FDAFDA · Food and drugs·Recall number Z-2160-2026

RelayPro Thoracic Stent-Graft System (various sizes) Reference numbers: 28-N4-32-104-32U 28-N4-32-164-28U 28-N4-32-164-32U 28-N4-32-209-28U 28-N4-32-209-32U 28-N4-32-259-32U 28-N4-34-109-34U 28-N4-34-154-30U 28-N4-34-154-34U 28-N4-34-209-30U 28-N4-34-209-34U 28-N4-34-259-34U 28-N4-36-109-36U 28-N4-36-154-32U 28-N4-36-154-36U 28-N4-36-199-32U 28-N4-36-199-36U 28-N4-36-259-36U 28-N4-38-109-38U 28-N4-38-154-34U 28-N4-38-154-38U 28-N4-38-199-34U 28-N4-38-199-38U 28-N4-38-259-38U 28-N4-40-114-40U 28-N4-40-154-36U 28-N4-40-154-40U 28-N4-40-204-36U 28-N4-40-204-40U 28-N4-40-259-40U 28-N4-42-114-42U 28-N4-42-159-38U 28-N4-42-159-42U 28-N4-42-204-38U 28-N4-42-204-42U 28-N4-42-259-42U 28-N4-44-114-44U 28-N4-44-164-40U 28-N4-44-164-44U 28-N4-44-209-40U 28-N4-44-209-44U 28-N4-44-259-40U 28-N4-44-259-44U 28-N4-46-114-46U 28-N4-46-164-42U 28-N4-46-164-46U 28-N4-46-209-42U 28-N4-46-209-46U 28-N4-46-259-42U

Published June 3, 2026 · Recall initiated: April 22, 2026

Summary

FDA records medical device Class I recall Z-2160-2026, published June 3, 2026 and initiated April 22, 2026. Recalling firm: Bolton Medical Inc.; Status: Ongoing; Lot: All Lots/(Ref Number/Proximal Diameter/DI)28-N4-32-104-32U/ 32/ (01)00843576150461; 28-N4-32-164-28U/ 32/ (01)00843576150959; 28-N4-32-164-32U/ 32/ (01)00843576150591; 28-N4-32-209-28U/ 32/ (01)00843576151055; 28-N4-32-209-32U/ 32/ (01)00843576150720; 28-N4-32-259-32U/ 32/ (01)00843576150850; 28-N4-34-109-34U/ 34/ (01)00843576150478; 28-N4-34-154-30U/ 34/ (01)00843576150966; 28-N4-34-154-34U/ 34/ (01)00843576150607; 28-N4-34-209-30U/ 34/ (01)00843576151062; 28-N4-34-209-34U/ 34/ (01)00843576150737; 28-N4-34-259-34U/ 34/ (01)00843576150867; 28-N4-36-109-36U/ 36/ (01)00843576150485; 28-N4-36-154-32U/ 36/ (01)00843576150973; 28-N4-36-154-36U/ 36/ (01)00843576150614; 28-N4-36-199-32U/ 36/ (01)00843576151079; 28-N4-36-199-36U/ 36/ (01)00843576150744; 28-N4-36-259-36U/ 36/ (01)00843576150874; 28-N4-38-109-38U/ 38/ (01)00843576150492; 28-N4-38-154-34U/ 38/ (01)00843576150980; 28-N4-38-154-38U/ 38/ (01)00843576150621; 28-N4-38-199-34U/ 38/ (01)00843576151086; 28-N4-38-199-38U/ 38/ (01)00843576150751; 28-N4-38-259-38U/ 38/ (01)00843576150881; 28-N4-40-114-40U/ 40/ (01)00843576150508; 28-N4-40-154-36U/ 40/ (01)00843576150997; 28-N4-40-154-40U/ 40/ (01)00843576150638; 28-N4-40-204-36U/ 40/ (01)00843576151093; 28-N4-40-204-40U/ 40/ (01)00843576150768; 28-N4-40-259-40U/ 40/ (01)00843576150898; 28-N4-42-114-42U/ 42/ (01)00843576150515; 28-N4-42-159-38U/ 42/ (01)00843576151000; 28-N4-42-159-42U/ 42/ (01)00843576150645; 28-N4-42-204-38U/ 42/ (01)00843576151109; 28-N4-42-204-42U/ 42/ (01)00843576150775; 28-N4-42-259-42U/ 42/ (01)00843576150904; 28-N4-44-114-44U/ 44/ (01)00843576150522; 28-N4-44-164-40U/ 44/ (01)00843576151017; 28-N4-44-164-44U/ 44/ (01)00843576150652; 28-N4-44-209-40U/ 44/ (01)00843576151116; 28-N4-44-209-44U/ 44/ (01)00843576150782; 28-N4-44-259-40U/ 44/ (01)00843576151215; 28-N4-44-259-44U/ 44/ (01)00843576150911; 28-N4-46-114-46U/ 46/ (01)00843576150539; 28-N4-46-164-42U/ 46/ (01)00843576151024; 28-N4-46-164-46U/ 46/ (01)00843576150669; 28-N4-46-209-42U/ 46/ (01)00843576151123; 28-N4-46-209-46U/ 46/ (01)00843576150799; 28-N4-46-259-42U/ 46/ (01)00843576151222. Class I means a reasonable probability of serious adverse health consequences or death.

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Labeling update concerning proximal clasp disconnection from the outer control tube which may prevent stent graft from being released from the delivery system.

