Back to search
FDAFDA · Food and drugs·Recall number Z-2138-2012

TANGO3 Water Storage Tank with Ozone Disinfection System, Tango3, LLC, 132 Citizens Blvd., Simpsonville, KY 40067. Models are WOZ-250-PE, WOZ-240-SS, WOZ-180-SS. The TANGO3 Water Storage Tank with Ozone Disinfection System is intended to be used for disinfection of the water distribution system of a dialysis facility.

Published August 15, 2012 · Recall initiated: March 26, 2012

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

During an FDA inspection, the firm was notified of Quality System deficiencies, which included lack of procedures for design control and customized modifications that were either not reviewed or received proper clearance from FDA. There was no assurance that all systems in the field were working properly and operating at a minimum of 0.2ppm of ozone.

What to do

Voluntary: Firm initiated · Letter

Recall details

Manufacturer
Tango3 LLC
Distributor
Product was shipped to the following states: CA, CO, CT, DE, FL, GA, IA, ID, IL, IN, KS, KY, MA, MD, ME, NH, NJ, NM, OH, PA, RI, TX and VA.
Lot
Model WOZ -250-PE, Serial #s: 1031, 1032, 1033, 1035, 1036, 1038, 1039, 1042, 1043, 1044, 1045, 1046, 1047, 1048, 1049, 1050, 1051, 1053, 1054, 1055, 1056, 1057, 1058, 1059, 1060, 1061, 1062, 1063, 1065, 1066, 1067, 1068, 1069, 1070, 1071, 1072, 1073, 1074, 1075, 1076 & 1077. Model WOZ-180-SS, Serial #s: 1034, 1037, 1040 & 1064. Model WOZ-240-SS, Serial #s: 1041 & 1052.
Risk classification
Class II
Area (FDA)
device
Country
United States