CytoGuard Closed Luer Connector, Intravenous access 2 cartons of 50 units each (100 units per case)
Published May 31, 2017 · Recall initiated: April 10, 2017
Summary
FDA records medical device Class II recall Z-2155-2017, published May 31, 2017 and initiated April 10, 2017. Recalling firm: B Braun Medical Inc; Status: Terminated; Reported quantity: 417,600 units; Lot: Lot # 1240156001, 1240156002, 1240156003, 1240156004, 1370249001, 1370251101, 1370257101, 1370259201, 1370268401, 1370270401, 1370273801, 1470311101, and 1570315301. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
B. Braun is voluntarily recalling 13 lots of CytoGuard Closed Luer Connector due to a potential for some blisters to be punctured resulting in a compromised sterility barrier.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- B Braun Medical Inc
- Distributor
- United States Nationwide distribution
- Lot
- Lot # 1240156001, 1240156002, 1240156003, 1240156004, 1370249001, 1370251101, 1370257101, 1370259201, 1370268401, 1370270401, 1370273801, 1470311101, and 1570315301
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAB Braun SAFSITE Injection Site, RV1000NC SAFSITE VALVE W/O EXTRA CAP. Catalog Number: 415068.October 28, 2026
- FDABrand Name: Extension Set Product Name: DISCOFIX FOUR WAY STOPCOCK Model/Catalog Number: 456025September 9, 2026
- FDABrand Name: B. BRAUN MEDICAL INC. Product Name: CESK RIVERSIDE HEALTH SYSTEM Model/Catalog Number: 552126June 10, 2026
- FDABrand Name: B. BRAUN MEDICAL INC. Product Name: NES1725KFXL SPINAL EPID CLOSED TIP CATH Model/Catalog Number:June 10, 2026
- FDAOpteryx Variable Rescue Screw, Part #CP2214, UDI: *+M711CP22140/$$801716012*. Packaged in heat-sealed plasticMay 31, 2017
- FDAAXIOM Sensis, Computer, Diagnostic, Programmable. Intended to be used as a diagnostic and administrative toolMay 31, 2017
Questions about this recall
Was CytoGuard Closed Luer Connector, Intravenous access 2 cartons of 50 units each (100 units per case) recalled?
Yes. FDA published recall Z-2155-2017 on May 31, 2017. Reported hazard: B. Braun is voluntarily recalling 13 lots of CytoGuard Closed Luer Connector due to a potential for some blisters to be punctured resulting in a compromised sterility barrier.
What should I do if I have this product?
Voluntary: Firm initiated · Letter