FDAFDA · Food and drugs·Recall number Z-2213-2026

OneLIF Interbody Cage (REF/Description): 010-107-1007/Small 7 Degree 10mm X 7mm, 010-107-1007-2/Small 7 Degree 10mm X 7mm - Oblique, 010-107-1209/Small 7 Degree 12mm X 9mm, 010-107-1411/Small 7 Degree 14mm X 11mm, 010-107-1613/Small 7 Degree 16mm X 13mm, 010-113-1006/Small, 13 Degree 10mm X 6mm, 010-113-1006-2/Small, 13 Degree, 10mm X 6mm - Oblique, 010-113-1208/Small 13 Degree 12mm X 8mm, 010-113-1410/Small 13 Degree 14mm X 10mm, 010-113-1612/Small 13 Degree 16mm X 12mm, 010-119-1407/19 Degree 14mm X 7mm, 010-119-1609/Small 19 Degree 16mm X 9mm, 010-119-1811/Small 19 Degree 18mm X 11mm, 010-125-1606/Small 25 Degree 16mm X 6mm, 010-125-1808/Small 25 Degree 18mm X 8mm, 010-207-1007/Medium 7 Degree 10mm X 7mm, 010-207-1007-2/Medium 7 Degree 10mm X 7mm - Oblique, 010-207-1209/Medium 7 Degree 12mm X 9mm, 010-207-1411/Medium 7 Degree 14mm X 11mm, 010-207-1613/Medium 7 Degree 16mm X 13mm, 010-213-1206/Medium 13 Degree 12mm X 6mm, 010-213-1408/Medium 13 Degree 14mm X 8mm, 010-213-1610/Medium 13 Degree 16mm X 10mm, 010-213-1812/Medium 13 Degree 18mm X 12mm, 010-219-1406/Medium 19 Degree 14mm X 6mm, 010-219-1608/Medium 19 Degree 16mm X 8mm, 010-219-1810/Medium 19 Degree 18mm X 10mm, 010-225-1807/Medium 25 Degree 18mm X 7mm, 010-225-2009/Medium 25 Degree 20mm X 9mm, 010-307-1006/Large 7 Degree,10mm X 6mm, 010-307-1006-2/Large 7 Degree 10mm X 6mm - Oblique, 010-307-1208/ Large 7 Degree 12mm X 8mm, 010-307-1410/Large 7 Degree 14mm X 10mm, 010-307-1612/Large 7 Degree 16mm X 12mm, 010-313-1205/Large 13 Degree 12mm X 5mm, 010-313-1407/Large 13 Degree 14mm X 7mm, 010-313-1609/Large 13 Degree 16mm X 9mm, 010-313-1811/Large 13 Degree 18mm X 11mm, 010-313-2013/Large 13 Degree 20mm X 13mm, 010-319-1606/Large 19 Degree 16mm X 6mm, 010-319-1808/ Large 19 Degree 18mm X 8mm, 010-319-2010/ Large 19 Degree 20mm X 10mm, 010-325-2007/Large 25 Degree 20mm X 7mm, 010-325-2209/ Large 25 Degree 22mm X 9mm, Implant Kits containing OneLIF Interbody Cages, Name/Kit Family: Kit-OneLIF-10 mm (716)/Kit-OneLIF-10 mm, Kit-OneLIF-10 mm (717)/Kit-OneLIF-10 mm, Kit-OneLIF-10 mm (718)/Kit-OneLIF-10 mm, Kit-OneLIF-10 mm (719)/Kit-OneLIF-10 mm, Kit-OneLIF-10 mm (722)/Kit-OneLIF-10 mm, Kit-OneLIF-10 mm (723)/Kit-OneLIF-10 mm, Kit-OneLIF-10 mm (724)/Kit-OneLIF-10 mm, Kit-OneLIF-10 mm (733)/Kit-OneLIF-10 mm, Kit-OneLIF-10 mm (734)/Kit-OneLIF-10 mm, Kit-OneLIF-10 mm (737)/Kit-OneLIF-10 mm, Kit-OneLIF-10 mm (738)/Kit-OneLIF-10 mm, Kit-OneLIF-10 mm (746)/Kit-OneLIF-10 mm, Kit-OneLIF-10 mm (751)/Kit-OneLIF-10 mm, Kit-OneLIF-10 mm (761)/Kit-OneLIF-10 mm, Kit-OneLIF-10 mm (762)/Kit-OneLIF-10 mm, Kit-OneLIF-10 mm (787)/Kit-OneLIF-10 mm, Kit-OneLIF- IMPOneLIF Implant Kit (701)/Kit-OneLIF-IMP - OneLIF Implant Kit, Kit-OneLIF-IMP - OneLIF Implant Kit (702)/Kit-OneLIF-IMP - OneLIF Implant Kit, Kit-OneLIF-IMP - OneLIF Implant Kit (703) NL/Kit-OneLIF-IMP - OneLIF Implant Kit, Kit-OneLIF-IMP - OneLIF Implant Kit (704)/Kit-OneLIF-IMP - OneLIF Implant Kit, Kit-OneLIF-IMP - OneLIF Implant Kit (705)/Kit-OneLIF-IMP - OneLIF Implant Kit, Kit-OneLIF-IMP - OneLIF Implant Kit (706)/Kit-OneLIF-IMP - OneLIF Implant Kit, Kit-OneLIF-IMP - OneLIF Implant Kit (708)/Kit-OneLIF-IMP - OneLIF Implant Kit, Kit-OneLIF-IMP - OneLIF Implant Kit (709)/Kit-OneLIF-IMP - OneLIF Implant Kit,

