Back to search
FDAFDA · Food and drugs·Recall number Z-2146-2017

Jackson-Pratt Silicone Round Drain, 10FR. Sterile, Rx Only. For Single Use Only. Jackson-Pratt Wound Drainage Systems are sterile single use devices. The wound drainage system consist of a wound drain and a fluid collection reservoir. The wound drain is a radio-opaque section of tubing which has perforations or ducts. The wound drain is surgically placed in a surgical wound site using a trocar. Blood and fluids are collected through a wound drain into a fluid collection reservoir. Wound drains are used to remove exudates from wound sites. Trocars, drains and reservoirs are for a single patient use only and should be discarded after use

Published May 31, 2017 · Recall initiated: April 21, 2017

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Product's seal possibly compromised, potentially compromising the sterility of the package contents. Use of impacted products could result in an increased risk of infection.

What to do

Voluntary: Firm initiated · Letter

Recall details

Manufacturer
Cardinal Health 200, LLC
Distributor
Nationwide Distribution
Lot
Catalog Number: SU130-1321 Lot Numbers: 1161206 1161207 1161370 1161371 1161372 1161476 1161477 1161478 1161606 1161607 1161608 1170036
Risk classification
Class II
Area (FDA)
device
Country
United States