The FL23SE Bed is an AC-powered adjustable hospital bed with four built-in electric DC motors and remote controls that is intended for medical purposes. It can be operated by the patient or the caregiver to adjust the height and surface contour of the bed. The bed includes movable and latchable siderails. The FL23SE Bed is intended for use with patients for procedures, therapy, recovery in healthcare environments, and to transport patients between bays and procedural rooms. The FL23SE Bed is intended for use with patients for procedures, therapy, recovery in healthcare environments, and to transport patients between bays and procedural rooms.
Published October 2, 2013 · Recall initiated: August 19, 2013
Summary
FDA records medical device Class II recall Z-2283-2013, published October 2, 2013 and initiated August 19, 2013. Recalling firm: Stryker Medical Division of Stryker Corporation; Status: Terminated; Reported quantity: 61 units; Lot: Model Numbers: FL23SE with nurse call option: OL235017 & OL235018 List of Affected Devices: J60267 J60425 J60009 J60001 J60094 J60095 J60096 J60348 J60349 J60350 J60351 J60352 J60353 J60354 J60360 J60361 J60362 J60363 J60364 J60365 J60366 J60367 J60368 J60369 J60370 J60371 J60372 J60373 J60374 J60375 J60376 J60377 J60378 J60379 J60380 J60381 J60382 J60383 J60384 J60385 J60386 J60387 J60388 J60389 J60390 J60391 J60392 J60393 J60394 J60395 J60396 J60397 J60398 J60399 J60400 J60401 J60402 J60403 J60404 J60405 J60512. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
The nurse call cable (Part Number: QDF23-0571) on the FL23SE Bed is located right under the opening of the bed head box. If an accessory (i.e. IV pole) is inserted into this slot, the circuit directing the nurse call system can be damaged.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Stryker Medical Division of Stryker Corporation
- Distributor
- Worldwide Distribution - USA including MI, WI, VA and internationally to Canada.
- Lot
- Model Numbers: FL23SE with nurse call option: OL235017 & OL235018 List of Affected Devices: J60267 J60425 J60009 J60001 J60094 J60095 J60096 J60348 J60349 J60350 J60351 J60352 J60353 J60354 J60360 J60361 J60362 J60363 J60364 J60365 J60366 J60367 J60368 J60369 J60370 J60371 J60372 J60373 J60374 J60375 J60376 J60377 J60378 J60379 J60380 J60381 J60382 J60383 J60384 J60385 J60386 J60387 J60388 J60389 J60390 J60391 J60392 J60393 J60394 J60395 J60396 J60397 J60398 J60399 J60400 J60401 J60402 J60403 J60404 J60405 J60512
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAProCuity bed series, model number 3009, item number: 3009PX-LE-400.December 13, 2023
- FDAProform- support surface is used to assist in the prevention and treatment of all categories/stages ofAugust 18, 2021
- FDAPrime Fifth Wheel Stretcher-stretcher may be used for minor procedures and short-term stay (treatment andAugust 18, 2021
- FDAThe Spirit Plus Bed, Model 5600 Spirit Plus /with high rail siderail optionNovember 21, 2018
- FDASynthes Oracle Spacer System Slap Hammer, Part Number 03.809.690, Lot Number 3723847 The Oracle Spacer isOctober 2, 2013
- FDAUniversal Titanium Prosthesis, Round Head, with Flex H/A titanium shoe. The ear prosthesis is implantable andOctober 2, 2013
Questions about this recall
Was The FL23SE Bed is an AC-powered adjustable hospital bed with four built-in electric DC motors and remote controls that is intended for medical purposes. It can be operated by the patient or the caregiver to adjust the height and surface contour of the bed. The bed includes movable and latchable siderails. The FL23SE Bed is intended for use with patients for procedures, therapy, recovery in healthcare environments, and to transport patients between bays and procedural rooms. The FL23SE Bed is intended for use with patients for procedures, therapy, recovery in healthcare environments, and to transport patients between bays and procedural rooms. recalled?
Yes. FDA published recall Z-2283-2013 on October 2, 2013. Reported hazard: The nurse call cable (Part Number: QDF23-0571) on the FL23SE Bed is located right under the opening of the bed head box. If an accessory (i.e. IV pole) is inserted into this slot, the circuit directing the nurse call system can be damaged.
What should I do if I have this product?
Voluntary: Firm initiated · Letter