FDAFDA · Food and drugs·Recall number Z-2289-2015

ANSPACH - Extra Coarse Diamond Ball; 8MM REF L-8DX-G1 Cutting shaping bone including spine and cranium.

Published August 5, 2015 · Recall initiated: April 13, 2015

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

G1 Dissection Tools are not compatible with all attachments listed on the labeling; G1 cutting tools are labeled with a Specified Shaft Exposure which may not be consistent when used with each attachment on the product label. Shaft exposure may be greater.

What to do

Voluntary: Firm initiated · Letter

Recall details

Manufacturer
The Anspach Effort, Inc.
Distributor
Worldwide Distribution - US including AZ, CA, CO, FL, GA, IL, KY, LA, MN, MO, NC, NJ, NY, OH, OR, PA, TX, and Internationally to Canada and China.
Lot
REF L-8DX-G1 ; ALL lot numbers
Risk classification
Class II
Area (FDA)
device
Country
United States