CORFLO PEG Kit, Pull 16FR. Product Usage: CORFLO PEG Tubes are intended for delivery of enteral nutrition, water and medications into the stomach. The PEG adapter is the connection point between the tubing and the feeding set/syringe. The adapter is designed as a replaceable component.

Published May 31, 2017 · Recall initiated: February 3, 2017

Summary

FDA records medical device Class II recall Z-2163-2017, published May 31, 2017 and initiated February 3, 2017. Recalling firm: CORPAK MedSystems, Inc.; Status: Terminated; Reported quantity: 8 cases (2 units per case); Lot: Catalogue Numbers: 50-6016. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Firm is initiating voluntary correction of CORFLO PEG with ENFit. Barb connector (stem) of the adapter may break unexpectedly when exposed to stress.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
CORPAK MedSystems, Inc.
Distributor
Worldwide distribution in the states of NH, VT, CA and the countries of UK, Ireland, France, Austria, Norway, Sweden, New Zealand
Lot
Catalogue Numbers: 50-6016
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was CORFLO PEG Kit, Pull 16FR. Product Usage: CORFLO PEG Tubes are intended for delivery of enteral nutrition, water and medications into the stomach. The PEG adapter is the connection point between the tubing and the feeding set/syringe. The adapter is designed as a replaceable component. recalled?

Yes. FDA published recall Z-2163-2017 on May 31, 2017. Reported hazard: Firm is initiating voluntary correction of CORFLO PEG with ENFit. Barb connector (stem) of the adapter may break unexpectedly when exposed to stress.

What should I do if I have this product?

Voluntary: Firm initiated · Letter