Mini-Infuser Model 150XL Single Speed Infusion Pump, an Rx battery operated syringe infusion pump; Product Usage: These product codes includes refurbished units. Intended for the controlled rate delivery of small volume parental fluids as prescribed by a physician. Usage: The Mini-Infuser System is designed for controlled intermittent administration of intravenous medication.
Published September 5, 2012 · Recall initiated: March 4, 2011
Summary
FDA records medical device Class III recall Z-2268-2012, published September 5, 2012 and initiated March 4, 2011. Recalling firm: Baxter Healthcare Corp.; Status: Terminated; Reported quantity: 26,485 units; Lot: product codes 2M8170 and 2M8170R, all serial numbers. Class III means the product is not likely to cause adverse health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Should a battery with the minimum contact diameter (.51 inches) be placed on the battery spring (.53 inches), an electrical circuit will not be completed and the Mini-Infuser will not operate. The reason being that the Mini-Infuser Battery Spring has a flat contact that measures .53 inches ¿.01 inches on the inner diameter of the spring coil. The IEC 60086-2 and ANSI 08 standards specify a minimum
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Baxter Healthcare Corp.
- Distributor
- 45,865 units
- Lot
- product codes 2M8170 and 2M8170R, all serial numbers
- Risk classification
- Class III
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAGAMBRO Capillary Dialyzer, Polyflux(R) Revaclear MAX, STERILE, Product code 110634 Product Usage: TheNovember 16, 2016
- FDASynovis VASCU-GUARD Peripheral Vascular Patch in the following sizes and product codes: 1 cm x 6 cm, productAugust 24, 2016
- FDA0.9% Sodium Chloride Injection, USP, 250 mL VIAFLEX Plastic Container Bag, Rx Only, Manufactured by BaxterJune 15, 2016
- FDADextrose Injection,USP,70% 2000 mL VIAFLEX Plastic Container Bag, Pharmacy Bulk Package, Not For DirectJune 15, 2016
- FDACaviCide, Part Numbers: 13-1025, 13-1000, 13-1024, MC-1000. Alternate brands: EnviroCide, part number:September 5, 2012
- FDAMini-Infuser Model 300XL Multispeed Infusion Pump, an Rx battery operated syringe infusion pump with an activeSeptember 5, 2012
Questions about this recall
Was Mini-Infuser Model 150XL Single Speed Infusion Pump, an Rx battery operated syringe infusion pump; Product Usage: These product codes includes refurbished units. Intended for the controlled rate delivery of small volume parental fluids as prescribed by a physician. Usage: The Mini-Infuser System is designed for controlled intermittent administration of intravenous medication. recalled?
Yes. FDA published recall Z-2268-2012 on September 5, 2012. Reported hazard: Should a battery with the minimum contact diameter (.51 inches) be placed on the battery spring (.53 inches), an electrical circuit will not be completed and the Mini-Infuser will not operate. The reason being that the Mini-Infuser Battery Spring has a flat contact that measures .53 inches ¿.01 inches on the inner diameter of the spring coil. The IEC 60086-2 and ANSI 08 standards specify a minimum
What should I do if I have this product?
Voluntary: Firm initiated · Letter