Back to search
FDAFDA · Food and drugs·Recall number Z-2215-2024

2nd Generation CentriMag Primary Console, REF: 201-90401, 201-90701, 201-90411, 201-90421, 201-30300, L201-90401, L201-90411, L201-90421, part of the CentriMag Circulatory Support System

Published July 3, 2024 · Recall initiated: May 22, 2024

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Due to a component change, the console part of a circulatory support system may not conform to the IEC 61000-4-5 standard, so if subjected to a power surge above 1.8kV and up to 2.0 kV, the console may shut down completely without alarm, pumping would cease, and the patient would have to be switched to a backup circulatory support system.

What to do

Voluntary: Firm initiated · Letter

Recall details

Manufacturer
Thoratec Switzerland GMBH
Distributor
Worldwide - US Nationwide distribution included in the states of CA, PA, FL, IL, NY, AR, GA, WI, AZ, TN, TX, NC, MT, VA, MA, NJ, NE, WV, ME, OK, CO, OH, KC, WA, IN, CT, MI, OR, ID, MO, SC, KY, LA, NM, NV, VT, IA, MD and the countries of ALGERIA, AUSTRALIA, AUSTRIA, BANGLADESH, BULGARIA, CANADA, CAYMAN ISLANDS, CHILE, COLOMBIA, COSTA RICA, CZECH REPUBLIC, DENMARK, DOMINICAN REPUBLIC, ESTONIA, FINLAND, FRANCE, GERMANY, GREECE, HONG KONG, INDIA, IRAN, IRAQ, ISRAEL, ITALY, KENYA, KOREA, KUWAIT, LITHUANIA, MACEDONIA MALAYSIA, MARTINIQUE, MEXICO, MOROCCO, NETHERLANDS, NORWAY, OMAN, PERU, POLAND, PORTUGAL, QATAR, ROMANIA, RUSSIA, SAUDI ARABIA, SINGAPORE, SLOVENIA, SOUTH AFRICA, SPAIN, SWEDEN, SWITZERLAND, TAIWAN, THAILAND, TUNISIA, UNITED ARAB EMIRATES, UNITED KINGDOM.
Lot
REF/Serial Number or Range: 201-30300/ L00610-0004-L06786-0002 201-90401/ 19234129-20119751 and L00610-0011-L08005-0001 201-90411/19234128-20190463 and L03849-0004-L08002-0001 201-90421/19474336-19606873 and L04780-0005-L07601-0001 201-90701/ 19415572 and L03774-0007-L08007-0001 L201-90401/ L00694-0006-L06579-0004 and L201-90401 L201-90411/ 19258710-19605455 and L01627-0007-L07980-0001 L201-90421/ L01716-0010-L06482-0005 UDI-DI: 07640135140689, 07640135140726, 07640135140702, 07640135140719, 07640135140856, 05415067037282, 07640135140894, 07640135149019, 07640135149002, 07640135140702, 07640135140719
Risk classification
Class II
Area (FDA)
device
Country
Switzerland

Sell products like this?

Recalert watches your whole catalog for recalls like this one, automatically.

Start free — no card