FDAFDA · Food and drugs·Recall number Z-2133-2013
CONFIDENCE SPINAL CEMENT SYSTEM 11CC PLUS KIT, Product Code: 2839-10-000 The CONFIDENCE Spinal Cement System is intended for percutaneous delivery of CONFIDENCE Spinal Cement, which is indicated for fixation of pathological fractures of the vertebral body during vertebroplasty or kyphoplasty procedures
Published September 11, 2013 · Recall initiated: April 17, 2013
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
During injection of cement, the water in the hydraulic pump leaks past the piston within the pump body resulting in the loss of pressure and inability to continue to inject cement.
What to do
Voluntary: Firm initiated · Letter
Recall details
- Manufacturer
- DePuy Spine, Inc.
- Distributor
- Worldwide Distribution - USA including AR, AZ, CA, CO, CT, FL, GA, HI, IA, ID, IN, KY, LA, MA, MD, MI, MN, MO, NC, NE, NJ, NY, OH, OK, OR, PA, SC, TN, TX, VA, WA, WV, WY; Worldwide: Algeria, Australia, Austria, Belgium, Brazil, Canada, Colombia, Cyprus, Czech Republic, France, Germany, Hong Kong, India, Israel, Italy, Latvia, Luxembourg, Malaysia, Mexico, Netherlands, Norway, Poland, Portugal, Singapore, Slovenia, South Africa, Spain, Sweden, Switzerland, Turkey, UAE, and UK
- Lot
- HPBB0K, HPBB3C,HPCB7J, HPDBD6, HPBB0L,HPBB3D,HPCB7W,HPDBF0, HPBB0M,HPBB3F,HPDBBC,HPDBG4, HPBB0N,HPBBWH,HPDBBD,HPDBGB, HPBB0P,HPBBWJ,HPDBBK,HPDBGC,HPBB0T,HPBBWL,HPDBBL,HPDBGJ, HPBB19,HPBBZ8,HPDBBY,HPDBGM,HPBB27,HPBBZD,HPDBCD,HPDBGN, HPBB29,HPBBZN,HPDBCF, HPDBGY,HPBB2B,HPCB7B,HPDBD0,HPBB2C, HPCB7C,HPDBD3,HPBB2G,HPCB7H,HPDBD4
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States