Back to search
FDAFDA · Food and drugs·Recall number Z-2275-2026

PhaseOne Antimicrobial Solution, Model/Catalog Number: 15050

Published June 10, 2026 · Recall initiated: April 27, 2026

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

The recall is being conducted due to corrosion of the internal metallic spring within the sprayer component. This condition is caused by material incompatibility between the product solution and the stainless steel spring and may result in visible discoloration and the presence of trace levels of metallic elements (including iron, chromium, and nickel) in the solution during use.

What to do

Voluntary: Firm initiated

Recall details

Manufacturer
Oculus Technologies of Mexico, S.A. de C.V.
Distributor
US Nationwide distribution in the states of AZ, CA, CO, FL, GA, IL, IN, KY, MI, MS, NC, NJ, NY, TN, TX, UT, VA.
Lot
UDI:00860009268616 Lot: 25G263
Risk classification
Class II
Area (FDA)
device
Country
Mexico

Sell products like this?

Recalert watches your whole catalog for recalls like this one, automatically.