FDAFDA · Food and drugs·Recall number Z-2238-2013

TWINFIX" Ultra 4.5 mm PLLA-HA Suture Anchor with two #2 ULTRABRAID Sutures (blue, blue-COBRAID") with Needles; Part Number: 72202612 Biodegradable suture anchor

Published September 25, 2013 · Recall initiated: August 6, 2013

Summary

FDA records medical device Class II recall Z-2238-2013, published September 25, 2013 and initiated August 6, 2013. Recalling firm: Smith & Nephew, Inc. Endoscopy Division; Status: Terminated; Reported quantity: 1571 US; Lot: 50373723 50389654 50389655 50397577 50397596 50397599 50397605 50417226 50431744 50431853 50433695 50437528 50440608 50441690 50447116 50447118 50447119 50448227 50454009 50454867 50461412 50467503 50469159 50470653. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Pinholes detected in a small number of primary pouches that may constitute a breach of the sterile barrier.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Smith & Nephew, Inc. Endoscopy Division
Distributor
Worldwide Distribution - USA (nationwide) Canada, Europe: Austria, Belgium, Cyprus, Czech Republic, Denmark, Germany, Greece, Finland, France, Italy, Ireland, Luxembourg, Netherlands, Norway, Poland, Portugal, Russia, Slovenia, Spain, Sweden, Switzerland, Turkey, Ukraine and United Kingdom, Asia-pacific: Australia, China, Dubai, Hong Kong, India, Japan, Malaysia, New Zealand, Philippines, Sri Lanka, Singapore, South Korea, Taiwan, Thailand, Argentina, Bermuda, Brazil, Chile, Colombia, Costa Rica, Israel, Mexico, Puerto Rico, South Africa, and Venezuela.
Lot
50373723 50389654 50389655 50397577 50397596 50397599 50397605 50417226 50431744 50431853 50433695 50437528 50440608 50441690 50447116 50447118 50447119 50448227 50454009 50454867 50461412 50467503 50469159 50470653
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was TWINFIX" Ultra 4.5 mm PLLA-HA Suture Anchor with two #2 ULTRABRAID Sutures (blue, blue-COBRAID") with Needles; Part Number: 72202612 Biodegradable suture anchor recalled?

Yes. FDA published recall Z-2238-2013 on September 25, 2013. Reported hazard: Pinholes detected in a small number of primary pouches that may constitute a breach of the sterile barrier.

What should I do if I have this product?

Voluntary: Firm initiated · Letter