6.5 mm Cancellous Screw 32 mm Thread Length Sterile zimmer
Published August 27, 2014 · Recall initiated: July 22, 2014
Summary
FDA records medical device Class II recall Z-2248-2014, published August 27, 2014 and initiated July 22, 2014. Recalling firm: Zimmer, Inc.; Status: Terminated; Reported quantity: 47,682 units total; Lot: Item Number 47-4865-045, 050, 055, 060, 065, 070, 075, 080, 085, 090, 100-02. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Zimmer is recalling various trauma screws that failed a leak test due to damage of the inner cavity only.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Zimmer, Inc.
- Distributor
- National Distribution: AK, AL, AZ, CA, CT, DC, FL, GA, US VIRGIN ISLANDS, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, NE, NJ, NM, NV, NY, OH, OK, OR, PA, PR, SC, SD, TN, TX, VA, WA, WI, WV, WY Foreign Distribution: Canada, Mexico, ARGENTINA, AUSTRALIA, BELGIUM,BRAZIL, CHINA, EL SALVADOR, ENGLAND, GERMANY, INDIA, ITALY, JAPAN, JORDAN, KOREA, NEW ZEALAND, SAUDI ARABIA, SINGAPORE, SPAIN, SWITZERLAND, TAIWAN, UAE
- Lot
- Item Number 47-4865-045, 050, 055, 060, 065, 070, 075, 080, 085, 090, 100-02
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAOff-Axis Comprehensive Shoulder System Small, Augment, Off-Axis, Reamer Guide, Model/Catalog Number: 110040240July 1, 2026
- FDABrand Name: Persona Revision Product Name: Persona Revision® Trabecular Metal® Femoral Distal AugmentJuly 1, 2026
- FDAOff-Axis Alliance Glenoid, Left, 4-Peg, Augment, Reamer Guide, Model/Catalog Number: 110040444July 1, 2026
- FDAOff-Axis Comprehensive Shoulder System, Medium, Augment, Off-Axis, Reamer Guide Model/Catalog Number:July 1, 2026
- FDA6.5 mm Cancellous Screw 16 mm Thread Length Sterile zimmerAugust 27, 2014
- FDA6.5 mm Cancellous Screw Fully Threaded Sterile zimmerAugust 27, 2014
Questions about this recall
Was 6.5 mm Cancellous Screw 32 mm Thread Length Sterile zimmer recalled?
Yes. FDA published recall Z-2248-2014 on August 27, 2014. Reported hazard: Zimmer is recalling various trauma screws that failed a leak test due to damage of the inner cavity only.
What should I do if I have this product?
Voluntary: Firm initiated · Letter