CARESCAPE Canvas Smart Display, Model Numbers: a) 5700023, b) 5700023-02, c) 5939590-102; monitor, physiological, patient
Published July 3, 2024 · Recall initiated: April 5, 2024
Summary
FDA records medical device Class II recall Z-2169-2024, published July 3, 2024 and initiated April 5, 2024. Recalling firm: GE Healthcare Finland Oy; Status: Ongoing; Reported quantity: 4649 units; Lot: GTIN 00195278522313: a) 5700023, Serial Numbers: 2534526-3-1, 2553213-5-1, 2606854-2-1, 5220256-1-1, 5215673-1-1, 5215634-1-1, 5264236-1-1, 5264113-1-1, 5245953-1-1, 5264089-1-1, 5275734-1-1, 5301535-1-1, 5300969-1-1, 5341313-1-1, 5395275-1-1, 5232613-1-1, 5232595-1-1, 5287222-1-1, 5390098-3-1, 5316614-6-1, 5316614-16-1, 5316614-22-1, 2615004-27-1, 5368586-1-1, 5223040-1-1, 5223040-1-2, 5223040-1-3, 5223040-1-4, 5274472-12-1, 5274472-71-1, 5350377-1-1, 5215675-1-1, 5223119-1-1, 5231395-1-1, 5245547-1-1, 5307272-2-1, 5318051-3-1, 5318978-2-1, 5317879-3-1, 5331606-1-1, 5338491-2-1, 5356848-4-1, 5356848-8-1, 5356848-7-1, 5356848-10-1, 5365520-6-1, 5373002-3-1, 5356866-18-1, 5356866-10-1, 5356866-26-1, 5356866-6-1, 5356866-2-1, 5374179-2-1, 2495499-1-1, 2495362-1-1, 2514941-1-1, 2522522-1-1, 2525199-1-1, 2520809-1-1, 2520809-1-2, 2526133-1-1, 2503865-2-1, 2514269-1-1, 2532289-1-1, 2534398-6-1, 2543443-1-1, 2538968-1-1, 2543938-15-1, 2547020-1-1, 2556233-14-1, 2556233-13-1, 2607079-5-1, 2556233-18-1, 2494479-1-1, 2494484-1-1, 2494487-1-1, 2515199-1-1, 2494485-1-1, 2494483-1-1, 2512827-1-1, 2512885-1-1, 2513077-1-1, 2518278-4-1, 2531556-1-1, 2525641-2-1, 2554943-1-1, 2551434-1-1, 2582151-1-1, 2585668-6-1, 2611315-1-1, 2611294-1-1, 2564890-3-1, 2619057-1-1, 2617112-1-1, 2618012-4-1, 2494648-5-1, 2500341-1-1, 2500318-2-1, 2509366-1-1, 2509367-1-1, 2512604-1-1, 2512604-1-2, 2514892-1-1, 2513939-2-1, 2516853-1-1, 2516851-1-1, 2523820-1-1, 2523643-2-1, 2524982-5-1, 2524982-1-1, 2537780-1-1, 2537780-1-2, 2522678-7-1, 2556763-18-1, 2556763-2-1, 2556763-18-2, 2562037-1-1, 2555090-2-1, 2576242-2-1, 2576259-2-1, 2554602-1-1, 2560382-1-1, 2560381-1-1, 2586284-1-1, 2612355-1-1, 2610438-1-1, 2611840-3-1, 2611351-2-1, 2611541-1-1, 2612895-4-1, 2604783-2-1, 2618900-1-1, 2618900-1-2, 2622551-1-1, 2623858-3-1, 2623105-1-1, 2623105-1-2, 2624426-2-1, 2626536-1-1, 2494497-1-1, 2494500-1-1, 2494525-1-1, 2503234-1-1, 2494526-1-1, 2495361-1-1, 2496099-1-1, 2516274-1-1, 2514690-1-1, 2495543-1-1, 2495400-1-1, 2495559-1-1, 2495409-1-1, 2519427-1-1, 2519427-1-2, 2519427-1-3, 2503764-1-1, 2522006-1-1, 2513114-3-1, 2514141-1-1, 2521852-1-1, 2512685-1-1, 2521540-1-1, 