Bard Medical SureStep Foley Tray System, 16 Fr., Catalog #A319416AM; and Bard Medical Bardex IC Complete Care Infection Control Temperature-Sensing Foley Catheter Tray with Bard Hydrogel and Bacti-Guard Silver Alloy Coating, 350 ml Urine Meter, 16 Fr, Catalog #319416AM. Intended for use in the drainage and/or collection and/or measurement of urine.
Published June 26, 2024 · Recall initiated: May 9, 2024
Summary
FDA records medical device Class II recall Z-2154-2024, published June 26, 2024 and initiated May 9, 2024. Recalling firm: C.R. Bard Inc; Status: Ongoing; Reported quantity: 6,420 devices; Lot: SureStep Foley Tray System - Lot #NGHZ1757, exp. 31Jan2026, UDI-DI 00801741073861 (eaches); 10801741073868 (case). Bardex IC Complete Care Infection Control Temperature-Sensing Foley Catheter Tray, Lot #NGHX2932, exp. 31May2025, UDI-DI 00801741024450 (each); 10801741024457 (case). Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Some of the devices may function incorrectly resulting in no temperature display or erratic temperatures being displayed on the temperature monitoring devices.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- C.R. Bard Inc
- Distributor
- Distribution was nationwide. There was no government/military/foreign distribution.
- Lot
- SureStep Foley Tray System - Lot #NGHZ1757, exp. 31Jan2026, UDI-DI 00801741073861 (eaches); 10801741073868 (case). Bardex IC Complete Care Infection Control Temperature-Sensing Foley Catheter Tray, Lot #NGHX2932, exp. 31May2025, UDI-DI 00801741024450 (each); 10801741024457 (case).
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAProducts that contain the Pericare Wipes: *Bardex® I.C. Complete Care® Foley Catheter Tray REF: 300316A UDISeptember 16, 2026
- FDAProducts that contain the Pericare Wipes: Bardex Lubricath Foley Catheter Tray REF: 899914 899916 899918 UDISeptember 16, 2026
- FDAProducts that contain the Pericare Wipes: Bard® Foley Tray With Lubri-Sil® All-Silicone Foley Catheter REF:September 16, 2026
- FDAProducts that contain the Pericare Wipes: SureStep Intermittent Catheter Tray REF: INTP16 INTS14 INTS14CSeptember 16, 2026
- FDAGray Revision Instrument System Accessory Tray/Case, Catalog/Part Number: 6210-9-910June 26, 2024
- FDAOPTETRAK Three Peg Patella, Item Numbers: a) 200-02-26, 26MM; b) 200-02-29, 29MM; c) 200-02-32, 32MM; d)June 26, 2024
Questions about this recall
Was Bard Medical SureStep Foley Tray System, 16 Fr., Catalog #A319416AM; and Bard Medical Bardex IC Complete Care Infection Control Temperature-Sensing Foley Catheter Tray with Bard Hydrogel and Bacti-Guard Silver Alloy Coating, 350 ml Urine Meter, 16 Fr, Catalog #319416AM. Intended for use in the drainage and/or collection and/or measurement of urine. recalled?
Yes. FDA published recall Z-2154-2024 on June 26, 2024. Reported hazard: Some of the devices may function incorrectly resulting in no temperature display or erratic temperatures being displayed on the temperature monitoring devices.
What should I do if I have this product?
Voluntary: Firm initiated · Letter