Back to search
FDAFDA · Food and drugs·Recall number Z-2133-2019

Nasal Alar SpO2 Sensor, Adult/Pediatric SpO2 Sensor, Rx Only Product Usage: The Assurance¿ Nasal Alar SpO2 Sensor is indicated for single patient use for continuous noninvasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate from the nasal ala of adult and pediatric patients, (at least 4 years and older and weighing > 15kg). The sensor can be used in a variety of healthcare environments where compatible pulse oximetry monitors are indicated for use, under professional supervision

Published August 7, 2019 · Recall initiated: May 8, 2019

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

These Instructions for Use has missing instructions related to the checking and changing of the application site procedure.

What to do

Voluntary: Firm initiated · Letter

Recall details

Manufacturer
Xhale Assurance, Inc.
Distributor
US Nationwide distribution to AZ, CA, FL, KS, LA, MA , MO, MS, NC, NY, OH, PA, TX, WI. International distribution to Canada, Germany, Malaysia, United Kingdom.
Lot
Reference Codes (Lot Numbers): 0303 (101619, 101624, 101628, 101632, 101637, 101642, 101646, 101647, 101651, 101701, 201651, 101707, 101715, XH101711, XH101720, XH101715, XH101726, XH101729, XH101733, XH101737, XH101820, XH101842), 301-11214 (TR101711, TR101715, TR101720, TR101726, TR101729, TR101744, TR101733, TR101737, TR101748, TR101742, TR101746, TR101750, TR101803, TR101811, TR101817, TR101820, TR101824, TR101829, TR101833, TR101838, TR101842, TR101847, TR101852), 0201-A01(PF101736, PF101806, PF101814, PF101817, PF101819, PF101822, PF101828, PF101833, PF101838). Instructions for Use for revisions 10412_7 and prior (i.e. 10412_6, 10412_5, etc) as well as 10358_6 and prior. Shelf life is 3 years for sensors manufactured prior to Oct 2018. 5 years for sensor manufactured on or after Oct 2018.
Risk classification
Class II
Area (FDA)
device
Country
United States