Anspach G1 High Speed Electric Handpiece Air Cooled (eG1A) Use for cutting and shaping bone including bones of the cranium and spine.

Published July 27, 2016 · Recall initiated: June 15, 2016

Summary

FDA records medical device Class II recall Z-2253-2016, published July 27, 2016 and initiated June 15, 2016. Recalling firm: The Anspach Effort, Inc.; Status: Terminated; Reported quantity: 18; Lot: Lot #'s: H29309439804, H44310033104, K09311396703, and K09311396704. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

The graphics at the attachment interface indicating locked and unlocked position are reversed.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
The Anspach Effort, Inc.
Distributor
Worldwide Distribution: US Distribution to states of: GA and FL; and country of: Switzerland.
Lot
Lot #'s: H29309439804, H44310033104, K09311396703, and K09311396704.
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was Anspach G1 High Speed Electric Handpiece Air Cooled (eG1A) Use for cutting and shaping bone including bones of the cranium and spine. recalled?

Yes. FDA published recall Z-2253-2016 on July 27, 2016. Reported hazard: The graphics at the attachment interface indicating locked and unlocked position are reversed.

What should I do if I have this product?

Voluntary: Firm initiated · Letter