Back to search
FDAFDA · Food and drugs·Recall number Z-2259-2026

Allura system; System Code Description (Model Numbers): Allura Xper FD10 (722003, 722010, 722026), Allura Xper FD10 OR Table (722022, 722033), Allura Xper FD10/10 (722005, 722011, 722027), Allura Xper FD20 (722006, 722012, 722028), Allura Xper FD20 Biplane (722008, 722013), Allura Xper FD20 OR Table (722015, 722023, 722035), Allura Xper FD20 Biplane OR Table (722025), Allura Xper FD20/10 (722029), Allura Xper FD20/15 (722058), Allura Xper FD20/15 OR Table (722059), Allura Xper FD20/20 (722038), Allura Xper FD20/20 OR Table (722039);

Published June 3, 2026 · Recall initiated: April 30, 2026

Summary

FDA records medical device Class II recall Z-2259-2026, published June 3, 2026 and initiated April 30, 2026. Recalling firm: PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.; Status: Ongoing; Lot: 1. Allura Xper FD10; Model: 722003; UDI-DI: N/A; Serial Numbers: 445, 1555, 1114, 1529, 472, 1349, 1594, 1311, 1472, 1187, 1700, 1031, 912, 913, 940, 285, 1637, 5, 797, 639, 452, 1585, 535, 1338, 1641, 1916, 614, 605, 492, 1612, 505, 239, 1216, 1169, 1118, 115, 519, 356, 715, 423, 1399, 789, 1509, 1016, 875, 1923, 313, 314, 1919, 1458, 369, 1176, 1230, 1925, 339, 1325, 1633, 596, 1784, 464, 796, 1928, 1025, 292, 1669, 1697, 708, 1, 1760, 1452, 346, 961, 1172, 354, 416, 363, 582, 161, 569, 922; 2. Allura Xper FD10; Model: 722010; UDI: (01)00884838059030(21)1123, (01)00884838059030(21)932, (01)00884838059030(21)239, (01)00884838059030(21)360, (01)00884838059030(21)500, (01)00884838059030(21)794, (01)00884838059030(21)939, (01)00884838059030(21)921, (01)00884838059030(21)286, (01)00884838059030(21)1440, (01)00884838059030(21)1144, (01)00884838059030(21)521, (01)00884838059030(21)33, (01)00884838059030(21)107, (01)00884838059030(21)1025, (01)00884838059030(21)668, (01)00884838059030(21)656, (01)00884838059030(21)1261, (01)00884838059030(21)1170, (01)00884838059030(21)585, (01)00884838059030(21)1128, (01)00884838059030(21)882, (01)00884838059030(21)1230, (01)00884838059030(21)280, (01)00884838059030(21)488, (01)00884838059030(21)787, (01)00884838059030(21)968, (01)00884838059030(21)890, (01)00884838059030(21)780, (01)00884838059030(21)190, (01)00884838059030(21)1181, (01)00884838059030(21)694, (01)00884838059030(21)1279, (01)00884838059030(21)344, (01)00884838059030(21)1149, (01)00884838059030(21)423, (01)00884838059030(21)1176, (01)00884838059030(21)1173, (01)00884838059030(21)1040, (01)00884838059030(21)1309, (01)00884838059030(21)763, (01)00884838059030(21)931, (01)00884838059030(21)478, (01)00884838059030(21)1298, (01)00884838059030(21)970, (01)00884838059030(21)1226, (01)00884838059030(21)290, (01)00884838059030(21)365, (01)00884838059030(21)651, (01)00884838059030(21)1292, (01)00884838059030(21)1275, (01)00884838059030(21)755, (01)00884838059030(21)671, (01)00884838059030(21)677, (01)00884838059030(21)1336, (01)00884838059030(21)477, (01)00884838059030(21)1078, (01)00884838059030(21)467, (01)00884838059030(21)1369, (01)00884838059030(21)792, (01)00884838059030(21)679, (01)00884838059030(21)485, (01)00884838059030(21)395, (01)00884838059030(21)592, (01)00884838059030(21)1312, (01)00884838059030(21)1302, (01)00884838059030(21)505, (01)00884838059030(21)714, (01)00884838059030(21)414, (01)00884838059030(21)785, (01)00884838059030(21)1283, (01)00884838059030(21)119, (01)00884838059030(21)662, (01)00884838059030(21)698, (01)00884838059030(21)690, (01)00884838059030(21)1235, (01)00884838059030(21)1233, (01)00884838059030(21)1236, (01)00884838059030(21)1234, (01)00884838059030(21)158, (01)00884838059030(21)1009, (01)00884838059030(21)992, (01)00884838059030(21)1214, (01)00884838059030(21)1064, (01)00884838059030(21)979, (01)00884838059030(21)764, (01)00884838059030(21)920, (01)00884838059030(21)788, (01)00884838059030(21)1396, (01)00884838059030(21)902, (01)00884838059030(21)1344, (01)00884838059030(21)1183, (01)00884838059030(21)453, (01)00884838059030(21)1118, (01)00884838059030(21)411, (01)00884838059030(21)1210, (01)00884838059030(21)1209, (01)00884838059030(21)1041, (01)00884838059030(21)1042, (01)00884838059030(21)1432, (01)00884838059030(21)308, (01)00884838059030(21)833, (01)00884838059030(21)1310, (01)00884838059030(21)728, (01)00884838059030(21)791, (01)00884838059030(21)480, (01)00884838059030(21)1263, (01)00884838059030(21)987, (01)00884838059030(21)986, (01)00884838059030(21)422, (01)00884838059030(21)905, (01)00884838059030(21)218, (01)00884838059030(21)1005, (01)00884838059030(21)1425, (01)00884838059030(21)1397, (01)00884838059030(21)1398, (01)00884838059030(21)1132, (01)00884838059030(21)468, (01)00884838059030(21)1442, (01)00884838059030(21)1048, (01)00884838059030(21)648, (01)00884838059030(21)646, (01)00884838059030(21)1126, (01)00884838059030(21)873, (01)00884838059030(21)252, (01)00884838059030(21)343, (01)0088. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

