FDAFDA · Food and drugs·Recall number Z-2243-2014
CloverSnare 4-Loop Vascular Retrieval Snare. Product is packaged in a Tyvek-film sterilizable outer package and is supplied one pouch in a box. The CloverSnare 4-Loop Vascular Retriever is intended for use in the cardiovascular system to manipulate and retrieve foreign objects, including but not limited to, wire guides, coils, balloons, catheters, and filters.
Published September 3, 2014 · Recall initiated: July 17, 2014
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
This remedial action is the result of six product complaints associated with separation of the snare from the distal tip of the wire.
What to do
Voluntary: Firm initiated · Letter
Recall details
- Manufacturer
- Cook Inc.
- Distributor
- Worldwide Distribution - USA including AK, AL, AZ, CA, CT, DC, FL, GA, HI, IA, ID, IL, IN, KY, LA, MD, ME, MI, MN, MO, MS, NC, NE, NH, NJ, NV, NY, OH, OK, PA, SC, TX, UT, VA, WA, WI, WV and Internationally to: Canada, Austria, Belgium, Denmark, Germany, Great Britain, Ireland, Italy, Spain, Sweden, and Switzerland.
- Lot
- Model Number(s): VRS-6.0-9.0 Lot numbers: 3583416, 3583418, 3583422, 3583424, 3583426, 3583428, 3583430, 3583432, 3583434, 3583436, 3583440, 3583442, 3583452, 3583456, 3583458, 3583462, 3583464, 3583466, 3583468, 3583470, 3583472, 3583474, 3583476, 3583478, 3583480, 3583482, 3583484, 3583486, 3583488, 3583490, 3583492, 3583494, 3583496, 3583498, 3583500, 3583502, 3583504, 4293921, 4293923, 4293925, 4293927, 4319573, 4319575, 4319577, 4319579, 4319581, 4319583, 4319585, 4319587, 4319589, 4319591, 4572365, 3583418X, 3583430X, 3583442X, 3583442XX, 3583464XXX, 3583480XX, and 3583486X.
- Risk classification
- Class I
- Area (FDA)
- device
- Country
- United States