Published May 27, 2026 · Recall initiated: April 23, 2026

Summary

FDA records medical device Class II recall Z-2213-2026, published May 27, 2026 and initiated April 23, 2026. Recalling firm: Novapproach Spine, LLC; Status: Ongoing; Reported quantity: 1259; Lot: REF/UDI-DI/ Cage Lots: 010-107-1007/G07001010710070/ 103003, 105701, 118001, 129301; 010-107-1007-2/G070010107100720/ 114301, 117509, 128703; 010-107-1209/G07001010712090/ 102714, 105702, 117501; 010-107-1411/G07001010714110/ 103019, 105703; 010-107-1613/G07001010716130/ 103020, 105704, 127701; 010-113-1006/G07001011310060/ 103021, 105705, 118002, 129801; 010-113-1006-2/G070010113100620/ 117510, 127716, 114302; 010-113-1208/G07001011312080/ 103022, 105706, 117502, 126902; 010-113-1410/G07001011314100/ 102715, 105707, 127702; 010-113-1612/G07001011316120/ 103023, 105708, 127703; 010-119-1407/G07001011914070/ 103024, 105709, 125101, 128701; 010-119-1609/G07001011916090/ 102716, 105710, 127001; 010-119-1811/G07001011918110/ 103025, 105711, 127003; 010-125-1606/G07001012516060/ 102717, 117901, 127002; 010-125-1808/G07001012518080/ 103005, 117902, 127704; 010-207-1007/G07001020710070/ 102702, 106201, 118003, 125102; 010-207-1007-2/G070010207100720/ 117511, 124106, 127717, 114303; 010-207-1209/G07001020712090/ 102701, 105401, 124101, 127705, 101201; 010-207-1411/G07001020714110/ 105001, 105402, 101203; 010-207-1613/G07001020716130/ 103007, 105403, 124102, 127706; 010-213-1206/G07001021312060/ 105002, 105404, 121202, 124103, 127707, 101204; 010-213-1408/G07001021314080/ 102704, 105405, 121203, 124104, 127708; 010-213-1610/G07001021316100/ 103008, 105406, 130101; 010-213-1812/G07001021318120/ 103009, 105407, 127709; 010-219-1406/G07001021914060/ 102705, 105408, 117503, 119301, 121201, 124105; 010-219-1608/G07001021916080/ 103010, 105409, 121204, 127710; 010-219-1810/G07001021918100/ 103011, 105410, 127711; 010-225-1807/G07001022518070/ 102706, 113701, 117001, 121205; 010-225-2009/G07001022520090/ 103001, 117903, 127004, 127712; 010-307-1006/G07001030710060/ 102707, 106202, 118004, 129302; 010-307-1006-2/G070010307100620/ 117512, 127718, 114304; 010-307-1208/G07001030712080/ 102709, 105601, 117504; 010-307-1410/G07001030714100/ 102710, 105602, 127713; 010-307-1612/G07001030716120/ 103012, 105603, 117505; 010-313-1205/G07001031312050/ 103013, 105604, 117506, 129303; 010-313-1407/G07001031314070/ 102711, 105605, 117507, 124107, 128702, 130102; 010-313-1609/G07001031316090/ 103014, 105606; 010-313-1811/G07001031318110/ 103015, 105607, 127714; 010-313-2013/G07001031320130/ 103016, 105608, 127715; 010-319-1606/G07001031916060/ 102712, 105609, 114905, 117508; 010-319-1808/G07001031918080/ 103017, 105610, 125103; 010-319-2010/G07001031920100/ 103018, 105611, 125104; 010-325-2007/G07001032520070/ 102713, 105612, 121206; 010-325-2209/G07001032522090/ 105613, 121207, 101206, 101207, 101208. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Intervertebral body fusion system straight inserter failed to properly attach to affected cages, which may result in an inability to securely engage the implant with the inserter, which could lead to procedural delay and/or the need to select an alternative implant.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Novapproach Spine, LLC