It was identified that HDDs used in the PCs of Azurion and Allura systems may show a decrease in performance as they age, particularly beyond six years of service. Issues with an HDD may, depending on the specific system PC in which the affected HDD is installed, result in loss of imaging functionality. Additionally, in Allura systems, this may lead to loss of motorized movement, or loss of data. In some cases, a system restart may temporarily restore functionality.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
Distributor
Worldwide distribution - US Nationwide and the countries of Canada, Afghanistan, Albania, Algeria, Argentina, Australia, Austria, Bahrain, Bangladesh, Belarus, Belgium, Bolivia, Bosnia and Herzegovina, Brazil, Brunei Darussal, Bulgaria, Chile, China, Colombia, Costa Rica, Croatia, Cuba, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, Estonia, Fiji, Finland, France, French Polynesia, Georgia, Germany, Greece, Guinea, Honduras, Hong Kong, Hungary, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kenya, Korea, South, Kosovo, Kuwait, Kyrgyzstan, Lao, Latvia, Lebanon, Libya, Lithuania, Macedonia, Malaysia, Maldives, Malta, Martinique, Mauritius, Mexico, Moldova, Mongolia, Morocco, Myanmar, Nepal, Netherlands, New Caledonia, New Zealand, Nicaragua, Nigeria, Norway, Oman, Pakistan, Palestine, Panama, Papua New Guinea, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, R®union, Romania, Russian Fed., Russian Federation, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, South Korea, Spain, Sri Lanka, Sudan, Sweden, Switzerland, Syrian Arab Republic, Taiwan, Tanzania, Thailand, Trinidad and Tobago, Tunisia, Turkey, Turkmenistan, Ukraine, United Arab Emirates, United Kingdom, Uruguay, Uzbekistan, Vietnam, Yemen.
Lot
1. Allura Xper FD10; Model: 722003; UDI-DI: N/A; Serial Numbers: 445, 1555, 1114, 1529, 472, 1349, 1594, 1311, 1472, 1187, 1700, 1031, 912, 913, 940, 285, 1637, 5, 797, 639, 452, 1585, 535, 1338, 1641, 1916, 614, 605, 492, 1612, 505, 239, 1216, 1169, 1118, 115, 519, 356, 715, 423, 1399, 789, 1509, 1016, 875, 1923, 313, 314, 1919, 1458, 369, 1176, 1230, 1925, 339, 1325, 1633, 596, 1784, 464, 796, 1928, 1025, 292, 1669, 1697, 708, 1, 1760, 1452, 346, 961, 1172, 354, 416, 363, 582, 161, 569, 922; 2. Allura Xper FD10; Model: 722010; UDI: (01)00884838059030(21)1123, (01)00884838059030(21)932, (01)00884838059030(21)239, (01)00884838059030(21)360, (01)00884838059030(21)500, (01)00884838059030(21)794, (01)00884838059030(21)939, (01)00884838059030(21)921, (01)00884838059030(21)286, (01)00884838059030(21)1440, (01)00884838059030(21)1144, (01)00884838059030(21)521, (01)00884838059030(21)33, (01)00884838059030(21)107, (01)00884838059030(21)1025, (01)00884838059030(21)668, (01)00884838059030(21)656, (01)00884838059030(21)1261, (01)00884838059030(21)1170, (01)00884838059030(21)585, (01)00884838059030(21)1128, (01)00884838059030(21)882, (01)00884838059030(21)1230, (01)00884838059030(21)280, (01)00884838059030(21)488, (01)00884838059030(21)787, (01)00884838059030(21)968, (01)00884838059030(21)890, (01)00884838059030(21)780, (01)00884838059030(21)190, (01)00884838059030(21)1181, (01)00884838059030(21)694, (01)00884838059030(21)1279, (01)00884838059030(21)344, (01)00884838059030(21)1149, (01)00884838059030(21)423, (01)00884838059030(21)1176, (01)00884838059030(21)1173, (01)00884838059030(21)1040, (01)00884838059030(21)1309, (01)00884838059030(21)763, (01)00884838059030(21)931, (01)00884838059030(21)478, (01)00884838059030(21)1298, (01)00884838059