Back to search
FDAFDA · Food and drugs·Recall number Z-2155-2017

CytoGuard Closed Luer Connector, Intravenous access 2 cartons of 50 units each (100 units per case)

Published May 31, 2017 · Recall initiated: April 10, 2017

Summary

FDA records medical device Class II recall Z-2155-2017, published May 31, 2017 and initiated April 10, 2017. Recalling firm: B Braun Medical Inc; Status: Terminated; Reported quantity: 417,600 units; Lot: Lot # 1240156001, 1240156002, 1240156003, 1240156004, 1370249001, 1370251101, 1370257101, 1370259201, 1370268401, 1370270401, 1370273801, 1470311101, and 1570315301. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

B. Braun is voluntarily recalling 13 lots of CytoGuard Closed Luer Connector due to a potential for some blisters to be punctured resulting in a compromised sterility barrier.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
B Braun Medical Inc
Distributor
United States Nationwide distribution
Lot
Lot # 1240156001, 1240156002, 1240156003, 1240156004, 1370249001, 1370251101, 1370257101, 1370259201, 1370268401, 1370270401, 1370273801, 1470311101, and 1570315301
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was CytoGuard Closed Luer Connector, Intravenous access 2 cartons of 50 units each (100 units per case) recalled?

Yes. FDA published recall Z-2155-2017 on May 31, 2017. Reported hazard: B. Braun is voluntarily recalling 13 lots of CytoGuard Closed Luer Connector due to a potential for some blisters to be punctured resulting in a compromised sterility barrier.

What should I do if I have this product?

Voluntary: Firm initiated · Letter