Distributor
US Nationwide distribution in the states of NY, CT, CA, TX, FL, MI.
Lot
REF/UDI-DI/ Cage Lots: 010-107-1007/G07001010710070/ 103003, 105701, 118001, 129301; 010-107-1007-2/G070010107100720/ 114301, 117509, 128703; 010-107-1209/G07001010712090/ 102714, 105702, 117501; 010-107-1411/G07001010714110/ 103019, 105703; 010-107-1613/G07001010716130/ 103020, 105704, 127701; 010-113-1006/G07001011310060/ 103021, 105705, 118002, 129801; 010-113-1006-2/G070010113100620/ 117510, 127716, 114302; 010-113-1208/G07001011312080/ 103022, 105706, 117502, 126902; 010-113-1410/G07001011314100/ 102715, 105707, 127702; 010-113-1612/G07001011316120/ 103023, 105708, 127703; 010-119-1407/G07001011914070/ 103024, 105709, 125101, 128701; 010-119-1609/G07001011916090/ 102716, 105710, 127001; 010-119-1811/G07001011918110/ 103025, 105711, 127003; 010-125-1606/G07001012516060/ 102717, 117901, 127002; 010-125-1808/G07001012518080/ 103005, 117902, 127704; 010-207-1007/G07001020710070/ 102702, 106201, 118003, 125102; 010-207-1007-2/G070010207100720/ 117511, 124106, 127717, 114303; 010-207-1209/G07001020712090/ 102701, 105401, 124101, 127705, 101201; 010-207-1411/G07001020714110/ 105001, 105402, 101203; 010-207-1613/G07001020716130/ 103007, 105403, 124102, 127706; 010-213-1206/G07001021312060/ 105002, 105404, 121202, 124103, 127707, 101204; 010-213-1408/G07001021314080/ 102704, 105405, 121203, 124104, 127708; 010-213-1610/G07001021316100/ 103008, 105406, 130101; 010-213-1812/G07001021318120/ 103009, 105407, 127709; 010-219-1406/G07001021914060/ 102705, 105408, 117503, 119301, 121201, 124105; 010-219-1608/G07001021916080/ 103010, 105409, 121204, 127710; 010-219-1810/G07001021918100/ 103011, 105410, 127711; 010-225-1807/G07001022518070/ 102706, 113701, 117001, 121205; 010-225-2009/G07001022520090/ 103001, 117903, 127004, 127712; 010-307-1006/G07001030710060/ 102707, 106202, 118004, 129302; 010-307-1006-2/G070010307100620/ 117512, 127718, 114304; 010-307-1208/G07001030712080/ 102709, 105601, 117504; 010-307-1410/G07001030714100/ 102710, 105602, 127713; 010-307-1612/G07001030716120/ 103012, 105603, 117505; 010-313-1205/G07001031312050/ 103013, 105604, 117506, 129303; 010-313-1407/G07001031314070/ 102711, 105605, 117507, 124107, 128702, 130102; 010-313-1609/G07001031316090/ 103014, 105606; 010-313-1811/G07001031318110/ 103015, 105607, 127714; 010-313-2013/G07001031320130/ 103016, 105608, 127715; 010-319-1606/G07001031916060/ 102712, 105609, 114905, 117508; 010-319-1808/G07001031918080/ 103017, 105610, 125103; 010-319-2010/G07001031920100/ 103018, 105611, 125104; 010-325-2007/G07001032520070/ 102713, 105612, 121206; 010-325-2209/G07001032522090/ 105613, 121207, 101206, 101207, 101208
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was OneLIF Interbody Cage (REF/Description): 010-107-1007/Small 7 Degree 10mm X 7mm, 010-107-1007-2/Small 7 Degree 10mm X 7mm - Oblique, 010-107-1209/Small 7 Degree 12mm X 9mm, 010-107-1411/Small 7 Degree 14mm X 11mm, 010-107-1613/Small 7 Degree 16mm X 13mm, 010-113-1006/Small, 13 Degree 10mm X 6mm, 010-113-1006-2/Small, 13 Degree, 10mm X 6mm - Oblique, 010-113-1208/Small 13 Degree 12mm X 8mm, 010-113-1410/Small 13 Degree 14mm X 10mm, 010-113-1612/Small 