AEQUALIS Reversed II or AEQUALIS Reversed Fracture Impactor. These products are contained within the Aequalis reversed II Instrument set tray no.YKAD83 and the Aequalis reversed fracture instrument set tray no.YKAD95. The items affected are MWD21, MWD023, MWD024 and MWD025. The impactor is an instrument used during shoulder arthroplasty; it is not implanted. The impactor is used to set the glenoid sphere, poly insert, spacer, and hemi adapter into place during surgery. The impactor is a two piece instrument which consists of a tip and a handle. The tip is made of plastic and metal that is pressed together for assembly.
Published September 25, 2013 · Recall initiated: August 13, 2013
Summary
FDA records medical device Class II recall Z-2199-2013, published September 25, 2013 and initiated August 13, 2013. Recalling firm: Tornier, Inc; Status: Terminated; Reported quantity: 2,298 (1045 US, 1253 OUS); Lot: all lots. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Following the identification of the disassociation of an impactor tip during surgery, Tornier is initiating a voluntary recall of all lots of Aequalis Reversed II and Reversed Fracture impactors. Upon review of the event, Tornier determined that a redesign of the instrument was warranted. These products are contained within the Aequalis Revered II instrument set tray no YKAD83 and the Aequalis R
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Tornier, Inc
- Distributor
- .Worldwide Distribution - USA including Puerto Rico and the states of AL, AZ, CA, CO,CT, FL, GA, IL, KY, MD, MA, MI, MN, NE, NJ, NC, OH, OR, PA and TX., and the countries of CANADA, MEXICO, HONG KONG, SOUTH AFRICA, ARGENTINA, COLOMBIA, ISRAEL, AUSTRALIA, BELGIUM, FINLAND,FRANCE, GERMANY, ITALY, NETHERLANDS, SLOVENIA, SPAIN, SWITZERLAND and UNITED KINGDOM
- Lot
- all lots
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAStryker Tornier Perform Humeral System, Reversed Insert, 45%, REF DWP2393, for shoulder arthroplastyOctober 9, 2024
- FDAstryker TORNIER PERFORM HUMERAL SYSTEM - STEMLESS US Anatomic Nucleus, REF DWM003A; anatomic totalSeptember 25, 2024
- FDAStryker Tornier Perform Reversed Glenoid Press-Fit Short Post Ti6AI4V, REF DWJ001, L: 7mm, Cementless Sterile.January 31, 2024
- FDASalto Talaris Tibial Guide, Part # MJU333 ,Tray/Kit Code YKAL11, Non sterile. Orthopedic Manual SurgicalSeptember 2, 2015
- FDAProTime Microcoagulation System; PROTIME CUVETTE AND PROTIME3 CUVETTE; Store the foil-pouched cuvettesSeptember 25, 2013
- FDAT8 REMOVAL HEAD, REF 8910641, QTY: 1 EA, Product Usage: The non-sterile T8 Removal Diriver is used with theSeptember 25, 2013
Questions about this recall
Was AEQUALIS Reversed II or AEQUALIS Reversed Fracture Impactor. These products are contained within the Aequalis reversed II Instrument set tray no.YKAD83 and the Aequalis reversed fracture instrument set tray no.YKAD95. The items affected are MWD21, MWD023, MWD024 and MWD025. The impactor is an instrument used during shoulder arthroplasty; it is not implanted. The impactor is used to set the glenoid sphere, poly insert, spacer, and hemi adapter into place during surgery. The impactor is a two piece instrument which consists of a tip and a handle. The tip is made of plastic and metal that is pressed together for assembly. recalled?
Yes. FDA published recall Z-2199-2013 on September 25, 2013. Reported hazard: Following the identification of the disassociation of an impactor tip during surgery, Tornier is initiating a voluntary recall of all lots of Aequalis Reversed II and Reversed Fracture impactors. Upon review of the event, Tornier determined that a redesign of the instrument was warranted. These products are contained within the Aequalis Revered II instrument set tray no YKAD83 and the Aequalis R
What should I do if I have this product?
Voluntary: Firm initiated · Letter