Brand Name: Vector TAS 8mm Screw. The Device Listing Number is E670535. Vector T AS 8mm Screw is intended to provide a fixed anchorage point for attachment of orthodontic appliances to facilitate the orthodontic movement of teeth. It is used temporarily and removed after orthodontic treatment has been completed. It is surgically placed in the bone of the upper or lower jaw arches to provide support for orthodontic devices.

Published September 18, 2013 · Recall initiated: July 12, 2013

Summary

FDA records medical device Class II recall Z-2194-2013, published September 18, 2013 and initiated July 12, 2013. Recalling firm: Ormco/Sybronendo; Status: Terminated; Reported quantity: 4,786; Lot: Lot Number 112911: 8 MM Vector TAS Screw, Part Number 601-0021 Lot Number 112911: Vector TAS Starter Kit (10 Patient), Part Number 601-0010. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Ormco Corporation is voluntarily recalling one lot of 8 MM Vector TAS Screws because some screws of this lot were machined undersized at the top of the threads may potentially cause a weakening of the screw. Use of the undersized product may result in breakage of the screw during insertion or removal when torqued, possibly requiring surgical removal of any remaining pieces.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Ormco/Sybronendo
Distributor
Worldwide Distribution - USA (nationwide) including the countries of Canada, Australia, New Zealand, EU and ROW consignees.
Lot
Lot Number 112911: 8 MM Vector TAS Screw, Part Number 601-0021 Lot Number 112911: Vector TAS Starter Kit (10 Patient), Part Number 601-0010
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was Brand Name: Vector TAS 8mm Screw. The Device Listing Number is E670535. Vector T AS 8mm Screw is intended to provide a fixed anchorage point for attachment of orthodontic appliances to facilitate the orthodontic movement of teeth. It is used temporarily and removed after orthodontic treatment has been completed. It is surgically placed in the bone of the upper or lower jaw arches to provide support for orthodontic devices. recalled?

Yes. FDA published recall Z-2194-2013 on September 18, 2013. Reported hazard: Ormco Corporation is voluntarily recalling one lot of 8 MM Vector TAS Screws because some screws of this lot were machined undersized at the top of the threads may potentially cause a weakening of the screw. Use of the undersized product may result in breakage of the screw during insertion or removal when torqued, possibly requiring surgical removal of any remaining pieces.

What should I do if I have this product?

Voluntary: Firm initiated · Letter