Siemens Syngo RT Therapist and Syngo RT Oncologist intended use of the SIEMENS branded Syngo RT Therapist and Syngo RT Oncologist software applications used with SIEMENS family of linear accelerator systems is to deliver X-Ray photon and electron radiation for the therapeutic treatment of cancer.
Published August 20, 2014 · Recall initiated: June 17, 2014
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Siemens became aware of a potential issue with the syngo RT Therapist 4.3.SP1 and syngo RT Oncologist 4.3SP1 software applications running on Siemens brand LINAC systems that may experience sporadic system crashes that will cause the inability to unload patient data.
What to do
Voluntary: Firm initiated · Letter
Recall details
- Manufacturer
- Siemens Medical Solutions USA, Inc
- Distributor
- UT, WI, NY
- Lot
- Syngo RT Therapist (RTT 4.3.SP1) - model no. 08162815 Syngo RT Oncologist (RTO 4.3.SP1) - model no. 10568604 Syngo RT Therapist (RTT 4.3.SP1) for U.S. affected units only: 10010 10046 10586 No Syngo RT Oncologist (RTO 4.3.SP1) listed for U.S.
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States