FDAFDA · Food and drugs·Recall number Z-2208-2012

Philips Easy Diagnost Classic-Stationary x-ray system Catalog Numbers: 70642, 70643, 706030, 706031, 706032 Stationary x-ray system, Universal RF system for general use.

Published August 22, 2012 · Recall initiated: July 11, 2012

Summary

FDA records medical device Class II recall Z-2208-2012, published August 22, 2012 and initiated July 11, 2012. Recalling firm: Philips Healthcare Inc.; Status: Terminated; Reported quantity: 12 units; Lot: Serial Numbers: SN07000141 SN07000109 SN5000162 SN5000421 SN07000667 200148 200115 200081 300114. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Inclination switch in the stand may have been mounted incorrectly, as a result, the switch could not stop the tilting movement of the table base due tol hardware or software errors.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Philips Healthcare Inc.
Distributor
Worldwide Distribution - USA including IN, MA, MI, MO, NC, NY PA, SC, and WI and to the countries of Canada, HongKong, New Zealand.
Lot
Serial Numbers: SN07000141 SN07000109 SN5000162 SN5000421 SN07000667 200148 200115 200081 300114
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was Philips Easy Diagnost Classic-Stationary x-ray system Catalog Numbers: 70642, 70643, 706030, 706031, 706032 Stationary x-ray system, Universal RF system for general use. recalled?

Yes. FDA published recall Z-2208-2012 on August 22, 2012. Reported hazard: Inclination switch in the stand may have been mounted incorrectly, as a result, the switch could not stop the tilting movement of the table base due tol hardware or software errors.

What should I do if I have this product?

Voluntary: Firm initiated · Letter