2522959-1-1, 2522761-1-1, 2519916-11-1, 2529981-1-1, 2529405-12-1, 2529406-11-1, 2529982-1-1, 2529409-9-1, 2529409-6-1, 2529434-1-1, 2543934-7-1, 2543933-7-1, 2543932-1-1, 2539102-7-1, 2539094-7-1, 2539436-1-1, 2540128-1-1, 2539084-8-1, 2539437-1-1, 2539438-1-1, 2528671-18-1, 2562722-1-1, 2568082-1-1, 2554514-1-1, 2547282-20-1, 2547282-6-1, 2547282-10-1, 2547282-8-1, 2547282-4-1, 2547231-10-1, 2547446-15-1, 2523479-16-1, 2552470-1-1, 2580467-1-1, 2561965-1-1, 2582525-6-1, 2582525-32-1, 2585579-2-1, 2592273-1-1, 2599020-1-1, 2579706-5-1, 2574524-13-1, 2562789-9-1, 2562789-2-1, 2562789-9-2, 2562789-5-1, 2562789-5-2, 2562789-4-1, 2562789-2-3, 2562789-37-1, 2562789-18-1, 2562789-14-1, 2562789-20-1, 2579723-1-1, 2580465-2-1, 2567049-24-1, 2580465-12-1, 2554559-19-1, 2604759-1-1, 2606427-1-1, 2575662-15-1, 2582438-1-1, 2612551-1-1, 2574524-13-2, 2611203-1-1, 2612612-10-1, 2611747-1-1, 2606474-1-1, 2611016-1-1, 2612734-18-1, 2612734-2-1, 2613185-15-1, 2613083-1-1, 2615930-1-1, 2613111-1-1, 2613111-20-1, 2618898-2-1, 2617993-2-1, 2617993-1-1, 2615926-1-1, 2616599-1-1, 2618691-30-1, 2616077-1-1, 2618690-20-1, 2621975-1-1, 2618903-10-1, 2622941-1-1, 2619537-1-1, 2621988-1-1, 2622940-1-1, 2620603-1-1, 2619182-2-1, 2625896-1-1, 2624865-1-1, 2624047-1-1, 2626740-2-1, 2624450-1-1, 2624725-18-1, 2624725-13-1, 2624725-5-1, 2624725-1-1, 2624725-13-2, 2577660-1-1, 2517501-1-1, 2545870-1-1, 2615937-3-1, 2616596-2-1, 2619984-2-1, 2616737-1-1, 2618481-5-1, 2499911-1-1, 2505712-1-1, 2503687-1-1, 2554926-1-1, 2572230-2-1, 2588785-1-1, 2495298-1-1, 2495264-1-1, 2509179-2-1, 2495358-1-1, 2515181-2-1, 2515181-1-1, 2513165-1-1, 2495299-1-1, 2531550-1-1, 2544853-2-1, 2574647-2-1, 2598884-1-1, 2618137-1-1, 2628573-1-1, 2494372-1-1, 2518561-9-1, 2522026-1-1, 2522834-1-1, 2524986-1-1, 2524986-7-1, 2524898-1-1, 2522845-1-1, 252. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
GE HealthCare has become aware of the potential for loss of monitoring on CARESCAPE Canvas 1000, CARESCAPE Canvas Smart Display, and certain versions of CARESCAPE B850 and CARESCAPE B650 monitors if, following the replacement of the CPU timekeeper battery, a mains power loss occurs during active monitoring. In the unlikely event that this occurs, it can result in the displays or monitors not turning on, after the mains power is restored. This can result in a delay in recognizing changes in patient condition until a replacement monitor is connected.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- GE Healthcare Finland Oy
- Distributor
- Worldwide distribution.
