 510(k) K964435 Lot number 2418897, 2423996, 2425137, 2430044, 2431861, 2436569, 2443559, 2447238, 2448099, 2452377, 2457453, 2459025, 2464569, 2467687. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
The firm initiated the recall due to a molding defect in a high pressure stopcock where a small percentage of the side ports could become totally occluded.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- ICU Medical, Inc.
- Distributor
- Nationwide distribution
- Lot
- Part Number B30059 510(k) K964435 Lot number 2418897, 2423996, 2425137, 2430044, 2431861, 2436569, 2443559, 2447238, 2448099, 2452377, 2457453, 2459025, 2464569, 2467687.
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAIV Sets with drip chambersSeptember 23, 2026
- FDAIV Sets with drip chambersSeptember 23, 2026
- FDAOncology sets with drip chambersSeptember 23, 2026
- FDAIV Non-Dedicated Gravity Administration Sets with drip chambersSeptember 23, 2026
- FDACardiac Catheterization Waste Management System w/1000mL Waste Bag, 2 Fluid Admin Sets Product Usage: ForAugust 8, 2012
- FDAFlush Kit w 3 Way Off Stopcock Product Usage: Flush KitAugust 8, 2012
Questions about this recall
Was 94" (239cm) 20 Drop Admin Set w/High Pressure 4-Way (300 psi) Stopcock, Rotating Luer, Non-DEHP Tubing recalled?
Yes. FDA published recall Z-2130-2012 on August 8, 2012. Reported hazard: The firm initiated the recall due to a molding defect in a high pressure stopcock where a small percentage of the side ports could become totally occluded.
What should I do if I have this product?
Voluntary: Firm initiated · Letter