13 Degree 16mm X 12mm, 010-119-1407/19 Degree 14mm X 7mm, 010-119-1609/Small 19 Degree 16mm X 9mm, 010-119-1811/Small 19 Degree 18mm X 11mm, 010-125-1606/Small 25 Degree 16mm X 6mm, 010-125-1808/Small 25 Degree 18mm X 8mm, 010-207-1007/Medium 7 Degree 10mm X 7mm, 010-207-1007-2/Medium 7 Degree 10mm X 7mm - Oblique, 010-207-1209/Medium 7 Degree 12mm X 9mm, 010-207-1411/Medium 7 Degree 14mm X 11mm, 010-207-1613/Medium 7 Degree 16mm X 13mm, 010-213-1206/Medium 13 Degree 12mm X 6mm, 010-213-1408/Medium 13 Degree 14mm X 8mm, 010-213-1610/Medium 13 Degree 16mm X 10mm, 010-213-1812/Medium 13 Degree 18mm X 12mm, 010-219-1406/Medium 19 Degree 14mm X 6mm, 010-219-1608/Medium 19 Degree 16mm X 8mm, 010-219-1810/Medium 19 Degree 18mm X 10mm, 010-225-1807/Medium 25 Degree 18mm X 7mm, 010-225-2009/Medium 25 Degree 20mm X 9mm, 010-307-1006/Large 7 Degree,10mm X 6mm, 010-307-1006-2/Large 7 Degree 10mm X 6mm - Oblique, 010-307-1208/ Large 7 Degree 12mm X 8mm, 010-307-1410/Large 7 Degree 14mm X 10mm, 010-307-1612/Large 7 Degree 16mm X 12mm, 010-313-1205/Large 13 Degree 12mm X 5mm, 010-313-1407/Large 13 Degree 14mm X 7mm, 010-313-1609/Large 13 Degree 16mm X 9mm, 010-313-1811/Large 13 Degree 18mm X 11mm, 010-313-2013/Large 13 Degree 20mm X 13mm, 010-319-1606/Large 19 Degree 16mm X 6mm, 010-319-1808/ Large 19 Degree 18mm X 8mm, 010-319-2010/ Large 19 Degree 20mm X 10mm, 010-325-2007/Large 25 Degree 20mm X 7mm, 010-325-2209/ Large 25 Degree 22mm X 9mm, Implant Kits containing OneLIF Interbody Cages, Name/Kit Family: Kit-OneLIF-10 mm (716)/Kit-OneLIF-10 mm, Kit-OneLIF-10 mm (717)/Kit-OneLIF-10 mm, Kit-OneLIF-10 mm (718)/Kit-OneLIF-10 mm, Kit-OneLIF-10 mm (719)/Kit-OneLIF-10 mm, Kit-OneLIF-10 mm (722)/Kit-OneLIF-10 mm, Kit-OneLIF-10 mm (723)/Kit-OneLIF-10 mm, Kit-OneLIF-10 mm (724)/Kit-OneLIF-10 mm, Kit-OneLIF-10 mm (733)/Kit-OneLIF-10 mm, Kit-OneLIF-10 mm (734)/Kit-OneLIF-10 mm, Kit-OneLIF-10 mm (737)/Kit-OneLIF-10 mm, Kit-OneLIF-10 mm (738)/Kit-OneLIF-10 mm, Kit-OneLIF-10 mm (746)/Kit-OneLIF-10 mm, Kit-OneLIF-10 mm (751)/Kit-OneLIF-10 mm, Kit-OneLIF-10 mm (761)/Kit-OneLIF-10 mm, Kit-OneLIF-10 mm (762)/Kit-OneLIF-10 mm, Kit-OneLIF-10 mm (787)/Kit-OneLIF-10 mm, Kit-OneLIF- IMPOneLIF Implant Kit (701)/Kit-OneLIF-IMP - OneLIF Implant Kit, Kit-OneLIF-IMP - OneLIF Implant Kit (702)/Kit-OneLIF-IMP - OneLIF Implant Kit, Kit-OneLIF-IMP - OneLIF Implant Kit (703) NL/Kit-OneLIF-IMP - OneLIF Implant Kit, Kit-OneLIF-IMP - OneLIF Implant Kit (704)/Kit-OneLIF-IMP - OneLIF Implant Kit, Kit-OneLIF-IMP - OneLIF Implant Kit (705)/Kit-OneLIF-IMP - OneLIF Implant Kit, Kit-OneLIF-IMP - OneLIF Implant Kit (706)/Kit-OneLIF-IMP - OneLIF Implant Kit, Kit-OneLIF-IMP - OneLIF Implant Kit (708)/Kit-OneLIF-IMP - OneLIF Implant Kit, Kit-OneLIF-IMP - OneLIF Implant Kit (709)/Kit-OneLIF-IMP - OneLIF Implant Kit, recalled?

Yes. FDA published recall Z-2213-2026 on May 27, 2026. Reported hazard: Intervertebral body fusion system straight inserter failed to properly attach to affected cages, which may result in an inability to securely engage the implant with the inserter, which could lead to procedural delay and/or the need to select an alternative implant.

What should I do if I have this product?

Voluntary: Firm initiated · Letter