- Lot
- GTIN 00195278522313: a) 5700023, Serial Numbers: 2534526-3-1, 2553213-5-1, 2606854-2-1, 5220256-1-1, 5215673-1-1, 5215634-1-1, 5264236-1-1, 5264113-1-1, 5245953-1-1, 5264089-1-1, 5275734-1-1, 5301535-1-1, 5300969-1-1, 5341313-1-1, 5395275-1-1, 5232613-1-1, 5232595-1-1, 5287222-1-1, 5390098-3-1, 5316614-6-1, 5316614-16-1, 5316614-22-1, 2615004-27-1, 5368586-1-1, 5223040-1-1, 5223040-1-2, 5223040-1-3, 5223040-1-4, 5274472-12-1, 5274472-71-1, 5350377-1-1, 5215675-1-1, 5223119-1-1, 5231395-1-1, 5245547-1-1, 5307272-2-1, 5318051-3-1, 5318978-2-1, 5317879-3-1, 5331606-1-1, 5338491-2-1, 5356848-4-1, 5356848-8-1, 5356848-7-1, 5356848-10-1, 5365520-6-1, 5373002-3-1, 5356866-18-1, 5356866-10-1, 5356866-26-1, 5356866-6-1, 5356866-2-1, 5374179-2-1, 2495499-1-1, 2495362-1-1, 2514941-1-1, 2522522-1-1, 2525199-1-1, 2520809-1-1, 2520809-1-2, 2526133-1-1, 2503865-2-1, 2514269-1-1, 2532289-1-1, 2534398-6-1, 2543443-1-1, 2538968-1-1, 2543938-15-1, 2547020-1-1, 2556233-14-1, 2556233-13-1, 2607079-5-1, 2556233-18-1, 2494479-1-1, 2494484-1-1, 2494487-1-1, 2515199-1-1, 2494485-1-1, 2494483-1-1, 2512827-1-1, 2512885-1-1, 2513077-1-1, 2518278-4-1, 2531556-1-1, 2525641-2-1, 2554943-1-1, 2551434-1-1, 2582151-1-1, 2585668-6-1, 2611315-1-1, 2611294-1-1, 2564890-3-1, 2619057-1-1, 2617112-1-1, 2618012-4-1, 2494648-5-1, 2500341-1-1, 2500318-2-1, 2509366-1-1, 2509367-1-1, 2512604-1-1, 2512604-1-2, 2514892-1-1, 2513939-2-1, 2516853-1-1, 2516851-1-1, 2523820-1-1, 2523643-2-1, 2524982-5-1, 2524982-1-1, 2537780-1-1, 2537780-1-2, 2522678-7-1, 2556763-18-1, 2556763-2-1, 2556763-18-2, 2562037-1-1, 2555090-2-1, 2576242-2-1, 2576259-2-1, 2554602-1-1, 2560382-1-1, 2560381-1-1, 2586284-1-1, 2612355-1-1, 2610438-1-1, 2611840-3-1, 2611351-2-1, 2611541-1-1, 2612895-4-1, 2604783-2-1, 2618900-1-1, 2618900-1-2, 2622551-1-1, 2623858-3-1, 2623105-1-1, 2623105-1-2, 2624426-2-1, 2626536-1-1, 2494497-1-1, 2494500-1-1, 2494525-1-1, 2503234-1-1, 2494526-1-1, 2495361-1-1, 2496099-1-1, 2516274-1-1, 2514690-1-1, 2495543-1-1, 2495400-1-1, 2495559-1-1, 2495409-1-1, 2519427-1-1, 2519427-1-2, 2519427-1-3, 2503764-1-1, 2522006-1-1, 2513114-3-1, 2514141-1-1, 2521852-1-1, 2512685-1-1, 2521540-1-1, 2522959-1-1, 2522761-1-1, 2519916-11-1, 2529981-1-1, 2529405-12-1, 2529406-11-1, 2529982-1-1, 2529409-9-1, 2529409-6-1, 2529434-1-1, 2543934-7-1, 2543933-7-1, 2543932-1-1, 2539102-7-1, 2539094-7-1, 2539436-1-1, 2540128-1-1, 2539084-8-1, 2539437-1-1, 2539438-1-1, 2528671-18-1, 2562722-1-1, 2568082-1-1, 2554514-1-1, 2547282-20-1, 2547282-6-1, 2547282-10-1, 2547282-8-1, 2547282-4-1, 2547231-10-1, 2547446-15-1, 2523479-16-1, 2552470-1-1, 2580467-1-1, 2561965-1-1, 2582525-6-1, 2582525-32-1, 2585579-2-1, 2592273-1-1, 2599020-1-1, 2579706-5-1, 2574524-13-1, 2562789-9-1, 2562789-2-1, 2562789-9-2, 2562789-5-1, 2562789-5-2, 2562789-4-1, 2562789-2-3, 2562789-37-1, 