Back to search
FDAFDA · Food and drugs·Recall number Z-2253-2016

Anspach G1 High Speed Electric Handpiece Air Cooled (eG1A) Use for cutting and shaping bone including bones of the cranium and spine.

Published July 27, 2016 · Recall initiated: June 15, 2016

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

The graphics at the attachment interface indicating locked and unlocked position are reversed.

What to do

Voluntary: Firm initiated · Letter

Recall details

Manufacturer
The Anspach Effort, Inc.
Distributor
Worldwide Distribution: US Distribution to states of: GA and FL; and country of: Switzerland.
Lot
Lot #'s: H29309439804, H44310033104, K09311396703, and K09311396704.
Risk classification
Class II
Area (FDA)
device
Country
United States