Brand Name: Vector TAS 8mm Screw. The Device Listing Number is E670535. Vector T AS 8mm Screw is intended to provide a fixed anchorage point for attachment of orthodontic appliances to facilitate the orthodontic movement of teeth. It is used temporarily and removed after orthodontic treatment has been completed. It is surgically placed in the bone of the upper or lower jaw arches to provide support for orthodontic devices.
Published September 18, 2013 · Recall initiated: July 12, 2013
Summary
FDA records medical device Class II recall Z-2194-2013, published September 18, 2013 and initiated July 12, 2013. Recalling firm: Ormco/Sybronendo; Status: Terminated; Reported quantity: 4,786; Lot: Lot Number 112911: 8 MM Vector TAS Screw, Part Number 601-0021 Lot Number 112911: Vector TAS Starter Kit (10 Patient), Part Number 601-0010. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Ormco Corporation is voluntarily recalling one lot of 8 MM Vector TAS Screws because some screws of this lot were machined undersized at the top of the threads may potentially cause a weakening of the screw. Use of the undersized product may result in breakage of the screw during insertion or removal when torqued, possibly requiring surgical removal of any remaining pieces.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Ormco/Sybronendo
- Distributor
- Worldwide Distribution - USA (nationwide) including the countries of Canada, Australia, New Zealand, EU and ROW consignees.
- Lot
- Lot Number 112911: 8 MM Vector TAS Screw, Part Number 601-0021 Lot Number 112911: Vector TAS Starter Kit (10 Patient), Part Number 601-0010
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAOrmco Damon Q2 Orthodontic Metal Bracket, Part No. 491-9834, L3R Hi Tq Hook, Tq +13 degrees, Rx. ManufacturedJune 17, 2020
- FDASnapLink, Ref Part No. 438-2190, L6R SnapLink, Slot .022, TQ--28, Ang +2, Rot +2, Qty 10, Made in Mexico, CEFebruary 26, 2020
- FDASealapex Xpress, Part No. 33639 The brand name of the device is Sealapex Xpress, a root canal filling resin.September 28, 2016
- FDASpirit MB Bracket, UR Central +14T +5A.018, Part Number 494-0110, Lot Number 081351013. Product Usage: AnApril 9, 2014
- FDA*** 1) Cat. 4341B; Qty. 10; Thoracentesis Tray; With Catheter; Sterile; CareFusion; 1500 Waukegan Road; McGawSeptember 18, 2013
- FDAAOA Mini RPE Screw Assembly , a component of Herbst, MARA and Fixed Metal devices. Product Usage: AOA Mini RPESeptember 18, 2013
Questions about this recall
Was Brand Name: Vector TAS 8mm Screw. The Device Listing Number is E670535. Vector T AS 8mm Screw is intended to provide a fixed anchorage point for attachment of orthodontic appliances to facilitate the orthodontic movement of teeth. It is used temporarily and removed after orthodontic treatment has been completed. It is surgically placed in the bone of the upper or lower jaw arches to provide support for orthodontic devices. recalled?
Yes. FDA published recall Z-2194-2013 on September 18, 2013. Reported hazard: Ormco Corporation is voluntarily recalling one lot of 8 MM Vector TAS Screws because some screws of this lot were machined undersized at the top of the threads may potentially cause a weakening of the screw. Use of the undersized product may result in breakage of the screw during insertion or removal when torqued, possibly requiring surgical removal of any remaining pieces.
What should I do if I have this product?
Voluntary: Firm initiated · Letter