2562789-18-1, 2562789-14-1, 2562789-20-1, 2579723-1-1, 2580465-2-1, 2567049-24-1, 2580465-12-1, 2554559-19-1, 2604759-1-1, 2606427-1-1, 2575662-15-1, 2582438-1-1, 2612551-1-1, 2574524-13-2, 2611203-1-1, 2612612-10-1, 2611747-1-1, 2606474-1-1, 2611016-1-1, 2612734-18-1, 2612734-2-1, 2613185-15-1, 2613083-1-1, 2615930-1-1, 2613111-1-1, 2613111-20-1, 2618898-2-1, 2617993-2-1, 2617993-1-1, 2615926-1-1, 2616599-1-1, 2618691-30-1, 2616077-1-1, 2618690-20-1, 2621975-1-1, 2618903-10-1, 2622941-1-1, 2619537-1-1, 2621988-1-1, 2622940-1-1, 2620603-1-1, 2619182-2-1, 2625896-1-1, 2624865-1-1, 2624047-1-1, 2626740-2-1, 2624450-1-1, 2624725-18-1, 2624725-13-1, 2624725-5-1, 2624725-1-1, 2624725-13-2, 2577660-1-1, 2517501-1-1, 2545870-1-1, 2615937-3-1, 2616596-2-1, 2619984-2-1, 2616737-1-1, 2618481-5-1, 2499911-1-1, 2505712-1-1, 2503687-1-1, 2554926-1-1, 2572230-2-1, 2588785-1-1, 2495298-1-1, 2495264-1-1, 2509179-2-1, 2495358-1-1, 2515181-2-1, 2515181-1-1, 2513165-1-1, 2495299-1-1, 2531550-1-1, 2544853-2-1, 2574647-2-1, 2598884-1-1, 2618137-1-1, 2628573-1-1, 2494372-1-1, 2518561-9-1, 2522026-1-1, 2522834-1-1, 2524986-1-1, 2524986-7-1, 2524898-1-1, 2522845-1-1, 252
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- Finland
Related recalls
- FDACARESCAPE Canvas 1000, Model Numbers: a) 5865770; b) 5865770-01040169; c) 5865770-01054142; d)July 3, 2024
- FDACARESCAPE Canvas Smart Display, Model Numbers: a) 5514418, b) 5514418 / MAEP2, c) 5514418-01059642, d)July 3, 2024
- FDACARESCAPE B850 Model Numbers: a) MBC303, Model Numbers: 1) 2078633-001, 2) 2078633-010, 3) 2078633-012, 4)July 3, 2024
- FDACARESCAPE B650: Model Numbers: a) MBB313, Model Numbers: 1) 2095801-001, 2) 2095801-001-01018416, 3)July 3, 2024
- FDACUSA Excel C2600 23KHz Straight Handpiece-Indicated for use in surgical procedures where fragmentation,July 3, 2024
- FDASENSE XL TORSO COIL 1.5T Mk2. 16-element receive only coil for torso and abdomen imaging in Philips MR InteraJuly 3, 2024
Questions about this recall
Was CARESCAPE Canvas Smart Display, Model Numbers: a) 5700023, b) 5700023-02, c) 5939590-102; monitor, physiological, patient recalled?
Yes. FDA published recall Z-2169-2024 on July 3, 2024. Reported hazard: GE HealthCare has become aware of the potential for loss of monitoring on CARESCAPE Canvas 1000, CARESCAPE Canvas Smart Display, and certain versions of CARESCAPE B850 and CARESCAPE B650 monitors if, following the replacement of the CPU timekeeper battery, a mains power loss occurs during active monitoring. In the unlikely event that this occurs, it can result in the displays or monitors not turning on, after the mains power is restored. This can result in a delay in recognizing changes in patient condition until a replacement monitor is connected.
What should I do if I have this product?
Voluntary: Firm initiated · Letter