Back to search
FDAFDA · Food and drugs·Recall number Z-2226-2024

Sterile surgical convenience kit: Medline Neuro IR Pack, Pack #DYNJ56073C, containing a Pure Pouch with component Medline Reorder #DYNJ04048, Kelly Forceps/STD STR Sterile.

Published July 10, 2024 · Recall initiated: March 14, 2024

Summary

FDA records medical device Class II recall Z-2226-2024, published July 10, 2024 and initiated March 14, 2024. Recalling firm: MEDLINE INDUSTRIES, LP - Northfield; Status: Ongoing; Reported quantity: 12,588 total sterile kits; Lot: Pack #DYNJ56073C - lot numbers 23IDA509, exp. 5/31/2026, and 23DDC159, exp. 11/30/2025, UDI-DI each-10193489441833, UDI-DI case-40193489441834. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Surgical convenience kits containing Pure Pouch component that has weak seals which may result in a breach in sterility of the contents if the seal fails.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
MEDLINE INDUSTRIES, LP - Northfield
Distributor
US Nationwide distribution in the states of AL, CA, CO, CT, FL, IL, IN, MD, MN, MT, NC, NJ, NV, NY, OH, PA, SC, TN, TX, VA, WA, WI, and WY.
Lot
Pack #DYNJ56073C - lot numbers 23IDA509, exp. 5/31/2026, and 23DDC159, exp. 11/30/2025, UDI-DI each-10193489441833, UDI-DI case-40193489441834.
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was Sterile surgical convenience kit: Medline Neuro IR Pack, Pack #DYNJ56073C, containing a Pure Pouch with component Medline Reorder #DYNJ04048, Kelly Forceps/STD STR Sterile. recalled?

Yes. FDA published recall Z-2226-2024 on July 10, 2024. Reported hazard: Surgical convenience kits containing Pure Pouch component that has weak seals which may result in a breach in sterility of the contents if the seal fails.

What should I do if I have this product?

Voluntary: Firm initiated · Letter