Trabecular Metal Reverse Shoulder Liner 40mm, part 00-4309-028-00 - Per surgical technique 97-4309-003-00 Revision 4: "Complete Poly Liner insertion by snapping the appropriate 60¿ (Standard) or 65¿ (Retentive) Poly Liner Impactor to the Poly Liner Impactor Handle. Then, place the Poly Liner Impactor on the Poly Liner articular surface and forcefully strike the Poly Liner Handle with a mallet."
Published August 22, 2012 · Recall initiated: July 18, 2012
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Lot specific Trabecular Metal (TM) Reverse Shoulder Liner Impactors are being recalled due to reports of alignment peg fractures. All lots of TM Reverse Shoulder Liner Impactors manufactured from 455 stainless steel are being retrieved, as all reported fractured instruments were manufactured with this material. The TM Reverse Shoulder Liner Impactor units that were manufactured with 13-8 stain
What to do
Voluntary: Firm initiated · Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
Recall details
- Manufacturer
- Zimmer, Inc.
- Distributor
- Worldwide Distribution - USA (nationwide) and internationally to Germany, Belgium, Denmark, Finland, France, Italy, Netherlands, Spain, Sweden, Switzerland, and the United Kingdom.
- Lot
- part 00-4309-028-00, lot 60443178, 60545332, 60549563, 60605851, 60612791, 60720047 60758647 60773944 60836577 60896973, 60973522, 61019783, 61024158
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States