030(21)970, (01)00884838059030(21)1226, (01)00884838059030(21)290, (01)00884838059030(21)365, (01)00884838059030(21)651, (01)00884838059030(21)1292, (01)00884838059030(21)1275, (01)00884838059030(21)755, (01)00884838059030(21)671, (01)00884838059030(21)677, (01)00884838059030(21)1336, (01)00884838059030(21)477, (01)00884838059030(21)1078, (01)00884838059030(21)467, (01)00884838059030(21)1369, (01)00884838059030(21)792, (01)00884838059030(21)679, (01)00884838059030(21)485, (01)00884838059030(21)395, (01)00884838059030(21)592, (01)00884838059030(21)1312, (01)00884838059030(21)1302, (01)00884838059030(21)505, (01)00884838059030(21)714, (01)00884838059030(21)414, (01)00884838059030(21)785, (01)00884838059030(21)1283, (01)00884838059030(21)119, (01)00884838059030(21)662, (01)00884838059030(21)698, (01)00884838059030(21)690, (01)00884838059030(21)1235, (01)00884838059030(21)1233, (01)00884838059030(21)1236, (01)00884838059030(21)1234, (01)00884838059030(21)158, (01)00884838059030(21)1009, (01)00884838059030(21)992, (01)00884838059030(21)1214, (01)00884838059030(21)1064, (01)00884838059030(21)979, (01)00884838059030(21)764, (01)00884838059030(21)920, (01)00884838059030(21)788, (01)00884838059030(21)1396, (01)00884838059030(21)902, (01)00884838059030(21)1344, (01)00884838059030(21)1183, (01)00884838059030(21)453, (01)00884838059030(21)1118, (01)00884838059030(21)411, (01)00884838059030(21)1210, (01)00884838059030(21)1209, (01)00884838059030(21)1041, (01)00884838059030(21)1042, (01)00884838059030(21)1432, (01)00884838059030(21)308, (01)00884838059030(21)833, (01)00884838059030(21)1310, (01)00884838059030(21)728, (01)00884838059030(21)791, (01)00884838059030(21)480, (01)00884838059030(21)1263, (01)00884838059030(21)987, (01)00884838059030(21)986, (01)00884838059030(21)422, (01)00884838059030(21)905, (01)00884838059030(21)218, (01)00884838059030(21)1005, (01)00884838059030(21)1425, (01)00884838059030(21)1397, (01)00884838059030(21)1398, (01)00884838059030(21)1132, (01)00884838059030(21)468, (01)00884838059030(21)1442, (01)00884838059030(21)1048, (01)00884838059030(21)648, (01)00884838059030(21)646, (01)00884838059030(21)1126, (01)00884838059030(21)873, (01)00884838059030(21)252, (01)00884838059030(21)343, (01)0088
Risk classification
Class II
Area (FDA)
device
Country
Netherlands

Related recalls

Questions about this recall

Was Allura system; System Code Description (Model Numbers): Allura Xper FD10 (722003, 722010, 722026), Allura Xper FD10 OR Table (722022, 722033), Allura Xper FD10/10 (722005, 722011, 722027), Allura Xper FD20 (722006, 722012, 722028), Allura Xper FD20 Biplane (722008, 722013), Allura Xper FD20 OR Table (722015, 722023, 722035), Allura Xper FD20 Biplane OR Table (722025), Allura Xper FD20/10 (722029), Allura Xper FD20/15 (722058), Allura Xper FD20/15 OR Table (722059), Allura Xper FD20/20 (722038), Allura Xper FD20/20 OR Table (722039); recalled?

Yes. FDA published recall Z-2259-2026 on June 3, 2026. Reported hazard: It was identified that HDDs used in the PCs of Azurion and Allura systems may show a decrease in performance as they age, particularly beyond six years of service. Issues with an HDD may, depending on the specific system PC in which the affected HDD is installed, result in loss of imaging functionality. Additionally, in Allura systems, this may lead to loss of motorized movement, or loss of data. In some cases, a system restart may temporarily restore functionality.

What should I do if I have this product?

Voluntary: Firm initiated · Letter