TRESTLE LUXE® Anterior Cervical Plating System, Product code KWQ. 510(k) K102820. The TRESTLE LUXE Anterior Cervical Plating System is a temporary device used to stabilize the cervical spine during bone fusion development. Device implants include a range of plate sizes and bone screws to provide the versatility required for the specific indications noted. Fixation is achieved by means of a rigid plate that is surgically attached to the spine with bone screws. The TRESTLE LUXE Anterior Cervical Plating System is intended for use in the anterior cervical spine (C2-C7). Patients with trauma (including fractures), spondylolisthesis, and pseudoarthrosis, degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies), spinal stenosis, tumors, and failed previous fusion.

Published September 5, 2012 · Recall initiated: May 29, 2012

Summary

FDA records medical device Class II recall Z-2273-2012, published September 5, 2012 and initiated May 29, 2012. Recalling firm: Alphatec Spine, Inc.; Status: Terminated; Reported quantity: 38; Lot: Product code KWQ. 510(k) K102820 Part Number 71721. Lot numbers: 6629701, 6594701, 6434301, 6434302, 6434302R1. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

On May 9th 2012 Alphatec Spine received a complaint which reported that the distal tip of the Trestle Luxe Slide Alignment Tool (Part Number 71721), does not properly fit into the hexalobe screw on the Trestle Luxe Plate.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Alphatec Spine, Inc.
Distributor
Nationwide Distribution.
Lot
Product code KWQ. 510(k) K102820 Part Number 71721. Lot numbers: 6629701, 6594701, 6434301, 6434302, 6434302R1.
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was TRESTLE LUXE® Anterior Cervical Plating System, Product code KWQ. 510(k) K102820. The TRESTLE LUXE Anterior Cervical Plating System is a temporary device used to stabilize the cervical spine during bone fusion development. Device implants include a range of plate sizes and bone screws to provide the versatility required for the specific indications noted. Fixation is achieved by means of a rigid plate that is surgically attached to the spine with bone screws. The TRESTLE LUXE Anterior Cervical Plating System is intended for use in the anterior cervical spine (C2-C7). Patients with trauma (including fractures), spondylolisthesis, and pseudoarthrosis, degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies), spinal stenosis, tumors, and failed previous fusion. recalled?

Yes. FDA published recall Z-2273-2012 on September 5, 2012. Reported hazard: On May 9th 2012 Alphatec Spine received a complaint which reported that the distal tip of the Trestle Luxe Slide Alignment Tool (Part Number 71721), does not properly fit into the hexalobe screw on the Trestle Luxe Plate.

What should I do if I have this product?

Voluntary: Firm initiated · Letter