What to do

Voluntary: Firm initiated

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Bolton Medical Inc.
Distributor
Worldwide - distribution in the US Territory of Puerto Rico and the countries of Argentina, Chile, Canada, Colombia, Mexico, United Kingdom.
Lot
All Lots/(Ref Number/Proximal Diameter/DI)28-N4-32-104-32U/ 32/ (01)00843576150461; 28-N4-32-164-28U/ 32/ (01)00843576150959; 28-N4-32-164-32U/ 32/ (01)00843576150591; 28-N4-32-209-28U/ 32/ (01)00843576151055; 28-N4-32-209-32U/ 32/ (01)00843576150720; 28-N4-32-259-32U/ 32/ (01)00843576150850; 28-N4-34-109-34U/ 34/ (01)00843576150478; 28-N4-34-154-30U/ 34/ (01)00843576150966; 28-N4-34-154-34U/ 34/ (01)00843576150607; 28-N4-34-209-30U/ 34/ (01)00843576151062; 28-N4-34-209-34U/ 34/ (01)00843576150737; 28-N4-34-259-34U/ 34/ (01)00843576150867; 28-N4-36-109-36U/ 36/ (01)00843576150485; 28-N4-36-154-32U/ 36/ (01)00843576150973; 28-N4-36-154-36U/ 36/ (01)00843576150614; 28-N4-36-199-32U/ 36/ (01)00843576151079; 28-N4-36-199-36U/ 36/ (01)00843576150744; 28-N4-36-259-36U/ 36/ (01)00843576150874; 28-N4-38-109-38U/ 38/ (01)00843576150492; 28-N4-38-154-34U/ 38/ (01)00843576150980; 28-N4-38-154-38U/ 38/ (01)00843576150621; 28-N4-38-199-34U/ 38/ (01)00843576151086; 28-N4-38-199-38U/ 38/ (01)00843576150751; 28-N4-38-259-38U/ 38/ (01)00843576150881; 28-N4-40-114-40U/ 40/ (01)00843576150508; 28-N4-40-154-36U/ 40/ (01)00843576150997; 28-N4-40-154-40U/ 40/ (01)00843576150638; 28-N4-40-204-36U/ 40/ (01)00843576151093; 28-N4-40-204-40U/ 40/ (01)00843576150768; 28-N4-40-259-40U/ 40/ (01)00843576150898; 28-N4-42-114-42U/ 42/ (01)00843576150515; 28-N4-42-159-38U/ 42/ (01)00843576151000; 28-N4-42-159-42U/ 42/ (01)00843576150645; 28-N4-42-204-38U/ 42/ (01)00843576151109; 28-N4-42-204-42U/ 42/ (01)00843576150775; 28-N4-42-259-42U/ 42/ (01)00843576150904; 28-N4-44-114-44U/ 44/ (01)00843576150522; 28-N4-44-164-40U/ 44/ (01)00843576151017; 28-N4-44-164-44U/ 44/ (01)00843576150652; 28-N4-44-209-40U/ 44/ (01)00843576151116; 28-N4-44-209-44U/ 44/ (01)00843576150782; 28-N4-44-259-40U/ 44/ (01)00843576151215; 28-N4-44-259-44U/ 44/ (01)00843576150911; 28-N4-46-114-46U/ 46/ (01)00843576150539; 28-N4-46-164-42U/ 46/ (01)00843576151024; 28-N4-46-164-46U/ 46/ (01)00843576150669; 28-N4-46-209-42U/ 46/ (01)00843576151123; 28-N4-46-209-46U/ 46/ (01)00843576150799; 28-N4-46-259-42U/ 46/ (01)00843576151222
Risk classification
Class I
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was RelayPro Thoracic Stent-Graft System (various sizes) Reference numbers: 28-N4-32-104-32U 28-N4-32-164-28U 28-N4-32-164-32U 28-N4-32-209-28U 28-N4-32-209-32U 28-N4-32-259-32U 28-N4-34-109-34U 28-N4-34-154-30U 28-N4-34-154-34U 28-N4-34-209-30U 28-N4-34-209-34U 28-N4-34-259-34U 28-N4-36-109-36U 28-N4-36-154-32U 28-N4-36-154-36U 28-N4-36-199-32U 28-N4-36-199-36U 28-N4-36-259-36U 28-N4-38-109-38U 28-N4-38-154-34U 28-N4-38-154-38U 28-N4-38-199-34U 28-N4-38-199-38U 28-N4-38-259-38U 28-N4-40-114-40U 28-N4-40-154-36U 28-N4-40-154-40U 28-N4-40-204-36U 28-N4-40-204-40U 28-N4-40-259-40U 28-N4-42-114-42U 28-N4-42-159-38U 28-N4-42-159-42U 28-N4-42-204-38U 28-N4-42-204-42U 28-N4-42-259-42U 28-N4-44-114-44U 28-N4-44-164-40U 28-N4-44-164-44U 28-N4-44-209-40U 28-N4-44-209-44U 28-N4-44-259-40U 28-N4-44-259-44U 28-N4-46-114-46U 28-N4-46-164-42U 28-N4-46-164-46U 28-N4-46-209-42U 28-N4-46-209-46U 28-N4-46-259-42U recalled?

Yes. FDA published recall Z-2160-2026 on June 3, 2026. Reported hazard: Labeling update concerning proximal clasp disconnection from the outer control tube which may prevent stent graft from being released from the delivery system.