Dimension® Integrated Chemistry Systems intended to measure a variety of analytes in human body fluids.

Published August 6, 2014 · Recall initiated: May 27, 2014

Summary

FDA records medical device Class III recall Z-2133-2014, published August 6, 2014 and initiated May 27, 2014. Recalling firm: Siemens Healthcare Diagnostics, Inc.; Status: Terminated; Reported quantity: 15; Lot: DE271317, DE271320, DE271323, DE271324, DE271326, DE271327, DE271328, DE271332 DE271341, DE271343, DE271325, 12252589, 2004082898, 2004082899, 224719-AX. Class III means the product is not likely to cause adverse health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

There is a misassembly issue with the photometer installed on Dimension® Integrated Chemistry Systems instruments. Through internal investigation by Siemens Healthcare Diagnostics has confirmed that a limited number of Dimension photometers were misassembled. These photometers are performing within specifications, but may lead to a shortened life of the source lamp.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Siemens Healthcare Diagnostics, Inc.
Distributor
FL LA AR IA OH OH WI IL LA KS MT
Lot
DE271317, DE271320, DE271323, DE271324, DE271326, DE271327, DE271328, DE271332 DE271341, DE271343, DE271325, 12252589, 2004082898, 2004082899, 224719-AX
Risk classification
Class III
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was Dimension® Integrated Chemistry Systems intended to measure a variety of analytes in human body fluids. recalled?

Yes. FDA published recall Z-2133-2014 on August 6, 2014. Reported hazard: There is a misassembly issue with the photometer installed on Dimension® Integrated Chemistry Systems instruments. Through internal investigation by Siemens Healthcare Diagnostics has confirmed that a limited number of Dimension photometers were misassembled. These photometers are performing within specifications, but may lead to a shortened life of the source lamp.