LVIS Intraluminal Support Device, REF/Outside Diameter(OD)xTotal Length(TL)xWorking Length(WL): 214022-CAS/4.0 mm x 22 mm x 18 mm, 214518-CAS/4.5 mm x 18 mm x 14 mm, 213517-CAS/3.5 mm x 17 mm x 13 mm
Published May 20, 2026 · Recall initiated: April 6, 2026
Summary
FDA records medical device Class II recall Z-2200-2026, published May 20, 2026 and initiated April 6, 2026. Recalling firm: MICROVENTION INC.; Status: Ongoing; Reported quantity: 38; Lot: Lots: 0000616462, 0000619851, 0000552987, 0000551725. UDI: (01)00842429115893(11)240325(17)270228(10)0000552987, (01)00842429115893(11)240325(17)270228(10)0000552987, (01)00842429115893(11)240325(17)270228(10)0000552987, (01)00842429115930(11)240325(17)270228(10)0000551725, (01)00842429115930(11)240325(17)270228(10)0000551725, (01)00842429115930(11)240610(17)270531(10)0000616462, (01)00842429115930(11)240610(17)270531(10)0000616462, (01)00842429115961(11)240610(17)270531(10)0000619851, (01)00842429115961(11)240610(17)270531(10)0000619851, (01)00842429115961(11)240610(17)270531(10)0000619851, (01)00842429115961(11)240610(17)270531(10)0000619851,. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Stent-assisted coiling device may due to commingling during manufacturing have dimensional nonconformance and/or incorrect device configuration. If device is oversized may lead to increased vessel wall stress, neointimal hyperplasia, vessel injury/rupture; if undersized may lead to incomplete expansion/malposition, acute stent thrombosis, distal embolization, ischemic neurological events
What to do
Voluntary: Firm initiated
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- MICROVENTION INC.
- Distributor
- International distribution to the country of China.
- Lot
- Lots: 0000616462, 0000619851, 0000552987, 0000551725. UDI: (01)00842429115893(11)240325(17)270228(10)0000552987, (01)00842429115893(11)240325(17)270228(10)0000552987, (01)00842429115893(11)240325(17)270228(10)0000552987, (01)00842429115930(11)240325(17)270228(10)0000551725, (01)00842429115930(11)240325(17)270228(10)0000551725, (01)00842429115930(11)240610(17)270531(10)0000616462, (01)00842429115930(11)240610(17)270531(10)0000616462, (01)00842429115961(11)240610(17)270531(10)0000619851, (01)00842429115961(11)240610(17)270531(10)0000619851, (01)00842429115961(11)240610(17)270531(10)0000619851, (01)00842429115961(11)240610(17)270531(10)0000619851,
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDABOBBY Balloon Guide Catheter REF: BOB-895-CN BOB-895-EMEA BOB-895-EMEA-AL BOB-895-US MV-BGA01095 For use inAugust 19, 2026
- FDAMedtronic Marinr Steerable, Product Number 72402; REPROCESSED ELECTROPHYSIOLOGY CATHETERMay 20, 2026
- FDAMedtronic SynchroMed Flex Infusion Mode - A810 Clinician Programmer Software Application, loaded on theMay 20, 2026
- FDAMicrovention, Headway" 17 Advanced Straight Microcatheter with Hydrophilic Coating, REF: MC172150SX-CN,January 29, 2025
- FDAERIC Retrieval Device REF: ER173020 ER174030 ER176044 The ERIC Retrieval Device is a mechanical thrombectomySeptember 25, 2024
- FDATERUMO HydroPearl Compressible Microspheres for Embolisation REF 8HP2S600February 7, 2024
Questions about this recall
Was LVIS Intraluminal Support Device, REF/Outside Diameter(OD)xTotal Length(TL)xWorking Length(WL): 214022-CAS/4.0 mm x 22 mm x 18 mm, 214518-CAS/4.5 mm x 18 mm x 14 mm, 213517-CAS/3.5 mm x 17 mm x 13 mm recalled?
Yes. FDA published recall Z-2200-2026 on May 20, 2026. Reported hazard: Stent-assisted coiling device may due to commingling during manufacturing have dimensional nonconformance and/or incorrect device configuration. If device is oversized may lead to increased vessel wall stress, neointimal hyperplasia, vessel injury/rupture; if undersized may lead to incomplete expansion/malposition, acute stent thrombosis, distal embolization, ischemic neurological events
What should I do if I have this product?
Voluntary: Firm initiated