The NCB Polyaxial Locking Plate System NCB screws 26mm: 02.03155.026, non-sterile Product is implantable screws indicated for temporary internal fixation and stabilization of fractures and osteotomies of long bones. These screws are intended for bicortical anchorage.
Published August 22, 2012 · Recall initiated: July 13, 2012
Summary
FDA records medical device Class II recall Z-2187-2012, published August 22, 2012 and initiated July 13, 2012. Recalling firm: Zimmer, Inc.; Status: Terminated; Reported quantity: 24 to US Distributor with 7 further distributed; Lot: Part number 02.03155.026, lot 2649941. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Zimmer investigation following a complaint found one lot of NCB Polyaxial Locking Plate System NCB screws, 26mm, distributed in the United States contained incorrect 32 mm screws.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Zimmer, Inc.
- Distributor
- Nationwide distribution: USA including states of: AZ, IL, IN, ME and NJ.
- Lot
- Part number 02.03155.026, lot 2649941
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAOff-Axis Comprehensive Shoulder System Small, Augment, Off-Axis, Reamer Guide, Model/Catalog Number: 110040240July 1, 2026
- FDABrand Name: Persona Revision Product Name: Persona Revision® Trabecular Metal® Femoral Distal AugmentJuly 1, 2026
- FDAOff-Axis Alliance Glenoid, Left, 4-Peg, Augment, Reamer Guide, Model/Catalog Number: 110040444July 1, 2026
- FDAOff-Axis Comprehensive Shoulder System, Medium, Augment, Off-Axis, Reamer Guide Model/Catalog Number:July 1, 2026
- FDATAUT, Cholangiogram Catheter Intraducer, Peritoneal Catheter, Teleflex Medical, Research Triangle Park, NC. AnAugust 22, 2012
- FDAEdwards Lifesciences Swan-Ganz Thermodilution VIP Catheters Model 131F7J. For use in patients who requireAugust 22, 2012
Questions about this recall
Was The NCB Polyaxial Locking Plate System NCB screws 26mm: 02.03155.026, non-sterile Product is implantable screws indicated for temporary internal fixation and stabilization of fractures and osteotomies of long bones. These screws are intended for bicortical anchorage. recalled?
Yes. FDA published recall Z-2187-2012 on August 22, 2012. Reported hazard: Zimmer investigation following a complaint found one lot of NCB Polyaxial Locking Plate System NCB screws, 26mm, distributed in the United States contained incorrect 32 mm screws.
What should I do if I have this product?
Voluntary: Firm initiated · Letter