Visual-ICE® Cryoablation System, Model FPRCH 6000 Product Usage: The Galil Medical Visual-ICE Cryoablation System is intended for cryoablative destruction of tissue during minimally invasive procedures; various Galil Medical accessory products are required to perform these procedures. The Visual-ICE System is indicated for use as a cryosurgical tool in the fields of general surgery, dermatology, neurology (including cryoanalgesia), thoracic surgery, ENT, gynecology, oncology, proctology, and urology. This system is designed to destroy tissue (including prostate and kidney tissue, liver metastases, tumors, skin lesions, and warts) by the application of extremely cold temperatures.

Published August 29, 2012 · Recall initiated: July 20, 2012

Summary

FDA records medical device Class II recall Z-2261-2012, published August 29, 2012 and initiated July 20, 2012. Recalling firm: Galil Medical, Inc.; Status: Terminated; Reported quantity: 4; Lot: VL0004 VL0014 VL0015 VL0016. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Galil Medical is recalling a number of Visual Ice Cryoablation Systems because the gas regulators may fail to regulate the gas pressure, rendering the devices inoperable.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Galil Medical, Inc.
Distributor
Worldwide Distribution - US nationwide including the state of MN and the countries of United Kingdom and Netherlands
Lot
VL0004 VL0014 VL0015 VL0016
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was Visual-ICE® Cryoablation System, Model FPRCH 6000 Product Usage: The Galil Medical Visual-ICE Cryoablation System is intended for cryoablative destruction of tissue during minimally invasive procedures; various Galil Medical accessory products are required to perform these procedures. The Visual-ICE System is indicated for use as a cryosurgical tool in the fields of general surgery, dermatology, neurology (including cryoanalgesia), thoracic surgery, ENT, gynecology, oncology, proctology, and urology. This system is designed to destroy tissue (including prostate and kidney tissue, liver metastases, tumors, skin lesions, and warts) by the application of extremely cold temperatures. recalled?

Yes. FDA published recall Z-2261-2012 on August 29, 2012. Reported hazard: Galil Medical is recalling a number of Visual Ice Cryoablation Systems because the gas regulators may fail to regulate the gas pressure, rendering the devices inoperable.

What should I do if I have this product?

Voluntary: Firm initiated · Letter