Back to search
FDAFDA · Food and drugs·Recall number Z-2133-2013

CONFIDENCE SPINAL CEMENT SYSTEM 11CC PLUS KIT, Product Code: 2839-10-000 The CONFIDENCE Spinal Cement System is intended for percutaneous delivery of CONFIDENCE Spinal Cement, which is indicated for fixation of pathological fractures of the vertebral body during vertebroplasty or kyphoplasty procedures

Published September 11, 2013 · Recall initiated: April 17, 2013

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

During injection of cement, the water in the hydraulic pump leaks past the piston within the pump body resulting in the loss of pressure and inability to continue to inject cement.

What to do

Voluntary: Firm initiated · Letter

Recall details

Manufacturer
DePuy Spine, Inc.
Distributor
Worldwide Distribution - USA including AR, AZ, CA, CO, CT, FL, GA, HI, IA, ID, IN, KY, LA, MA, MD, MI, MN, MO, NC, NE, NJ, NY, OH, OK, OR, PA, SC, TN, TX, VA, WA, WV, WY; Worldwide: Algeria, Australia, Austria, Belgium, Brazil, Canada, Colombia, Cyprus, Czech Republic, France, Germany, Hong Kong, India, Israel, Italy, Latvia, Luxembourg, Malaysia, Mexico, Netherlands, Norway, Poland, Portugal, Singapore, Slovenia, South Africa, Spain, Sweden, Switzerland, Turkey, UAE, and UK
Lot
HPBB0K, HPBB3C,HPCB7J, HPDBD6, HPBB0L,HPBB3D,HPCB7W,HPDBF0, HPBB0M,HPBB3F,HPDBBC,HPDBG4, HPBB0N,HPBBWH,HPDBBD,HPDBGB, HPBB0P,HPBBWJ,HPDBBK,HPDBGC,HPBB0T,HPBBWL,HPDBBL,HPDBGJ, HPBB19,HPBBZ8,HPDBBY,HPDBGM,HPBB27,HPBBZD,HPDBCD,HPDBGN, HPBB29,HPBBZN,HPDBCF, HPDBGY,HPBB2B,HPCB7B,HPDBD0,HPBB2C, HPCB7C,HPDBD3,HPBB2G,HPCB7H,HPDBD4
Risk classification
Class II
Area (FDA)
device
Country
United States

Sell products like this?

Recalert watches your whole catalog for recalls like this one, automatically.

Start free — no card