SmartSite Low Sorbing Infusion Set, Model#72313E The SmartSite Low Sorbing Infusion Sets are used to administer fluid and medications through a needle or catheter inserted into the patient's artery or vein. The SmartSite Low Sorbing set is comprised of components commonly found on intravascular administration sets and extension sets. It includes a 1.2 micron filter, needleless SmartSite connector, low sorbing tubing segment, pinch clamps, and luer connectors.
Published October 2, 2013 · Recall initiated: August 30, 2013
Summary
FDA records medical device Class II recall Z-2279-2013, published October 2, 2013 and initiated August 30, 2013. Recalling firm: CareFusion 303, Inc.; Status: Terminated; Reported quantity: 600 units; Lot: Lot Number 13016314. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
CareFusion is recalling the SmartSite Low Sorbing Infusion Set, Model 72313E, Lot Number 13016314, because of an incorrect filter was assembled on this lot. The correct filter is a 1.2 micron filter (blue color) but a 0.2 micron filter (clear color) may have been added instead.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- CareFusion 303, Inc.
- Distributor
- CA, FL, AZ, and TN
- Lot
- Lot Number 13016314
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDABD Pyxis Anesthesia Station 4000 REF: 338 Medication cabinetJune 10, 2026
- FDABD Pyxis MedBank Tower REF: 1145-00 Medication cabinetJune 10, 2026
- FDABD Pyxis Mini Main REF: 349 Medication cabinetJune 10, 2026
- FDABD Pyxis MedStation ES REF: 323 Medication cabinetJune 10, 2026
- FDASynthes 3.5 MM LCP(R) Distal Humerus System The Synthes 3.5 MM LCP Distal Humerus System is intended forOctober 2, 2013
- FDAABX PENTRA Calcium CP Reagent, Model # A11A01633 Calcium plays an essential role in many cell functions:October 2, 2013
Questions about this recall
Was SmartSite Low Sorbing Infusion Set, Model#72313E The SmartSite Low Sorbing Infusion Sets are used to administer fluid and medications through a needle or catheter inserted into the patient's artery or vein. The SmartSite Low Sorbing set is comprised of components commonly found on intravascular administration sets and extension sets. It includes a 1.2 micron filter, needleless SmartSite connector, low sorbing tubing segment, pinch clamps, and luer connectors. recalled?
Yes. FDA published recall Z-2279-2013 on October 2, 2013. Reported hazard: CareFusion is recalling the SmartSite Low Sorbing Infusion Set, Model 72313E, Lot Number 13016314, because of an incorrect filter was assembled on this lot. The correct filter is a 1.2 micron filter (blue color) but a 0.2 micron filter (clear color) may have been added instead.
What should I do if I have this product?
Voluntary: Firm initiated · Letter