SENSATION 7 Fr. IAB, Intra-Aortic Balloon Catheter and Accessories, used to provide counter pulsation therapy in the aorta Product Description Part Number SENSATION 7 Fr. 34cc IAB WITH ACCESSORIES (CHINA) D684-00-0469-01C SENSATION 7 Fr. 34cc IAB WITH ACCESSORIES (INDIA) D684-00-0469-07 SENSATION 7 Fr. 40cc IAB WITH ACCESSORIES (CHINA) D684-00-0470-01C SENSATION 7 Fr. 40cc IAB WITH ACCESSORIES (INDIA) D684-00-0470-07 SENSATION 7 Fr. 40cc IAB WITH ACCESSORIES (TURKEY) D684-00-0470-09 SENSATION 7 Fr. 40cc IAB WITH ACCESSORIES D684-00-0470-01 SENSATION 7 Fr. 40cc IAB WITH ACCESSORIES, US ONLY D684-00-0470-01U SENSATION 7 Fr. 34cc IAB WITH ACCESSORIES, US ONLY D684-00-0469-01U SENSATION 7 Fr. 34cc IAB WITH ACCESSORIES D684-00-0469-01 SENSATION 7 Fr. 34cc IAB WITH ACCESSORIES, TURKEY D684-00-0469-09 PACKAGED INSERTION KIT - SENSATION 7 Fr. 34/40 cc IABs D884-00-0019-16
Published August 2, 2023 · Recall initiated: June 23, 2023
Summary
FDA records medical device Class II recall Z-2217-2023, published August 2, 2023 and initiated June 23, 2023. Recalling firm: Datascope Corp.; Status: Ongoing; Reported quantity: 321,609 total kits; Lot: Product Description UDI SENSATION 7 Fr. 34cc IAB WITH ACCESSORIES (CHINA) 10607567109367 SENSATION 7 Fr. 34cc IAB WITH ACCESSORIES (INDIA) 10607567109732 SENSATION 7 Fr. 40cc IAB WITH ACCESSORIES (CHINA) 10607567109374 SENSATION 7 Fr. 40cc IAB WITH ACCESSORIES (INDIA) 10607567109749 SENSATION 7 Fr. 40cc IAB WITH ACCESSORIES (TURKEY) 10607567113135 SENSATION 7 Fr. 40cc IAB WITH ACCESSORIES 10607567106779 SENSATION 7 Fr. 40cc IAB WITH ACCESSORIES, US ONLY 10607567109558 SENSATION 7 Fr. 34cc IAB WITH ACCESSORIES, US ONLY 10607567109541 SENSATION 7 Fr. 34cc IAB WITH ACCESSORIES 10607567106755 SENSATION 7 Fr. 34cc IAB WITH ACCESSORIES, TURKEY 10607567113128 PACKAGED INSERTION KIT - SENSATION 7 Fr. 34/40 cc IABs 10607567106724. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
During IAB catheter insertion, the introducer dilator may fracture at the hub, leaving the body of the introducer dilator housed within the sheath. Risks to health include damage to the femoral artery, descending aorta, or embolization; surgical intervention and delay of IAB therapy may also result. Firm has received 10 complaints, including 3 serious adverse events and 1 patient death.
What to do
Voluntary: Firm initiated
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Datascope Corp.
- Distributor
- Domestic distribution nationwide. International distribution worldwide.
- Lot
- Product Description UDI SENSATION 7 Fr. 34cc IAB WITH ACCESSORIES (CHINA) 10607567109367 SENSATION 7 Fr. 34cc IAB WITH ACCESSORIES (INDIA) 10607567109732 SENSATION 7 Fr. 40cc IAB WITH ACCESSORIES (CHINA) 10607567109374 SENSATION 7 Fr. 40cc IAB WITH ACCESSORIES (INDIA) 10607567109749 SENSATION 7 Fr. 40cc IAB WITH ACCESSORIES (TURKEY) 10607567113135 SENSATION 7 Fr. 40cc IAB WITH ACCESSORIES 10607567106779 SENSATION 7 Fr. 40cc IAB WITH ACCESSORIES, US ONLY 10607567109558 SENSATION 7 Fr. 34cc IAB WITH ACCESSORIES, US ONLY 10607567109541 SENSATION 7 Fr. 34cc IAB WITH ACCESSORIES 10607567106755 SENSATION 7 Fr. 34cc IAB WITH ACCESSORIES, TURKEY 10607567113128 PACKAGED INSERTION KIT - SENSATION 7 Fr. 34/40 cc IABs 10607567106724
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDACardiosave Rescue. Intra-Aortic Balloon Pump system.December 17, 2025
- FDAMEGA 7.5Fr. 40cc Intra-Aortic Balloon with Accessories, Model Number 0684-00- 0295-01November 13, 2024
- FDACARDIOSAVE RESCUE INTRA-AORTIC BALLOON PUMP (IABP). Electromechanical system used to inflate and deflateOctober 2, 2024
- FDACARDIOSAVE RESCUE INTRA-AORTIC BALLOON PUMP (IABP). Electromechanical system used to inflate and deflateOctober 2, 2024
- FDAHAMILTON C2 Ventilator Is intended to provide positive pressure ventilatory support to adults and pediatrics,August 2, 2023
- FDALINEAR 7.5Fr.IAB, Intra-Aortic Balloon Catheter and Accessories, used to provide counter pulsation therapy inAugust 2, 2023
Questions about this recall
Was SENSATION 7 Fr. IAB, Intra-Aortic Balloon Catheter and Accessories, used to provide counter pulsation therapy in the aorta Product Description Part Number SENSATION 7 Fr. 34cc IAB WITH ACCESSORIES (CHINA) D684-00-0469-01C SENSATION 7 Fr. 34cc IAB WITH ACCESSORIES (INDIA) D684-00-0469-07 SENSATION 7 Fr. 40cc IAB WITH ACCESSORIES (CHINA) D684-00-0470-01C SENSATION 7 Fr. 40cc IAB WITH ACCESSORIES (INDIA) D684-00-0470-07 SENSATION 7 Fr. 40cc IAB WITH ACCESSORIES (TURKEY) D684-00-0470-09 SENSATION 7 Fr. 40cc IAB WITH ACCESSORIES D684-00-0470-01 SENSATION 7 Fr. 40cc IAB WITH ACCESSORIES, US ONLY D684-00-0470-01U SENSATION 7 Fr. 34cc IAB WITH ACCESSORIES, US ONLY D684-00-0469-01U SENSATION 7 Fr. 34cc IAB WITH ACCESSORIES D684-00-0469-01 SENSATION 7 Fr. 34cc IAB WITH ACCESSORIES, TURKEY D684-00-0469-09 PACKAGED INSERTION KIT - SENSATION 7 Fr. 34/40 cc IABs D884-00-0019-16 recalled?
Yes. FDA published recall Z-2217-2023 on August 2, 2023. Reported hazard: During IAB catheter insertion, the introducer dilator may fracture at the hub, leaving the body of the introducer dilator housed within the sheath. Risks to health include damage to the femoral artery, descending aorta, or embolization; surgical intervention and delay of IAB therapy may also result. Firm has received 10 complaints, including 3 serious adverse events and 1 patient death.
What should I do if I have this product?
Voluntary: Firm initiated