FDAFDA · Food and drugs·Recall number Z-2131-2014

Zimmer Periarticular Locking Plate System Distal Radial Volar Plate Lateral Column (1st generation stainless steel only). Orthopedic fixation plate. For temporary internal fixation and stabilization of osteotomies and fractures.

Published August 6, 2014 · Recall initiated: June 10, 2014

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

A complaint was received in which it was reported that an 8-hole plate from lot 62276042 had an enlarged hole that is unable to lock the screw, which resulted in the screw head pulling through plate hole.

What to do

Voluntary: Firm initiated

Recall details

Manufacturer
Zimmer, Inc.
Distributor
Worldwide Distribution - US including the states of CA and OH., and the countries of BRAZIL, CHINA, KOREA and TAIWAN.
Lot
Part Number 00-2358-014-04; lot 62276042
Risk classification
Class II
Area (FDA)