PhaseOne Antimicrobial Solution, Model/Catalog Number: 15050
Published June 10, 2026 · Recall initiated: April 27, 2026
Summary
FDA records medical device Class II recall Z-2275-2026, published June 10, 2026 and initiated April 27, 2026. Recalling firm: Oculus Technologies of Mexico, S.A. de C.V.; Status: Ongoing; Reported quantity: 5904 units; Lot: UDI:00860009268616 Lot: 25G263. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
The recall is being conducted due to corrosion of the internal metallic spring within the sprayer component. This condition is caused by material incompatibility between the product solution and the stainless steel spring and may result in visible discoloration and the presence of trace levels of metallic elements (including iron, chromium, and nickel) in the solution during use.
What to do
Voluntary: Firm initiated
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Oculus Technologies of Mexico, S.A. de C.V.
- Distributor
- US Nationwide distribution in the states of AZ, CA, CO, FL, GA, IL, IN, KY, MI, MS, NC, NJ, NY, TN, TX, UT, VA.
- Lot
- UDI:00860009268616 Lot: 25G263
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- Mexico
Related recalls
- FDAThe Large Volume Pump (LVP) of the Ivenix Infusion System (IIS). Model Number: LVP-0004.June 10, 2026
- FDASpinal Needle procedure kits: Material Description (Material Number) 24GA. PENCAN SPINAL BUP. TRAY W/DURAPREPJune 10, 2026
- FDACovidien Mon-a-Therm general purpose temperature probe size: 9 Fr REF: 90050June 10, 2026
- FDASpinocan Spinal Needle procedure kits: Material Description (Material Number): Spinocan 26 Ga. x 3-1/2 in. (90June 10, 2026
- FDACovidien Mon-a-Therm General Purpose Temperature Probe REF: 90044June 10, 2026
- FDADesign Options Spinal Tray Spinal Needle Anesthesia procedure kits: Material Description (Material Number):June 10, 2026
Questions about this recall
Was PhaseOne Antimicrobial Solution, Model/Catalog Number: 15050 recalled?
Yes. FDA published recall Z-2275-2026 on June 10, 2026. Reported hazard: The recall is being conducted due to corrosion of the internal metallic spring within the sprayer component. This condition is caused by material incompatibility between the product solution and the stainless steel spring and may result in visible discoloration and the presence of trace levels of metallic elements (including iron, chromium, and nickel) in the solution during use.
What should I do if I have this product?
Voluntary: Firm initiated